| Literature DB >> 31510855 |
Stefan M Froschauer1,2, Matthias Holzbauer2, Dietmar Hager3, Richard Schnelzer1, Oskar Kwasny1,2, Dominik Duscher1,4.
Abstract
The present study compares 34 patients with thumb carpometacarpal osteoarthritis (37 thumbs) treated with the Elektra® prosthesis, with 18 patients (18 thumbs) treated with resection-suspension arthroplasty, with an overall mean follow-up period of 13.3 years. Evaluation with disability of arm and shoulder scores, pain via visual analogue scale and range of motion (radial and palmar abduction, and opposition) indicated no significant difference. However, the cohort with a surviving prosthesis showed significantly better subjective grip strength (p = 0.04). Complications occurred in 23 of the 37 thumbs in the prosthesis group compared with two in the resection-suspension arthroplasty patients. Seventeen prostheses required revision. At revision operations, we observed local signs of metallosis in 15 of 17 cases. The patients receiving resection-suspension arthroplasty were more satisfied with their treatment (p = 0.003). Therefore, we cannot recommend the implantation of Elektra® prosthesis and we speculate that the key problem of aseptic cup loosening is a result of the metal-on-metal bearing. Level of evidence: III.Entities:
Keywords: Arthroplasty; Elektra prosthesis; joint replacement; resection-suspension arthroplasty; thumb carpometacarpal osteoarthritis
Mesh:
Year: 2019 PMID: 31510855 PMCID: PMC7232778 DOI: 10.1177/1753193419873230
Source DB: PubMed Journal: J Hand Surg Eur Vol ISSN: 0266-7681
Demographics of patients undergoing different operations.
| Demographics | Elektra prosthesis | Resection-suspension arthroplasty |
|---|---|---|
| Patients/thumbs (number) | 29/32 | 13/13 |
| Right/left thumbs (number) | 13/19 | 7/6 |
| Women/men (number) | 26/3 | 12/1 |
| Mean age at time of operation (range) | 54 years (36–71) | 58 years (51–64) |
Figure 1.Radiograph showing failure of an Elektra prosthesis (aseptic cup loosening).
Figure 2.Kaplan–Meier survival curve for the Elektra prosthesis.
Clinical outcomes for the surviving prosthesis and resection-suspension arthroplasty groups at long-term follow-up.
| Evaluation | DASH score (mean (SD)) | VAS score (median (IQR)) | Radial abduction (mean (SD)) | Palmar abduction (mean (SD)) | Opposition (median (IQR)) | Subjective grip strength (median (IQR)) |
|---|---|---|---|---|---|---|
| Elektra prosthesis | 23 (26) | 0 (40) | 56° (12°) | 50° (17°) | 0 cm (0) | 0 (0) |
| Resection-suspension arthroplasty | 37 (26) | 0 (20) | 51° (8°) | 57° (5°) | 0 cm (1) | −1 (1) |
| 0.08 | 0.62 | 0.32 | 0.08 | 0.65 |
|
Unpaired t-test was used for DASH scores, palmar, and radial abduction. Mann–Whitney U-test was used for VAS scores, opposition, and subjective grip strength. Bold indicates statistically significant p-values.