Fusion Imaging to Guide Thoracic Endovascular Aortic Repair (TEVAR): A Randomized Comparison of Two Methods, 2D/3D Versus 3D/3D Image Fusion.

P-A Barral1,2, M A Demasi-Jacquier3,4, L Bal3,4, V Omnes3,4, A Bartoli5,3, P Piquet3,4, A Jacquier5,3, M Gaudry3,4.   

Abstract

PURPOSE: To compare the accuracy of two-dimensional (2D) versus three-dimensional (3D) image fusion for thoracic endovascular aortic repair (TEVAR) image guidance.
MATERIALS AND METHODS: Between December 2016 and March 2018, all eligible patients who underwent TEVAR were prospectively included in a single-center study. Image fusion methods (2D/3D or 3D/3D) were randomly assigned to guide each TEVAR and compared in terms of accuracy, dose area product (DAP), volume of contrast medium injected, fluoroscopy time and procedure time.
RESULTS: Thirty-two patients were prospectively included; 18 underwent 2D/3D and 14 underwent 3D/3D TEVAR. The 3D/3D method allowed more accurate positioning of the aortic mask on top of the fluoroscopic images (proximal landing zone error vector: 1.7 ± 3.3 mm) than was achieved by the 2D/3D method (6.1 ± 6.1 mm; p = 0.03). The 3D/3D image fusion method was associated with significantly lower DAP than the 2D/3D method (50.5 ± 30.1 Gy cm2 for 3D/3D vs. 99.5 ± 79.1 Gy cm2 for 2D/3D; p = 0.03). The volume of contrast medium injected was significantly lower for the 3D/3D method than for the 2D/3D method (50.6 ± 22.9 ml vs. 98.4 ± 47.9 ml; p = 0.002).
CONCLUSION: Higher image fusion accuracy and lower contrast volume and irradiation dose were observed for 3D/3D image fusion than for 2D/3D during TEVAR. LEVEL OF EVIDENCE: II, Randomized trial.

Entities:  

Keywords:  Aorta; Endovascular procedures; Multimodal imaging; Radiation protection; Thoracic

Mesh:

Year:  2019        PMID: 31482337     DOI: 10.1007/s00270-019-02303-9

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  2 in total

1.  Transesophageal Ultrasound Guidance for Endovascular Interventions on the Aorta.

Authors:  Mireya Castro-Verdes; Xun Yuan; Andreas Mitsis; Wei Li; Christoph A Nienaber
Journal:  Aorta (Stamford)       Date:  2022-05-31

2.  Use of three-dimensional-three-dimensional digital subtraction angiography spin fusion protocol for complex aortoiliac endovascular interventions.

Authors:  Edvard Skripochnik; Nicole D'Ambrosio; William E Crawford; Shang A Loh
Journal:  J Vasc Surg Cases Innov Tech       Date:  2021-04-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.