Wendy McDougald1,2, Christian Vanhove3, Adrienne Lehnert4, Barbara Lewellen4, John Wright5, Marco Mingarelli6, Carlos Alcaide Corral7,2, Jurgen E Schneider5, Sven Plein5, David E Newby7, Andy Welch6, Robert Miyaoka4, Stefaan Vandenberghe3, Adriana Alexandre S Tavares7,2.
Abstract
Preclinical PET/CT is a well-established noninvasive imaging tool for studying disease development/progression and the development of novel radiotracers and pharmaceuticals for clinical applications. Despite this pivotal role, standardization of preclinical PET/CT protocols, including CT absorbed dose guidelines, is essentially nonexistent. This study (1) quantitatively assesses the variability of current preclinical PET/CT acquisition and reconstruction protocols routinely used across multiple centers and scanners; and (2) proposes acquisition and reconstruction PET/CT protocols for standardization of multicenter data, optimized for routine scanning in the preclinical PET/CT laboratory.Entities:
Keywords: Hounsfield units; absorbed dose; preclinical PET/CT; recovery coefficient; standardization
Mesh:
Year: 2019 PMID: 31562220 PMCID: PMC7067528 DOI: 10.2967/jnumed.119.231308
Source DB: PubMed Journal: J Nucl Med ISSN: 0161-5505 Impact factor: 11.082
Criteria for Quantitative Analysis of PET and CT Results
| Parameter | Data |
| PET IQ phantom | |
| Uniformity | <15% |
| RCs | 1 |
| SOR | <0.20 |
| SUV | <10% bias |
| CT air/water phantom | |
| Air | 0 HU |
| Water | −1,000 HU |
| CT TEM phantom | |
| Lung | −700 HU |
| Soft tissue (adipose/muscle) | >0 HU |
| Bone (soft/cortical) | >200 HU |
*HU for TEM originally defined based on literature (27).
FIGURE 1.Recovery coefficients (RCs) for hot rods of 1, 2, 3, 4, and 5 mm of the PET IQ phantom extracted for each scanner. (A) Default reconstruction methods for scanners 1 (2D OSEM 2 iterations 16 subsets), 2 (3D OSEM 4 iterations 6 subsets, PSF), 3 (3D OSEM 2 iterations 18 subsets, PSF), 4 (3D MLEM 12 iterations), and 5 (2D MLEM 50 iterations). (B) RCs with standardization for scanner 4 (3D MLEM 25) and scanner 5 (2D MLEM 25) leaving scanners 1, 2, and 3 with the default reconstruction method. (C) RCs for each site using FBP reconstruction with the exception of scanner 4 (reconstruction option not available yet).
PET IQ, Measured Uniformity, and Spill-Over Ratios (SOR) Using the Default Reconstruction Methods and the Standardized Reconstruction Method
| Default protocol | Standardized protocol | |||||
| Scanner | Uniformity (SD%) | SOR water | SOR air | Uniformity (SD%) | SOR water | SOR air |
| 1 | 6.4±0.01 | 0.18±0.04 | 0.13±0.03 | 6.4±0.01 | 0.18±0.04 | 0.13±0.03 |
| 2 | 4.1±1.00 | 0.09±0.01 | 0.09±0.01 | 4.1±1.00 | 0.09±0.01 | 0.09±0.01 |
| 3 | 3.4±0.17 | 0.01±0.01 | 0.02±0.01 | 3.4±0.17 | 0.01±0.01 | 0.02±0.01 |
| 4 | 5.2±0.60 | 0.28±0.04 | 0.22±0.04 | 6.4±0.00 | 0.21±0.00 | 0.13±0.00 |
| 5 | 16.7±0.55 | 0.24±0.01 | 0.12±0.02 | 10.6±0.00 | 0.27±0.00 | 0.17±0.00 |
Values expressed as mean ± standard deviation, n = 3.
