| Literature DB >> 31553793 |
Huanan Wang1, Jianchi Guan2, Xiangnan Liu2,3, Yue Shi4, Qiwen Wu3, Mengzhen Luo3, Yujun Zhu5, Zizengchen Wang1, Lefeng Wang1, Yu Pan3.
Abstract
We developed a rapid fluorescent microsphere immunochromatographic test strip (FM-ICTS) assay for the quantitative detection of avian leukosis virus (ALV). A monoclonal antibody specific for the ALV major capsid protein encoded by the gag gene was coupled to label fluorescent microspheres. ALV antibodies were coated on a nitrocellulose membrane to prepare a test line for sample detection. The fluorescence signals of the test and control lines can be read either visually by exposure to UV light or using a fluorescence analyzer. ALV could be detected quantitatively using the ratio of fluorescence signals of the test and control lines (T/C). The assay threshold was determined as a T/C value of 0.0606. The fitting curve equation was established between 1 and 2,048 ng/mL P27 protein with an r2 value of 0.9998. The assay showed no cross reactivity with Newcastle disease virus, infectious laryngotracheitis virus, infectious bronchitis virus, Marek's disease virus, infectious bursal disease, Reoviridae virus, or avian influenza virus. The repeatability was satisfactory with an overall average CV of 8.65%. The Kappa coefficient between a commercial ELISA kit was 0.7031 using clinical chicken meconium samples. Thus, a simple, rapid, sensitive, and specific fluorescent microsphere immunochromatographic test strip was developed based on specific anti-capsid protein p27 monoclonal antibodies.Entities:
Keywords: T/C value; avian leukosis virus; fitting curve equation; fluorescent microsphere immunochromatographic test strip
Mesh:
Substances:
Year: 2019 PMID: 31553793 PMCID: PMC8913972 DOI: 10.3382/ps/pez547
Source DB: PubMed Journal: Poult Sci ISSN: 0032-5791 Impact factor: 3.352
Figure 1Positive and negative detection with fluorescent microsphere test strip for ALV antigens. ALV P27 proteins were added to the test strips at 10, 100, 1,000, and 10,000 ng/mL. (a) 10,000 ng/mL; (b) 1,000 ng/mL; (c) 100 ng/mL; (d)10 ng/mL; (e) PBS.
Positive and negative detection with fluorescent microsphere test strip for ALV antigens.
| c (ng/mL) | T value | C value | T/C |
|---|---|---|---|
| 10,000 | 23,729 | 3,442 | 6.8940 |
| 1,000 | 24,152 | 3,624 | 6.6645 |
| 100 | 14,860 | 4,567 | 3.2538 |
| 10 | 7,710 | 8,320 | 0.9267 |
| 0 | 39 | 10,900 | 0.0036 |
Note: c (concentration); T value (T-line value); C value (C-line value).
The threshold assay of FM-ICTS with 40 album samples from SPF chickens.
| Samples | T/C value | |||
|---|---|---|---|---|
| Negative albumin samples | 0.0134 | 0.0224 | 0.0031 | 0.0018 |
| 0.0126 | 0.0324 | 0.0909 | 0.0025 | |
| 0.0109 | 0.0236 | 0.0027 | 0.0367 | |
| 0.0306 | 0.0129 | 0.0039 | 0.0136 | |
| 0.0155 | 0.0003 | 0.0016 | 0.0258 | |
| 0.0042 | 0.0197 | 0.0025 | 0.0056 | |
| 0.0131 | 0.0021 | 0.0016 | 0.0013 | |
| 0.0017 | 0.0016 | 0.0025 | 0.0020 | |
| 0.0026 | 0.0041 | 0.0031 | 0.0012 | |
| 0.0242 | 0.0004 | 0.0008 | 0.0026 | |
| Mean = 0.0114 | ||||
| Standard deviation = 0.0164 | ||||
| Threshold = 0.0606 | ||||
Note: T (T-line value); C (C-line value).
Sensitivity assay of FM-ICTS testing ALV P27 protein.
| c (ng/mL) | T/C | Results |
|---|---|---|
| 1,024 | 5.0122 | + |
| 512 | 4.0763 | + |
| 256 | 3.2056 | + |
| 128 | 2.1832 | + |
| 64 | 1.5959 | + |
| 32 | 0.9245 | + |
| 16 | 0.5233 | + |
| 8 | 0.3028 | + |
| 4 | 0.2303 | + |
| 2 | 0.1372 | + |
| 1 | 0.0736 | + |
| 0.5 | 0.0380 | − |
| 0.25 | 0.0090 | − |
| 0.125 | 0.0003 | − |
| 0 | 0.0015 | − |
| Threshold | 0.0606 | / |
Note: c (concentration); T (T-line value); C (C-line value).
