Performance of automated urine analyzers using flow cytometric and digital image-based technology in routine urinalysis.

Canan Kucukgergin1, Evin Ademoglu1, Beyhan Omer1, Sema Genc1.   

Abstract

The purpose of this study was to evaluate the analytical performances of Sysmex UF-5000 and Dirui FUS-200 and to compare the results with manual microscopy and between each other. Two hundred fifty urine samples were analyzed for evaluation. Mid-stream specimens were studied sequentially using Dirui FUS-200 and Sysmex UF-5000, and also with manual microscopy within one hour. The physical and chemical components of urinalysis, and sediment results were investigated. The precision results of the FUS-200 and UF-5000 for WBCs, RBCs, and ECs were acceptable. The both analyzers demonstrated good linearity (r > 0.97), with no carry-over. The comparisons of FUS-200 and UF-5000 with manual microscopy for RBCs, WBCs, and ECs on 250 samples exhibited good agreement with little bias (R > 0.780). Only, the moderate agreements were obtained for calcium oxalate for both analyzers (R = 0.512, and 0.648, respectively). The sensitivities of the FUS-200 and UF-5000 were 75.8% and 86.8%, with specificities of 92.3% and 87.8% for WBCs, for RBCs the sensitivities were 91.1%, and 84.4% with specificities of 82.2%, and 89.6% for both analyzers. Kappa values of the UF-5000 were higher than FUS-200 for WBCs, RBCs, ECs, and calcium oxalate. The FUS-200 and UF-5000 urine analyzers, are both accurate, very precise systems and can be safely used in clinical laboratories. However, due to the technological characteristics of the UF-5000 analyzer, its positive impacts on the morphologic recognition and enumeration of RBCs and WBCs should be taken into account, particularly in university hospital laboratories with high patient volumes.

Entities:  

Keywords:  Automated urinalysis; FUS-200; UF-5000; analytical performance; urine microscopy

Mesh:

Year:  2019        PMID: 31460810     DOI: 10.1080/00365513.2019.1658894

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  3 in total

1.  Analysis of Volleyball Video Intelligent Description Technology Based on Computer Memory Network and Attention Mechanism.

Authors:  Zhongzi Zhang
Journal:  Comput Intell Neurosci       Date:  2021-12-28

2.  Consistency analysis of the Sysmex UF-5000 and Atellica UAS 800 urine sedimentation analyzers.

Authors:  Hai Liu; Qing Li; Yiduo Zhang; Dongyue Huang; Fan Yu
Journal:  J Clin Lab Anal       Date:  2022-08-16       Impact factor: 3.124

3.  Sysmex UN2000 detection of protein/creatinine ratio and of renal tubular epithelial cells can be used for screening lupus nephritis.

Authors:  Yabin Chen; Yuan Zhao; Zhishan Zhang; Xiang Cheng; Jie Lin; Jiaming Li; Yibo Wu; Zhen Zhong Lin; Jing Jing
Journal:  BMC Nephrol       Date:  2022-10-05       Impact factor: 2.585

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.