Numerical analysis of aortic hemodynamics under the support of venoarterial extracorporeal membrane oxygenation and intra-aortic balloon pump.

Kaiyun Gu1, Zhiyuan Guan1, Xuanqi Lin2, Yunzhen Feng3, Jieli Feng1, Yujie Yang1, Zhe Zhang4, Yu Chang5, Yunpeng Ling1, Feng Wan3.   

Abstract

BACKGROUND AND
OBJECTIVE: A gap still exists in the hemodynamic effect of intra-aortic balloon pump (IABP), venoarterial extracorporeal membrane oxygenation (VA-ECMO), and VA-ECMO plus IABP on the blood perfusion of the coronary artery, brain, and lower limb; the relation between heart flow and ECMO flow; and the wall stress of vessels.
METHODS: A finite-element model of the aorta, ECMO, and IABP was proposed to calculate the mechanical response via fluid-structure interaction. Heart failure (HF), IABP, ECMO, and ECMO plus IABP were utilized to study the effect of support models.
RESULTS: For the pressure curve, VA-ECMO weakened the dicrotic notch of pressure compared with HF and the pulsatile index (0.494 vs. 0.706 vs. 0.471 vs. 0.613). IABP, ECMO, and ECMO plus IABP increased the perfusion of the coronary, brain, and renal artery compared with HF. However, ECMO and ECMO plus IABP clearly reduced the blood flow of the left arteria femoralis compared to that of the right arteria femoralis (ECMO: 194.04 vs. 730.80 mL/min; ECMO plus IABP: 342.15 vs. 947.22 mL/min). In addition, the flow of ECMO accessed the renal artery more than the left ventricular flow. Greater ventricular flow perfused to the renal artery at a diastolic period for ECMO plus IABP, especially at the time points of 2.192 s and 2.304 s. Compared to the velocity distribution with ECMO, the flow of the right arteria femoralis was increased in the process of IABP-on. According to these four cases, the stress of the vascular wall was increased for ECMO support at the systolic period. The peak wall stress of ECMO is increased by 20% at 1.68 s.
CONCLUSIONS: ECMO plus IABP is more conducive to the blood supply than other cases from the result of numerical simulation. The location of blood intersection was generated in the region of the renal artery, which is estimated carefully.
Copyright © 2019 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Fluid–structure interaction; Hemodynamics; Intra-aortic balloon pump; Venoarterial extracorporeal membrane oxygenation

Mesh:

Year:  2019        PMID: 31465978     DOI: 10.1016/j.cmpb.2019.105041

Source DB:  PubMed          Journal:  Comput Methods Programs Biomed        ISSN: 0169-2607            Impact factor:   5.428


  2 in total

1.  Intra-aortic balloon pump impacts the regional haemodynamics of patients with cardiogenic shock treated with femoro-femoral veno-arterial extracorporeal membrane oxygenation.

Authors:  Bo Xu; Chenglong Li; Tong Cai; Yongchao Cui; Zhongtao Du; Qiushi Fan; Dong Guo; Chunjing Jiang; Zhichen Xing; Meng Xin; Pengcheng Wang; Liangshan Wang; Feng Yang; Ming Jia; Hong Wang; Xiaotong Hou
Journal:  ESC Heart Fail       Date:  2022-05-29

2.  Statistical Analysis of IABP-Surgery Data with the Co-use of Anticoagulants, Pulse of Dorsalis Pedis Artery, D-Dimer Data, and Coagulation Function.

Authors:  Huali Chen; Zhoumin Shen; Yan Zhang; Zhaofen Zheng; Lihua Li; Xinyi Tian; Jianqiang Peng; Xi Peng; Yi Zhou
Journal:  Front Surg       Date:  2022-05-16
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.