Use of 3D-Printed Heel Support Insoles Based on Arch Lift Improves Foot Pressure Distribution in Healthy People.

Hui Jin1, Rui Xu2, Shuxin Wang2, Jincheng Wang3.   

Abstract

BACKGROUND 3D-printed insoles are widely used. This study was conducted to test a customized three-dimensional (3D)-printed heel support insole based on arch lift and to investigate whether the pressure distribution on the sole was improved while maintaining foot function. MATERIAL AND METHODS The design was based on a 3D plantar contour scanning modeling technique. Thirty healthy male participants walked along a 10-m track under 3 self-controlled interventions. A customized 3D-printed heel support insole based on arch lift was inserted into the socks for the experimental condition A. For condition B, a customized 3D-printed heel-supporting insole was inserted into the socks, and a standardized pre-made heel-supporting insole was inserted into the socks as a control (condition C). We used the Footscan® pressure plate to measure the plantar parameters in the forefoot contact and foot flange phases in each condition. RESULTS Compared with condition B and the control condition, the peak pressure under the heel was significantly lower in condition A (P<0.05), and the peak pressure in the midfoot region was not significantly increased (P>0.05). CONCLUSIONS The biomechanical properties of the customized 3D-printed heel support are better than those of the traditional heel support insole, especially when there is a need for an additional increase in heel height. Patients do not decrease midfoot motion function while using this customized insole.

Entities:  

Mesh:

Year:  2019        PMID: 31549689      PMCID: PMC6777389          DOI: 10.12659/MSM.918763

Source DB:  PubMed          Journal:  Med Sci Monit        ISSN: 1234-1010


Background

Heel support is used as a recovery method for patients with some lower-extremity diseases, such as Achilles tendinitis [1] and leg length difference [2]. The increased heel height from the lift reduces the distance between the calcaneus and the tibia, as well as wear on the plantar aponeurosis [3]. However, most heel supports do not provide arch lift, resulting in insufficient support for the midfoot area. Each area of the sole has a corresponding biomechanical function, and the heel is primarily involved in absorbing shock [4]. Many studies have focused on the motor control of the heel [5,6] and forefoot load mitigation [7,8]. The midfoot, which connects the forefoot and the rearfoot, has received little attention compared to the heel and forefoot. The midfoot includes the lateral and medial arches, and the midfoot arch acts as a spring during exercise, which allows the storage of energy and rebound to benefit gait efficiency [9]. Additionally, midfoot behavior is very important before the load is shifted to the forefoot. During normal foot movements, the side longitudinal arch is always touching the ground, which is important for stabilizing the movement from heel to toe [10]. Therefore, we hypothesized that decreased stability caused by lift of the heel is associated with reduced contact of the midfoot lateral arch with the ground. The midfoot support has good biomechanical advantage in many aspects [11-14], and it can maintain the stability of motion and increase the contact area between the midfoot and the ground. To improve midfoot function when using the heel support, we designed and tested a custom three-dimensional (3D)-printed arch lift design. Arched supports have been widely used for midfoot improvement, with good biomechanical effects such as redistributing plantar pressure and reducing knee joint retraction torque [15]. Low-pressure areas can transfer the burden of high-pressure areas [16]. Therefore, to more effectively distribute the plantar pressure of patients with lower-limb diseases, we designed a heel support device based on an increase in the midfoot contact area. We hypothesized that a customized 3D-printed heel support insole based on arch lift has the function of maintaining the midfoot. For example, it can reduce the peak pressure of the forefoot and midfoot, the reach and velocity of the medial-lateral center of pressure (COP) [17], and the negative impact of simply raising the heel. Considering the special nature of the 3D-printed insole, we also set up a set of 3D-printed heel support insoles without arch lift. The aim of this research was to study the effect of a customized 3D-printed heel support insole based on arch lift on foot pressure distribution and motion stability control. Our results may have implications for relevant orthopedic insole design.

Material and Methods

Subjects and ethics

Thirty healthy men, with a mean age of 20.52 years (standard deviation [SD]=1.50), participated in this study. Their average mass was 62.74 kg (SD=1.90), with an average height of 170.42 cm (SD=1.80). The subjects had no history of lower-limb injury within 1 year and all subjects’ arches were in the normal range, with arch index (AI) ranging from 0.23 to 0.29. The AI is based on the foot model data generated by Footscan® (RSscan International, Belgium) pressure plates. All participants provided written informed consent before the beginning of the study and the study was approved by the local ethics committee.

