| Literature DB >> 314895 |
W Matek, P Frühmorgen, B Kaduk, H D Reidenbach, F Bodem, L Demling.
Abstract
In acute experiments in the stomach of dogs electrocoagulation with simultaneous instillation of water (EHT-electrode i.e. electro-hydro-thermo-electrode) proved the most effective electrocoagulation method for controlling standardized bleeding: Optimal visibility at the bleeding site, extension of the therapeutic range as well as the avoidance of charring at the bleeding site are its major advantages. Disturbing adhesion of coagulated material on the electrode is also avoided. The electro-hydro-thermo-electrode is a cheap and portable means of controlling bleeding.Entities:
Mesh:
Substances:
Year: 1979 PMID: 314895 DOI: 10.1055/s-0028-1098361