Intrasession Repeatability of the Contact Angle Measured Using the Captive Bubble Method and Agreement Assessed Between Different Analysis Software Programs.

Irene Sanchez1, Francisco Javier Carmona, Sara Gonzalez-Puertas, Angela Valiente, Raul Martin.   

Abstract

OBJECTIVES: The material biocompatibility of hydrogel and silicone hydrogel (SiHy) contact lens (CL) is of paramount importance in CL wear because a decrease in CL wettability reduces wearer comfort and increases wearer dropout. The aim of this study is to report on the repeatability and agreement between two different software programs that measure the contact angle with the captive bubble method in marketed CLs, which will help to translate this information into clinical practice.
METHODS: The contact angle of 23 different CLs was measured with the captive bubble method using 2 software programs: FTÅ200 and ImageJ. Three consecutive measurements were conducted for each CL. Reproducibility, repeatability, and agreement values were calculated according to the British Standards Institute and the International Organization for Standardization.
RESULTS: All methods showed good repeatability values in both CL materials (coefficient of variation <1.51%, Sw <2.26°, intraclass correlation coefficient >0.89, and the range of limits of agreement was between 7.22° and 7.57°). Higher concordance was achieved between the spherical and nonspherical options when using FTÅ200 software than when using ImageJ software. Statistically significant differences (P<0.05) between the 2 software programs were found, and they ranged between 5° and 10°.
CONCLUSIONS: The captive bubble method showed great repeatability in measuring the contact angle in marketed CLs with both software programs (FTÅ200 and ImageJ) assessed in this study. However, differences in the measured contact angles suggest that these techniques are not interchangeable. Therefore, standardization is recommended for contact angle measurement in hydrogel CL materials to facilitate comparisons, to improve clinical use of this information, and to analyze their impact in CL user comfort.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 31469756     DOI: 10.1097/ICL.0000000000000655

Source DB:  PubMed          Journal:  Eye Contact Lens        ISSN: 1542-2321            Impact factor:   2.018


  3 in total

1.  Low-cost, open-source contact angle analyzer using a mobile phone, commercial tripods and 3D printed parts.

Authors:  Won Han; Jaeho Shin; Joong Ho Shin
Journal:  HardwareX       Date:  2022-06-16

2.  Hydrogels as Corneal Stroma Substitutes for In Vitro Evaluation of Drug Ocular Permeation.

Authors:  Susi Burgalassi; Erica Zucchetti; Leonardo Ling; Patrizia Chetoni; Silvia Tampucci; Daniela Monti
Journal:  Pharmaceutics       Date:  2022-04-13       Impact factor: 6.525

3.  Resveratrol-Loaded Hydrogel Contact Lenses with Antioxidant and Antibiofilm Performance.

Authors:  María Vivero-Lopez; Andrea Muras; Diana Silva; Ana Paula Serro; Ana Otero; Angel Concheiro; Carmen Alvarez-Lorenzo
Journal:  Pharmaceutics       Date:  2021-04-11       Impact factor: 6.321

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.