Evaluation equipment

We used the Footscan® system (RSscan International, Belgium, 1×1 m Footscan® pressure plate with 8192 resistance sensors, 250 Hz frequency) to measure COP position changes, COP moving speed, peak pressure, and peak contact area. Figure 1A–1C show the area of the plantar, foot contact phases, and the range of the COP. The displacement of the center of pressure in the medial-lateral is relatively vertical to the x-axis of the longitudinal foot axis, which is set as the base line from the middle of the heel to the forefoot, between the 2nd and 3rd metatarsal heads. The Bodyarch® The Foot Scanning System (China) was used to measure the size of the foot.
Figure 1

(A) The Footscan® plantar pressure system divides the sole into 10 areas: the big toe (T1), toes 2–5 (T2–T5), from the 1st to the 5th metatarsal (M1, M2, M3, M4, and M5), midfoot (MF), medial heel (H1), and lateral heel (H2). For the convenience of analysis, in addition to the toes, the sole was divided into 3 areas: the forefoot area (FF), the midfoot area (MF), and the rear foot (RF) area. (B) The plantar movement has 4 phases: initial contact phase (ICP), forefoot contact phase (FFCP), foot flange phase (FFP) and forefoot push-out phase (FFPOP). (C) The center of pressure (COP) varies and this range is shown on the x-axis. (D) Customized 3D printed heel support and midfoot lift insole for experimental condition A.

Design of the heel support

The traditional hand-drawn arched support structure based on the barefoot model may not conform to the ideal biomechanical structure when used for heel support [18]. By capturing the size of the foot with a 3D scanner, the heel support and arch lift structure can be more accurately designed. Based on previous research [5,18-21], a customized 3D-printed heel support insole was designed with the following parameters: (1) a height of 24 mm; (2) material with moderate hardness (using EVA with A30 hardness); and (3) an arched arch lift and a raised heel structure. The subject stood in the center of the scanning plate and raised the heel 24 mm to capture the contour of the foot through the Bodyarch® Foot Scanning System. The insole design for experimental conditions A and B was based on the Bodyarch Designer® software. The arch lift of experimental condition A was designed to achieve full support by adjusting the lateral longitudinal portion of the arch to make the insole fully contact the ground. Experimental condition B did not utilize the arch lift structure, and the control condition used a standardized pre-made heel support insole. The model was designed utilizing the Bodyarch X1 printer®. Figure 1D shows the customized 3D printed heel support insole based on arch lift.

Process

Subjects were asked to wear standardized socks to walk along a 10-m walkway, and self-control was performed according to the following 3 conditions: (1) insert customized 3D-printed heel supports and midfoot support insoles in the socks of experimental condition A subjects; (2) insert custom 3D-printed heel support insoles into the socks of subjects in experimental condition B; (3) standardized pre-made heel support insoles were inserted into the socks of the control condition. After walking for 10 min along the track, participants passed the Footscan® plantar pressure scan plate at a normal walking speed. The pressure was recorded by the Footscan® system and 4 successful walks under each condition were recorded for each participant.

Data analyses

To reduce error, all data analysis only considered the left foot. The foot data – force-time (s), peak pressure (Kpa), and maximum contact area (cm2) – were recorded by the Footscan® pressure test system, which included 10 areas of the sole: the big toe (T1), toes 2–5 (T2–T5), from the 1st to the 5th metatarsal (M1, M2, M3, M4, and M5), midfoot (MF), medial heel (H1), and the lateral heel (H2). To facilitate analysis, in addition to the toes, the sole was divided into 3 regions: the forefoot region (FF), the midfoot region (MF), and the heel (RF) region. Plantar motion has 4 phases: initial contact phase (ICP), forefoot contact phase (FFCP), foot flange phase (FFP), and forefoot push-out phase (FFPOP). We analyzed the COP data during the FFCP and the FFP because maintaining a dynamic stability is most critical at these stages [21-23]. The FFCP period starts with the first metatarsal contact with the floor and ends when all of the humeral head regions are in contact with the floor. The calculation range of the COP is the absolute value of the difference between the maximum and minimum x-coordinate values of the COP during the phases.

Statistical analyses

Data analyses were done using the Social Science Statistics Program (SPSS) 24.0th Edition (SPSS, Inc., Chicago, IL, USA). One-way analysis of variance (ANOVA) was used to analyze the difference in maximum plantar pressure and COP variables in condition A and B and the control condition. P<0.05 was considered statistically significant for the variables under each condition. In Figures 2–6, error bars: ±SD; * P<0.05.
Figure 2

Contact area (CA) of the 3 plantar areas in experimental conditions A and B and the control condition.

Figure 3

Peak pressure (PP) of the 3 plantar areas in experimental conditions A and B and the control condition.

Figure 4

Percentage of force-time integral (FTI) in the 3 plantar areas in experimental conditions A and B and the control condition.

