This study evaluated a novel robotic direct reading method that used a real-time location system to measure the spatial-concentration distribution of volatile organic compounds (VOCs) in a chemistry laboratory. The CEMWIP II is a custom-made sensor that measures VOCs, temperature, humidity, and location, sending data wirelessly in real time to a remote location for display and storage. In this study, the CEMWIP II device was mounted on a robotic platform to create a CEMWIP II-mobile platform. The autonomous mobile platform was released from a corner of the room and allowed to travel randomly along an open floor with the goal of characterizing the spatial distribution of VOCs and identifying their sources in the laboratory. The experiment consisted of 12 runs made of permutations of four corner release sites and four beaker locations, with two beakers containing water and two containing the solvent acetone. The autonomous mobile platform was tasked with locating the two beakers of acetone. The sensor had a detection limit of 100 ppb and the confidence of detecting a source within a 1.46 m2 area was p = 0.0005 by ANOVA. The CEMWIP II-mobile platform was able to measure the spatial distribution of VOCs within a laboratory that were associated with open solvent containers.

Entities:  

Keywords:  CEMWIP; VOC; direct reading; exposure assessment; real-time location system; robotic

Year:  2019        PMID: 31532332     DOI: 10.1080/15459624.2019.1657580

Source DB:  PubMed          Journal:  J Occup Environ Hyg        ISSN: 1545-9624            Impact factor:   2.155


  1 in total

1.  A real-time gas monitoring system based on ion mobility spectrometry for workplace environmental measurements.

Authors:  Kazunari Takaya; Masayoshi Hagiwara; Shiro Matoba; Mitutoshi Takaya; Nobuyuki Shibata
Journal:  Ind Health       Date:  2021-11-08       Impact factor: 2.179

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.