Bhirawich Pholpoke, Techapon Songthawornpong, Woradorn Wattanapanitch.
Abstract
This work presents the design and analysis of a compact low-power motion artifact estimator for reducing the input dynamic range in wearable ECG acquisition systems. The estimator employs a novel mixed-signal architecture that performs adaptive filtering on the electrode impedance information to derive a cancellation signal to suppress the motion artifact at the input of the acquisition system. A detailed noise analysis and optimization strategies are also provided to minimize the estimator's noise referred to the acquisition system's input. Fabricated in a 0.18- μm CMOS process and operating from a 1-V supply, the estimator occupies an active area of 0.11 [Formula: see text] and consumes 2.6-3.2 μW of power depending on the final weights used. The low power consumption and small area thus make the estimator suitable for local placement at each recording channel in multi-channel ECG acquisition systems.Mesh:
Year: 2019 PMID: 31478870 DOI: 10.1109/TBCAS.2019.2937536
Source DB: PubMed Journal: IEEE Trans Biomed Circuits Syst ISSN: 1932-4545 Impact factor: 3.833