Ramin Salouti1,2, Mohammad Reza Khalili1, Mohammad Zamani2, Maryam Ghoreyshi3, M Hossein Nowroozzadeh1.
Abstract
PURPOSE: To assess the changes in biomechanical properties of the cornea after treatment of keratoconus patients with UV-A/riboflavin corneal collagen cross-linking (CXL) using Corvis ST (Oculus, Wetzlar, Germany) and Ocular Response Analyzer (ORA; Reichert Ophthalmic Instruments, Inc., Buffalo, NY, USA) devices.Entities:
Keywords: Corneal biomechanics; Corneal collagen cross-linking; Corvis ST; Ocular response analyzer
Year: 2019 PMID: 31528759 PMCID: PMC6742757 DOI: 10.1016/j.joco.2019.02.002
Source DB: PubMed Journal: J Curr Ophthalmol ISSN: 2452-2325
Comparison of Corvis ST parameters between measurements obtained before and after corneal collagen cross-linking.
| Baseline | After CXL | ||
|---|---|---|---|
| IOP, mmHg | 15.2 ± 2.1 | 15.0 ± 2.0 | 0.702 |
| CCT, μm | 473 ± 29 | 475 ± 27 | 0.688 |
| T1, milliseconds | 7.08 ± 0.28 | 7.05 ± 0.31 | 0.538 |
| L1, mm | 1.80 ± 0.25 | 1.77 ± 0.20 | 0.439 |
| V1, m/s | 0.12 ± 0.02 | 0.12 ± 0.03 | 0.512 |
| T2, milliseconds | 19.87 ± 0.20 | 19.85 ± 0.26 | 0.367 |
| L2, mm | 0.83 ± 0.25 | 1.15 ± 0.57 | 0.001 |
| V2, m/s | −0.81 ± 0.08 | −0.94 ± 0.26 | 0.032 |
| T, milliseconds | 17.05 ± 0.57 | 17.01 ± 0.52 | 0.394 |
| DA, mm | 1.16 ± 0.12 | 1.12 ± 0.09 | 0.062 |
| PD, mm | 4.54 ± 0.94 | 4.34 ± 1.10 | 0.348 |
| R, mm | 5.69 ± 0.64 | 5.54 ± 0.50 | 0.165 |
CXL: Corneal collagen crosslinking; IOP: Intraocular pressure; CCT: Central corneal thickness; T1: Time of applanation 1; L1: Length of applanation 1; V1: Velocity of applanation 1; T2: Time of applanation 2; L2: Length of applanation 2; V2: Velocity of applanation 2; T: Time of highest concavity; DA: Deformation amplitude; PD: Peak distance; R: Radius.
Comparison of Ocular Response Analyzer (ORA) parameters between measurements obtained before and after corneal collagen cross-linking (CXL).
| Baseline | After CXL | ||
|---|---|---|---|
| IOPg, mmHg | 12.4 ± 2.7 | 11.5 ± 2.6 | 0.062 |
| IOPcc, mmHg | 15.1 ± 2.4 | 14.5 ± 3.0 | 0.297 |
| CH, mmHg | 8.57 ± 1.58 | 8.37 ± 1.52 | 0.596 |
| CRF, mmHg | 7.82 ± 1.72 | 7.21 ± 1.05 | 0.036 |
| WS | 4.58 ± 2.55 | 6.12 ± 1.92 | 0.002 |
CXL: Corneal collagen crosslinking; IOP: Intraocular pressure; IOPg: Goldmann-related IOP; IOPcc: Cornea-compensated IOP; CH: Corneal hysteresis; CRF: Corneal resistance factor; WS: Waveform score.
Fig. 1Box and Whisker plot for length of applanation 2 (L2) and velocity of applanation 2 (V2) parameters of the Corvis ST at baseline vs. after corneal collagen cross-linking (CXL).
Fig. 2Box and Whisker plot for corneal resistance factor (CRF) and waveform score (WS) parameters of the Ocular Response Analyzer (ORA) at baseline vs. after corneal collagen cross-linking (CXL).