Confirmation of endotracheal tube placement using disposable fiberoptic bronchoscopy in the emergent setting.

Avir Mitra1, Asaf Gave2, Kelsey Coolahan1, Thomas Nguyen1.   

Abstract

BACKGROUND: Patients intubated in the prehospital setting require quick and definitive confirmation of endotracheal (ET) tube placement upon arrival to the emergency department (ED). Direct and adjunct strategies exist, but each has limitations and there is no definitive gold standard. The utility of bronchoscopy in ED intubation has been studied, but scant literature exists on its use for ET tube confirmation. This study aims to assess effectiveness, ease and speed with which ET tube placement can be confirmed with disposable fiberoptic bronchoscopy.
METHODS: Emergency medicine residents recruited from a 3-year urban residency program received 5 minutes of active learning on a simulation mannequin using a disposable, flexible Ambu aScope interfaced with a monitor. With residents blinded, the researcher randomly placed the ET tube in the trachea, esophagus or right mainstem. Residents identified ET tube position by threading the bronchoscope through the tube and viewing distal anatomy. Each resident underwent 4 trials. Accuracy, speed and perceptions of difficulty were measured.
RESULTS: Residents accurately identified the location of the ET tube in 88 out of 92 trials (95.7%). The median time-to-guess was 7.0 seconds, IQR (5.0-10.0). Average perceived difficulty was 1.6 on a scale from 1-5 (1 being very easy and 5 being very difficult). No tubes were damaged or dislodged.
CONCLUSION: While simulation cannot completely replicate the live experience, fiberoptic bronchoscopy appears to be a quick and accurate method for ET tube confirmation. Further studies directly comparing this novel approach to established practices on actual patients are warranted.

Entities:  

Keywords:  Airway; Bronchoscopy; Simulation; Tube confirmation

Year:  2019        PMID: 31534594      PMCID: PMC6732169          DOI: 10.5847/wjem.j.1920-8642.2019.04.003

Source DB:  PubMed          Journal:  World J Emerg Med        ISSN: 1920-8642


  23 in total

1.  Fibre-optic aided bougie (FAB) for simulated difficult tracheal intubation.

Authors:  J W Chiu; M H Goh; P C Ip-Yam
Journal:  Ann Acad Med Singapore       Date:  2000-01       Impact factor: 2.473

2.  Multicenter study of a portable, hand-size, colorimetric end-tidal carbon dioxide detection device.

Authors:  J P Ornato; J B Shipley; E M Racht; C M Slovis; K D Wrenn; P E Pepe; S L Almeida; V F Ginger; T V Fotre
Journal:  Ann Emerg Med       Date:  1992-05       Impact factor: 5.721

Review 3.  Esophageal intubation: a review of detection techniques.

Authors:  P K Birmingham; F W Cheney; R J Ward
Journal:  Anesth Analg       Date:  1986-08       Impact factor: 5.108

4.  End-Tidal Carbon Dioxide Use for Tracheal Intubation: Analysis From the National Emergency Airway Registry for Children (NEAR4KIDS) Registry.

Authors:  Melissa L Langhan; Beth L Emerson; Sholeen Nett; Matthew Pinto; Ilana Harwayne-Gidansky; Kyle J Rehder; Conrad Krawiec; Keith Meyer; John S Giuliano; Erin B Owen; Keiko M Tarquinio; Ron C Sanders; Michael Shepherd; Gokul Kris Bysani; Asha N Shenoi; Natalie Napolitano; Sandeep Gangadharan; Simon J Parsons; Dennis W Simon; Vinay M Nadkarni; Akira Nishisaki
Journal:  Pediatr Crit Care Med       Date:  2018-02       Impact factor: 3.624

5.  Esophageal detector device versus detection of end-tidal carbon dioxide level in emergency intubation.

Authors:  W P Bozeman; D Hexter; H K Liang; G D Kelen
Journal:  Ann Emerg Med       Date:  1996-05       Impact factor: 5.721

6.  Comparison of arterial-end-tidal PCO2 difference and dead space/tidal volume ratio in respiratory failure.

Authors:  M K Yamanaka; D Y Sue
Journal:  Chest       Date:  1987-11       Impact factor: 9.410

7.  Endotracheal tube placement confirmation: 100% sensitivity and specificity with sustained four-phase capnographic waveforms in a cadaveric experimental model.

Authors:  Salvatore Silvestri; Jay G Ladde; James F Brown; Jesus V Roa; Christopher Hunter; George A Ralls; Linda Papa
Journal:  Resuscitation       Date:  2017-01-19       Impact factor: 5.262

8.  Verification of endotracheal tube placement with colorimetric end-tidal CO2 detection.

Authors:  B A MacLeod; M B Heller; J Gerard; D M Yealy; J J Menegazzi
Journal:  Ann Emerg Med       Date:  1991-03       Impact factor: 5.721

9.  A simple blind placement of the left-sided double-lumen tubes.

Authors:  Zhi Jun Zong; Qi Ying Shen; Yao Lu; Yuan Hai Li
Journal:  Medicine (Baltimore)       Date:  2016-11       Impact factor: 1.889

10.  Comparison of intubation times using a manikin with an immobilized cervical spine: Macintosh laryngoscope vs. GlideScope vs. fiberoptic bronchoscope.

Authors:  Jung-In Ko; Sang Ook Ha; Min Seok Koo; Miyoung Kwon; Jieun Kim; Jin Jeon; So Hee Park; Sangwoo Shim; Youjin Chang; Taejin Park
Journal:  Clin Exp Emerg Med       Date:  2015-12-28
View more
  1 in total

Review 1.  Screening Performance Characteristics of Ultrasonography in Confirmation of Endotracheal Intubation; a Systematic Review and Meta-analysis.

Authors:  Mehrdad Farrokhi; Bardia Yarmohammadi; Amir Mangouri; Yasaman Hekmatnia; Yaser Bahramvand; Moein Kiani; Elham Nasrollahi; Milad Nazari-Sabet; Niusha Manoochehri-Arash; Maria Khurshid; Shima Mosalanejad; Vida Hajizadeh; Reza Amani-Beni; Masoumeh Moallem; Maryam Farahmandsadr
Journal:  Arch Acad Emerg Med       Date:  2021-10-26
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.