Literature DB >> 31550188

Prehospital Laryngeal Tube Airway Device Placement Resulting in Hypopharyngeal Perforation: A Case Report.

Matthew Chinn, Lynda Biedrzycki.   

Abstract

A 26-year-old female patient presented in cardiac arrest from presumed opioid overdose. An Ambu King LTS-D laryngeal device was placed by EMS providers for airway management during the resuscitation. There was no documented difficulty with placement and breath sounds and waveform capnography were consistent with appropriate placement. The resuscitation was terminated on scene after extensive resuscitative efforts by the EMS crew. Upon autopsy of the patient, it was discovered that the laryngeal tube device had caused a deep 5 cm perforation to the left piriform recess. The laryngeal tube had bent and was pushed into the perforation in the piriform recess; had the patient had regain of spontaneous circulation this could have caused significant morbidity. Laryngeal tube airway devices have shown increased usage in healthcare settings, in particular in the prehospital arena. Studies of these airway devices have shown they have quick insertion times, high success rates, and low complications. Tongue swelling and minor trauma are common complications of laryngeal tube airway devices. The case report describes a rare, yet potentially life-threatening, complication of laryngeal tube airway device placement- hypopharyngeal injury. If unrecognized, this injury could lead to serious complications. Providers should be aware of the common and uncommon injuries that are associated with prehospital laryngeal tube airway device placement.

Entities:  

Keywords:  King airway; Laryngeal tube; complication; prehospital; supraglottic airway

Mesh:

Year:  2019        PMID: 31550188     DOI: 10.1080/10903127.2019.1671565

Source DB:  PubMed          Journal:  Prehosp Emerg Care        ISSN: 1090-3127            Impact factor:   3.077


  1 in total

1.  Idiopathic hypopharyngeal perforation presenting as supraglottitis and parapharyngeal abscess: case report and literature review.

Authors:  Heerani Woodun; Heyrumb Woodun; Neeraj Sethi; Anthony Simons
Journal:  J Surg Case Rep       Date:  2022-01-11
  1 in total