Load imbalances existed as determined by a sensor after conventional gap balancing with a tensiometer in total knee arthroplasty.

Sang Jun Song1, Hyun Woo Lee1, Kang Il Kim1, Cheol Hee Park2.   

Abstract

PURPOSE: To evaluate intercompartmental load intraoperatively with a sensor after conventional gap balancing with a tensiometer during total knee arthroplasty (TKA).
METHODS: Fifty sensor-assisted TKA procedures were performed prospectively between August and September 2018 with a cruciate-retaining prosthesis. After applying a modified measured resection technique, conventional balancing between resected surfaces was achieved. The equal and rectangular flexion-extension gaps were confirmed using a tensiometer at 90° and 5°-7° (due to posterior tibial slope) of knee flexion. Then, the load distribution was evaluated intraoperatively with a sensor placed on trial implants in the positions of knee flexion (90° flexion) and extension (10° flexion).
RESULTS: The proportion of coronal load imbalance (medial load - lateral load ≥  ± 15 lb) was 56% in extension and 32% in flexion (p = 0.023). The proportion of sagittal load imbalance (extension load - flexion load ≥  ± 15 lb) was 36% in the medial compartment and 4% in the lateral compartment (p < 0.001). An additional procedure for load balancing was performed in 74% of knees.
CONCLUSIONS: Coronal and sagittal load imbalances existed as determined by the sensor even after the achievement of appropriate conventional gap balance. The additional rebalancing procedure was performed for balanced loads in 74% of the knees after conventional balancing. The use of an intraoperative load sensor offers the advantage of direct evaluation of the load on TKA implants. LEVEL OF EVIDENCE: IV.

Keywords:  Gap; Knee; Load; Sensor; Tensiometer; Total knee arthroplasty

Year:  2019        PMID: 31485685     DOI: 10.1007/s00167-019-05699-6

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


  5 in total

Review 1.  Sensor-Assisted Total Knee Arthroplasty: A Narrative Review.

Authors:  Cheol Hee Park; Sang Jun Song
Journal:  Clin Orthop Surg       Date:  2021-02-15

2.  Use of Fulcrum Positioning as a Balancing Tool During Total Knee Arthroplasty on a Robotic Platform.

Authors:  William F Sherman; Christina Freiberger
Journal:  Arthroplast Today       Date:  2021-04-05

3.  Comparison of the contact stress between the sensor and real polyethylene insert in total knee arthroplasty: a finite element analysis.

Authors:  Sang Jun Song; Kang Il Kim; Cheol Hee Park
Journal:  Ann Transl Med       Date:  2020-11

4.  Tibiofemoral dynamic stressed gap laxities correlate with compartment load measurements in robotic arm-assisted total knee arthroplasty.

Authors:  Richard J Allom; Jil A Wood; Darren B Chen; Samuel J MacDessi
Journal:  Bone Jt Open       Date:  2021-11

5.  Intraoperative Load Sensing in Total Knee Arthroplasty Leads to a Functional but Not Clinical Difference: A Comparative, Gait Analysis Evaluation.

Authors:  Michele Giuntoli; Michelangelo Scaglione; Enrico Bonicoli; Nicola Piolanti; Gianmarco Puccioni; Karlos Zepeda; Emanuele Giannini; Stefano Marchetti; Pier Francesco Indelli
Journal:  J Funct Morphol Kinesiol       Date:  2022-02-18
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.