Sensitivity assay of FM-ICTS testing ALV virus.
| c(TCID50/mL) | T/C | Results |
|---|---|---|
| 1,000 | 1.8076 | + |
| 100 | 0.562 | + |
| 10 | 0.0841 | + |
| 1 | 0.0212 | / |
| Negative | 0.0118 | / |
| Threshold | 0.0606 | / |
Note: c (concentration); T (T-line value); C (C-line value).
The relationship between concentration and T/C value.
| c(ng/mL) | Log2 ( c) | T/C value | Calculated c (ng/mL) | Recovery (%) |
|---|---|---|---|---|
| 1 | 0 | 0.0625 | 1.1 | 109.08 |
| 2 | 1 | 0.0996 | 1.7 | 86.40 |
| 4 | 2 | 0.1938 | 4.5 | 112.00 |
| 8 | 3 | 0.2717 | 7.3 | 90.91 |
| 16 | 4 | 0.4510 | 14.9 | 93.20 |
| 32 | 5 | 0.7807 | 33.1 | 103.41 |
| 64 | 6 | 1.2481 | 69.3 | 108.32 |
| 128 | 7 | 1.7716 | 126.7 | 98.98 |
| 256 | 8 | 2.5112 | 243.9 | 95.29 |
| 512 | 9 | 3.6357 | 524.8 | 102.50 |
| 1,024 | 10 | 4.8734 | 1022.3 | 99.83 |
| 2,048 | 11 | 6.4601 | 2054.9 | 100.34 |
| 4,096 | 12 | 6.6667 | 2230.0 | / |
| 8,192 | 13 | 5.5414 | 1394.7 | / |
| 16,384 | 14 | 3.8901 | 609.1 | / |
Note: c (concentration); T (T-line value); C (C-line value).
Specificity assay of FM-ICTS for ALV.
| Samples | T/C | Results |
|---|---|---|
| ILTV | 0.0063 | − |
| IBV | 0.0009 | − |
| MDV | 0.0041 | − |
| IBDV | 0.0022 | − |
| AIV | 0.0046 | − |
| ALV A | 1.0305 | + |
| ALV B | 6.0246 | + |
| ALV J | 2.9043 | + |
| REOV | 0.0270 | − |
| Control | 0.0040 | − |
| ALV P27 Protein | 1.2481 | + |
| Threshold | 0.0606 |
Note: T (T-line value); C (C-line value).
Repeatability assay of FM-ICTS for ALV.
| T/C | |||||
|---|---|---|---|---|---|
| c ng/mL | Repeat 1 | Repeat 2 | Repeat 3 | Mean | CV (%) |
| 0 | 0.0152 | 0.0157 | 0.0152 | 0.0154 | 1.82 |
| 1 | 0.0598 | 0.0579 | 0.0697 | 0.0625 | 10.14 |
| 2 | 0.0912 | 0.1066 | 0.1010 | 0.0996 | 7.79 |
| 4 | 0.1879 | 0.1749 | 0.2185 | 0.1938 | 11.57 |
| 8 | 0.2389 | 0.3182 | 0.2580 | 0.2717 | 15.23 |
| 16 | 0.4308 | 0.4658 | 0.4565 | 0.4510 | 4.02 |
| 32 | 0.7826 | 0.7464 | 0.8131 | 0.7807 | 4.28 |
| 64 | 1.3010 | 1.3805 | 1.0629 | 1.2481 | 13.24 |
| 128 | 2.0188 | 1.5518 | 1.7443 | 1.7716 | 13.25 |
| 256 | 2.3415 | 2.7000 | 2.4921 | 2.5112 | 7.17 |
| 512 | 3.4857 | 3.7738 | 3.6477 | 3.6357 | 3.97 |
| 1,024 | 4.5743 | 4.4618 | 5.5842 | 4.8734 | 12.68 |
| 2,048 | 5.9342 | 6.8531 | 6.5929 | 6.4601 | 7.33 |
| Mean CV: | 8.65 | ||||
Note: T (T-line value); C (C-line value); variation coefficient (CV).
Clinical sample detection with FM-ICTS and ELISA.
| Result | ELISA positive | ELISA negative | Total | Kappa coefficient |
|---|---|---|---|---|
| FM-ICTS positive | 68 | 23 | 91 | 0.7031 |
| FM-ICTS negative | 27 | 866 | 893 | |
| Total | 95 | 889 |