Measured and Expected Standard Uptake Values (SUVs) for Each Scanner Using the Default, Standardized Reconstruction Iterative Method, and FBP
| Scanner | Expected | Default | Measured/mean | Standard | Measured/mean | FBP | Measured/mean |
| Scanner 1 | 3.61±0.59 | 3.24±0.34 | 1.04±0.11 | 3.24±0.34 | 0.96±0.09 | 3.26±0.14 | 1.03±0.04 |
| Scanner 2 | 3.87±0.62 | 3.77±1.06 | 1.21±0.34 | 3.77±1.06 | 1.12±0.31 | 3.18±0.39 | 1.00±0.12 |
| Scanner 3 | 4.11±0.12 | 3.63±0.19 | 1.17±0.06 | 3.63±0.19 | 1.08±0.05 | 3.29±0.17 | 1.04±0.05 |
| Scanner 4 | 3.64±0.31 | 2.10±0.07 | 0.68±0.00 | 3.24±0.01 | 0.96±0.00 | NA | NA |
| Scanner 5 | 3.52±1.12 | 2.82±0.15 | 0.91±0.05 | 2.93±0.46 | 0.87±0.13 | 2.98±0.02 | 0.94±0.00 |
Expected SUVs are measured from the dose calibrator and decay corrected. Measured SUVs are the mean SUV value extracted from PMOD. The "Average" SUV value per scanner is the averaged of the mean SUVs per site for n = 3 measurements, expressed as mean ± standard deviation. SUV data are also presented as normalized to the mean SUV measurement per scanner.
NA = not available; FBP = filtered back projection. ANOVA: Default P < 0.001, Standard P < 0.205, FBP P < 0.388 (FBP scanner 4 not included), n = 3 per group.
Hounsfield Units (HU) Measured Using the CT Air/Water Phantom and Default/Standardized Protocols
| Scanner | Average HU water (0) | Water STDEV | Average HU air ( | Air STDEV |
| 1: Default | 133.05±5.94 | 284.35±4.07 | −967.86±5.35 | 149.97±0.73 |
| Standardized | −77.91±1.15 | 122.32±35.89 | −990.46±2.72 | 82.21±19.61 |
| 2: Default | −29.62±0.49 | 32.28±0.08 | −993.54±0.08 | 11.34±11.30 |
| Standardized | −27.88±0.40 | 35.59±0.10 | −993.29±0.05 | 12.06±0.13 |
| 3: Default | 16.97±3.68 | 43.18±0.10 | −994.98±0.61 | 15.76±6.10 |
| Standardized | 28.78±2.33 | 45.95±0.081 | −996.92±0.08 | 7.15±0.10 |
| 4: Default | 24.85±6.77 | 24.42±1.05 | −1000±0.00 | 8.85E−12 |
| Scanner 5, not calibrated to measure HU (output in linear gray scale), converted to HU | ||||
| 5: Default | −10.12 | 64.9 | −1008.26 | 92.03 |
| Standardized | −3.42 | 142.52 | −1024.19 | 73.47 |
FIGURE 2.CT TEM, air and water HU results. For each material, each data point represents a measurement from a scanner (n = 3) from 4 different sites. Densities 1.08 to 1.57 g/mL include rod sizes 2 mm and 4 mm as reported by the manufacturer. The x-axis clearly shows the spread of HU values per density (A) and displays the significant variations measured using the default protocols (P < 0.0001, 1-way ANOVA, n = 3 per group). (B) Improved precision across scanners and densities when the standardized protocol is used.
CT Absorbed Doses Determined Using Default Protocols and a Standardized Protocol for Mice and Rats
| Default (mGy) | Standard (mGy) | Measured dose difference default to standard (%) | ||||
| Scanner | Mouse | Rat | Mouse | Rat | Mouse | Rat |
| 1 | 11±0.10 | 7±0.10 | 20±0.09 | 13±0.16 | +77 | +86 |
| 2 | 40±0.11 | 28±0.02 | 31±0.23 | 21±0.08 | −23 | −23 |
| 3 | 59±0.03 | 48±0.11 | 39±0.23 | 28±0.08 | −34 | −42 |
| 4 | 32±0.18 | 15±0.10 | 56±0.76 | 25±0.05 | +71 | +60 |
| 5 | 216±0.02 | 100±0.17 | 41±0.02 | 31±0.03 | −81 | −69 |
Results expressed as mean ± standard deviation, n = 3.
Proposed Preclinical Standard Protocols for Daily Routine Use Irrespective of Scanner/Site
| Parameter | Protocol |
| PET reconstruction | Iterative algorithms OSEM or MLEM total updates (iterations*subsets or iterations) to be in the range of 24 to 36. FBP is also recommended for use in conjunction with iterative methods. |
| CT image acquisition parameters for FBP reconstruction methods | Tube voltage at 50 kVp Number of projections at 360 Exposer of 300 ms. |