Literature DB >> 31475356

International external quality assessment for measurements of direct oral anticoagulants: results and recommendations.

Martine J Hollestelle1, Piet Meijer1.   

Abstract

There is limited information regarding the performance of tests for direct oral anticoagulants (DOACs). To generate more knowledge, the accuracy of DOAC tests were evaluated using external quality assessment data from multiple years. This data demonstrated a good correlation for the tests with a small overall interlaboratory variability (10% for dabigatran, rivaroxaban and apixaban and 12% for edoxaban). The greatest differences between the various reagents were observed for rivaroxaban, especially for concentrations below 100 ng/ml. In conclusion, the results show overall reliable DOAC levels with some differences between reagent groups. Important finding: clinical decision criteria could be affected by the choice of reagent.
© 2019 British Society for Haematology and John Wiley & Sons Ltd.

Entities:  

Keywords:  anticoagulants; chromogenic anti-factor IIa and anti-factor Xa assays; diluted thrombin time; quality control

Year:  2019        PMID: 31475356     DOI: 10.1111/bjh.16179

Source DB:  PubMed          Journal:  Br J Haematol        ISSN: 0007-1048            Impact factor:   6.998


  1 in total

1.  Analytical verification and comparison of chromogenic assays for dabigatran, rivaroxaban and apixaban determination on BCSXP and STA Compact Max analysers.

Authors:  Ivana Ćelap; Sandra Margetić; Marija Brčić; Roman Mihić
Journal:  Biochem Med (Zagreb)       Date:  2020-02-15       Impact factor: 2.313

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.