Complications of the urinary incontinence system ATOMS: description of risk factors and how to prevent these pitfalls.

Sandra Mühlstädt1, Javier C Angulo2, Nasreldin Mohammed3, André Schumann3, Paolo Fornara3.   

Abstract

BACKGROUND: We report on our multicentre evaluation of the occurrence and management of complications with the ATOMS device (A.M.I., Austria). PATIENTS AND METHODS: Between 10/09 and 09/18, a total of 187 patients with persistent postoperative stress urinary incontinence received an ATOMS device in Madrid (n = 101) or Halle (n = 86). This prospective evaluation was carried out consecutively. In addition to complications, patient age, BMI, comorbidities and previous treatments for prostate cancer and urinary incontinence were examined. Statistical analysis was performed with GraphPad Prism 7® (GraphPad Software, Inc., La Jolla, USA); p < 0.05 was considered significant.
RESULTS: The average age of the overall population was 69.4 years, with a mean CCI of 4.3. In total, 18.2% of the patients had diabetes mellitus, 21.9% were irradiated, and 14.1% and 22.9% underwent previous surgery for urinary incontinence and urethral stricture, respectively. The overall success rate of the device was 80.2%. A total of 51 postoperative complications (51/187, 27.3%) were observed, with 18 grade I (9.6%), 1 grade II (0.5%), 3 grade IIIa (1.6%) and 29 grade IIIb (15.5%) complications. The most serious postoperative complications were primary wound infection (8/187, 4.3%) and long-term cutaneous erosion with subsequent infection (15/187, 8.02%). The main risk factors for complications were previous radiotherapy (p = 0.003) and surgery for urethral stricture (p = 0.017). No other parameters were significant in bivariate analysis (p > 0.05).
CONCLUSION: The most severe complications of ATOMS implantation were primary wound infection and cutaneous erosion, especially in the presence of previous radiotherapy. Particular focus on performing hygienic and accurate implantation techniques is recommended.

Entities:  

Keywords:  ATOMS; Complications; Male stress urinary incontinence; Management; Risk factors

Mesh:

Year:  2019        PMID: 31542824     DOI: 10.1007/s00345-019-02962-w

Source DB:  PubMed          Journal:  World J Urol        ISSN: 0724-4983            Impact factor:   4.226


  2 in total

1.  Systematic review and meta-analysis comparing Adjustable Transobturator Male System (ATOMS) and male Readjustment Mechanical External (REMEEX) system for post-prostatectomy incontinence.

Authors:  Javier C Angulo; Sonia Ruiz; Martín Lozano; Ignacio Arance; Miguel Virseda; David Lora
Journal:  World J Urol       Date:  2020-06-11       Impact factor: 4.226

2.  Excessive fibrosis and non-function after ATOMS (adjustable transobturator male system): A case of capsular contracture?

Authors:  Marco Pedrini; Jannik Stuehmeier; Lukas Andrius Jelisejevas; Alexandra Gulacsi; Wolgang Horninger; Peter Rehder
Journal:  Urol Case Rep       Date:  2020-05-06