Figure 5

The range of medial-lateral center of pressure (ML-COP) during the forefoot contact phase (FFCP) and foot flange phase (FFP) in experimental conditions A and B and the control condition.

Figure 6

The velocity of the medial-lateral center of pressure (ML-COP) during forefoot contact phase (FFCP) and foot flange phase (FFP) in experimental conditions A and B and the control condition.

Results

In Figures 2 and 3, comparisons of the peak pressure (PP) and contact area (CA) of the forefoot, midfoot, and heel of experimental conditions A and B and the control condition are shown. The CA under the middle foot of experimental condition A was significantly increased compared with the control condition (P<0.05). Compared with the control condition, there was not a significant increase in the peak pressure of the forefoot and midfoot regions (P>0.05) in experimental conditions A and B. Conversely, in the heel region, the peak pressures in experimental conditions A and B were significantly lower than in the control condition (P<0.05). Figure 4 shows the percentage of force-time integrals (FTI) for the 3 plantar regions of experimental condition A, experimental condition B, and the control condition. The FTIs of the forefoot areas of experimental condition A and experimental condition B were significantly smaller than those of the control condition (P<0.05). No significant differences were found in the percentage of FTIs between the 3 conditions (P>0.05). Figures 5 and 6 show the extent and speed of the medial-lateral’s center of pressure in the forefoot contact phase (FFCP) and foot flange phase (FFP) in experimental conditions A and B and control condition, respectively. In the FFCP, the medial-lateral COP velocity and range of experimental condition A were significantly lower than those of experimental condition B and the control condition (P<0.05). In FFP, although the COP velocity of experimental condition A did not change significantly, the range of COP was significantly lower in experimental condition B and control condition than in condition A (P<0.05).

Discussion

Patients with foot disease need to wear a special insole to recover. Going barefoot only aggravates the disease, so the main purpose of this experiment was to compare the effects of different insoles on plantar pressure. The experimental results of healthy individuals are instructive for patients. In healthy people, the pressure distribution on the plantar area is balanced and no additional intervention is needed to change the pressure distribution. However, for the patient, the patient’s own plantar pressure distribution has changed and additional intervention is needed to improve the plantar pressure distribution. The result of wearing insoles for normal subjects can be used to indicate the change in pressure on the sole of the insole. Relevant experiments have been conducted to select normal people to participate and have produced many instructive results [5,11,21], so normal subjects can be used to conduct experiments to provide guidance for patients. Heel elevation significantly increases midfoot contact area [5]. This is partly because the heel supports the foot–floor contact area to the metatarsophalangeal joint, leaving the midfoot and rearfoot off the ground, so the heel support insole does not support the midfoot well. The customized 3D-printed insole, which simply lifts the heel, also insufficiently stabilizes the midfoot when the heel is raised. In this study, compared with the control condition and experimental condition B, the COP velocity in experimental condition A was significantly lower. Conversely, experimental condition B and the control condition exhibited no significant difference in these 2 aspects, indicating that the designed arch lift structure was maintained and effective in terms of exercise stability. The customized insole based on the arch lift may be used to compensate for the adverse effects of reduced stability of the midfoot while raising the heel. Since the contact area of experimental condition A was significantly higher than that of the control condition, the hypothesis that the decreased stability is related to the decrease in contact between the midfoot lateral arch and the ground is supported. Studies have shown that due to the high peak pressure under the forefoot, heel elevation may negatively affect the forefoot [5,24]. However, in this study, the peak pressure of the forefoot remained almost unchanged in the experimental condition A compared with experimental condition B and the control condition. This can be explained by the foot arch lift sharing a portion of the load. The 3D-printed heel section also helps protect the soft tissue of the heel and absorbs the impact on the heel [4,25,26]. The midfoot links the forefoot and the rearfoot, so in the load transfer phase, the midfoot keeps the balance of the sole [27,28]. Increasing the range of the COP in the medial-lateral direction shows a greater need for controlling stability in the lower limbs [29-31]. A larger range of changes in COP and faster speeds indicate that the stability of motion is lower. The results show that, compared with experimental condition B in the stability control phase [21,22,32] (i.e., the FFCP and the FFP), the range of the inner and outer COP of the experimental condition A was significantly lower than that of the experimental condition B and the control condition. The COP rate of experimental condition A was significantly lower than the control condition, indicating that the customized 3D-printed heel support insole based on arch lift improved control of balance in the contact and flange phases. Although we demonstrated that a customized 3D-printed heel support insole based on arch lift improves foot pressure distribution and improves stability, there are some inevitable limitations in this study. First, we simply measured foot pressure data and COP variation range and speed to assess foot function without recording 3D motion data. As a result, we were unable to study the effects of the new insole on the biomechanics of lower-limb joints. Second, although the main purpose of our study was to compare the differences between wearing different insoles, it is still important to compare these with barefoot conditions, as manual intervention insoles may have some adverse effects. Since our sample size was small, the error was large when we made a comprehensive comparison. We will expand the sample in future experiments. In future research, we will further analyze the effects of customized 3D-printed heel support insoles based on arch lift on the wearer’s joint angle and torque, especially for patients with related lower-extremity diseases requiring rehabilitation.

