Candy-Plug Generation II for False Lumen Occlusion in Chronic Aortic Dissection: Feasibility and Early Results.

Ahmed Eleshra1, Tilo Kölbel1, Nikolaos Tsilimparis1, Giuseppe Panuccio1, Martin Scheerbaum1, E Sebastian Debus1, John Mogensen2, Fiona Rohlffs1.   

Abstract

Purpose: To present the early results of false lumen (FL) occlusion in chronic aortic dissection using the Candy-Plug generation II (CP II), which has a self-closing fabric channel that obviates the need for separate occlusion of its center. Materials and
Methods: Fourteen consecutive patients (mean age 60±11 years; 10 men) with persistent FL backflow and aneurysm formation at the thoracic segment in chronic aortic dissection underwent thoracic endovascular aortic repair (TEVAR) with FL occlusion using the refined CP II. Primary endpoints were technical success (successful deployment) and clinical success (no FL backflow at the CP II level). Secondary endpoints included 30-day mortality and morbidity and aortic remodeling during follow-up.
Results: Technical success was 100%. One patient required additional intraprocedural FL embolization at the CP II level due to persistent FL backflow on final angiography (clinical success 93%), though there was no flow through the CP II center. There were no intraprocedural complications. Immediate complete FL occlusion was achieved in 12 patients; the other 2 required reintervention. One had contrast enhancement in the distal FL proximal to the CP II and was treated with coil embolization. The other patient had persistent type I endoleak at the level of the left subclavian artery (LSA) and underwent left carotid-LSA bypass and proximal stent-graft extension. One patient died due to retrograde type A aortic dissection that was not related to CP II placement. Over a mean 8-month follow-up (range 3-12), 9 patients had computed tomography angiography; 8 patients had evidence of aortic remodeling, while 1 aneurysm sac was stable.
Conclusion: The CP II reduces the number of procedural steps and offers good seal, with minimal morbidity and mortality and a high rate of aortic remodeling.

Entities:  

Keywords:  aneurysm; aortic dissection; aortic remodeling; endoleak; false lumen occlusion; occluder; reintervention; thoracic endovascular aortic repair

Mesh:

Year:  2019        PMID: 31478456     DOI: 10.1177/1526602819871613

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  4 in total

1.  All type B aortic dissections should undergo thoracic endovascular aneurysm repair.

Authors:  Saket Singh; Naiem Nassiri; Prashanth Vallabhajosyula
Journal:  JTCVS Tech       Date:  2021-08-09

2.  A Novel Solution for Distal Dilation of Chronic Dissection After Repair Involving Visceral Branches: The Road Block Strategy.

Authors:  Yiming Li; Zhenjiang Li; Jiaxuan Feng; Rui Feng; Jian Zhou; Zaiping Jing
Journal:  Front Cardiovasc Med       Date:  2022-03-09

Review 3.  Endovascular strategies for post-dissection aortic aneurysm (PDAA).

Authors:  Zhaoxiang Zeng; Yuxi Zhao; Mingwei Wu; Xianhao Bao; Tao Li; Jiaxuan Feng; Rui Feng; Zaiping Jing
Journal:  J Cardiothorac Surg       Date:  2020-10-01       Impact factor: 1.637

Review 4.  Role of Endoluminal Techniques in the Management of Chronic Type B Aortic Dissection.

Authors:  Konstantinos Spanos; Tilo Kölbel
Journal:  Cardiovasc Intervent Radiol       Date:  2020-06-29       Impact factor: 2.740

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.