Literature DB >> 31539347

Practical approach to method verification in plasma and validation in cerebrospinal fluid under accreditation using a flexible scope in molecular virology: setting up the HIV, HBV and HCV Aptima™ Quant Dx assays.

Caroline Lefeuvre1, Adeline Pivert1, Cong T Tran1, Françoise Lunel-Fabiani1, Alexandra Ducancelle1, Hélène Le Guillou-Guillemette1.   

Abstract

Background Our laboratory obtained the ISO 15189 accreditation for the plasmatic HIV-1, HBV and HCV viral load (VL) using the m2000 RealTime™ system, which was recently changed for the platform Panther®. Here, we discuss a strategy for performing method validation/verification very quickly. Methods We performed the mandatory (repeatability, internal quality assessment [IQA], measurement uncertainty [MU]) and optional technical verifications for CE/IVD assays using the flexible scope range A. We also performed the mandatory assays for the validation of HIV-1 VL in the cerebrospinal fluid (CSF) using the flexible scope range B. The change was checked by following up on the turnaround time (TAT). Results The coefficient of variation (CV%) for repeatability and IQA complied with the limit of 0.25 log. The MU results ranged from 0.04 to 0.25 log copies or IU/mL. The comparisons of methods showed excellent correlations (R2 = 0.96 for the three parameters) but a delayed centrifugation on HCV VL showed variations of up to 2 log IU/mL. An excellent linearity for HIV-1 in the CSF was obtained from 1.5 to 5 log copies/mL with R2 = 0.99. The TAT increased (84%-98%) in routine usage. Conclusions The three Aptima assays are well suited for routine laboratory use and can be integrated within less than 2 weeks in accordance with flexible scope range A. Our data allows us to confidently perform HIV-1 VL in CSF following flexible scope range B. Finally, we provide an organizational guide for flexible scope management in molecular virology within a short time frame.

Entities:  

Keywords:  HBV; HCV; HIV-1; ISO 15189; flexible scope management; method validation; method verification; viral load

Mesh:

Substances:

Year:  2020        PMID: 31539347     DOI: 10.1515/cclm-2019-0502

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  1 in total

1.  Comparison of performance between three SARS-CoV-2 molecular assays (Aptima™, Laboratory Developed Test-Fusion, and R-GENE®) with special attention to turnaround time, a key point in laboratory management.

Authors:  Caroline Lefeuvre; Adeline Pivert; Emilie Przyrowski; Elise Bouthry; Estelle Darviot; Rafaël Mahieu; Françoise Lunel-Fabiani; Alexandra Ducancelle; Hélène Le Guillou-Guillemette
Journal:  J Med Virol       Date:  2022-03-07       Impact factor: 20.693