Conclusions

Use of customized 3D-printed heel support insoles based on arch lift improves biomechanical performance in plantar pressure distribution and the stability of movement, while reducing peak pressure of the heel and midfoot and forefoot load. This new type of arch lift design method may have a positive impact on the related orthopedic insole design.
  32 in total

1.  Changes in gait and plantar foot loading upon using vibrotactile wearable biofeedback system in patients with stroke.

Authors:  Christina Zong-Hao Ma; Yong-Ping Zheng; Winson Chiu-Chun Lee
Journal:  Top Stroke Rehabil       Date:  2017-09-27       Impact factor: 2.119

Review 2.  Insertional Achilles Tendinopathy.

Authors:  Gage M Caudell
Journal:  Clin Podiatr Med Surg       Date:  2016-12-23       Impact factor: 1.231

3.  Lower extremity joint loads in habitual rearfoot and mid/forefoot strike runners with normal and shortened stride lengths.

Authors:  Elizabeth R Boyer; Timothy R Derrick
Journal:  J Sports Sci       Date:  2017-05-08       Impact factor: 3.337

4.  Midfoot arthritis.

Authors:  Amar Patel; Smita Rao; Deborah Nawoczenski; Adolf S Flemister; Benedict DiGiovanni; Judith F Baumhauer
Journal:  J Am Acad Orthop Surg       Date:  2010-07       Impact factor: 3.020

Review 5.  Plantar heel pain.

Authors:  Andrew J Rosenbaum; John A DiPreta; David Misener
Journal:  Med Clin North Am       Date:  2013-12-10       Impact factor: 5.456

6.  Endoscopic Decompression of the First Branch of the Lateral Plantar Nerve and Release of the Plantar Aponeurosis for Chronic Heel Pain.

Authors:  Tun Hing Lui
Journal:  Arthrosc Tech       Date:  2016-06-06

7.  Are custom-made foot orthoses of any interest on the treatment of foot pain for prolonged standing workers?

Authors:  Tristan Tarrade; Fabrice Doucet; Nicolas Saint-Lô; Maxime Llari; Michel Behr
Journal:  Appl Ergon       Date:  2019-05-28       Impact factor: 3.661

8.  A comparison of the spatiotemporal parameters, kinematics, and biomechanics between shod, unshod, and minimally supported running as compared to walking.

Authors:  Everett B Lohman; Kanikkai Steni Balan Sackiriyas; R Wesley Swen
Journal:  Phys Ther Sport       Date:  2011-10-17       Impact factor: 2.365

9.  Effectiveness of a heel cup with an arch support insole on the standing balance of the elderly.

Authors:  Tzu-Hsuan Chen; Li-Wei Chou; Mei-Wun Tsai; Ming-Jor Lo; Mu-Jung Kao
Journal:  Clin Interv Aging       Date:  2014-02-20       Impact factor: 4.458

10.  Complementary mechanisms for upright balance during walking.

Authors:  Hendrik Reimann; Tyler D Fettrow; Elizabeth D Thompson; Peter Agada; Bradford J McFadyen; John J Jeka
Journal:  PLoS One       Date:  2017-02-24       Impact factor: 3.240

View more
  3 in total

1.  Effects of Custom-made Insoles on Plantar Biomechanics and Upper Extremity Muscle Performance.

Authors:  Yi Xu; Qing-Hua Hou; Xiu-Lan Han; Chu-Huai Wang; Dong-Feng Huang
Journal:  Curr Med Sci       Date:  2021-11-30

2.  Design of Personalized Cervical Fixation Orthosis Based on 3D Printing Technology.

Authors:  Yangyang Xu; Xiangyu Li; Yafei Chang; Yi Wang; Lifang Che; Guopeng Shi; Xiaofen Niu; Haiyan Wang; Xiaohe Li; Yujie He; Baoqing Pei; Guoqiang Wei
Journal:  Appl Bionics Biomech       Date:  2022-04-30       Impact factor: 1.664

3.  Design and Preliminary Validation of Individual Customized Insole for Adults with Flexible Flatfeet Based on the Plantar Pressure Redistribution.

Authors:  Yangzheng Jiang; Duojin Wang; Jiming Ying; Pengfei Chu; Yu Qian; Wenming Chen
Journal:  Sensors (Basel)       Date:  2021-03-04       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.