Debarun Sengupta1, Yutao Pei1, Ajay Giri Prakash Kottapalli1,2. 1. Department of Advanced Production Engineering (APE), Engineering and Technology Institute Groningen (ENTEG) , University of Groningen , Groningen 9747 AG , The Netherlands. 2. MIT Sea Grant College Program , Massachusetts Institute of Technology (MIT) , 77 Massachusetts Avenue , NW98-151, Cambridge , Massachusetts 02139 , United States.
Abstract
The growing demand for flexible, ultrasensitive, squeezable, skin-mountable, and wearable sensors tailored to the requirements of personalized health-care monitoring has fueled the necessity to explore novel nanomaterial-polymer composite-based sensors. Herein, we report a sensitive, 3D squeezable graphene-polydimethylsiloxane (PDMS) foam-based piezoresistive sensor realized by infusing multilayered graphene nanoparticles into a sugar-scaffolded porous PDMS foam structure. Static and dynamic compressive strain testing of the resulting piezoresistive foam sensors revealed two linear response regions with an average gauge factor of 2.87-8.77 over a strain range of 0-50%. Furthermore, the dynamic stimulus-response revealed the ability of the sensors to effectively track dynamic pressure up to a frequency of 70 Hz. In addition, the sensors displayed a high stability over 36000 cycles of cyclic compressive loading and 100 cycles of complete human gait motion. The 3D sensing foams were applied to experimentally demonstrate accurate human gait monitoring through both simulated gait models and real-time gait characterization experiments. The real-time gait experiments conducted demonstrate that the information of the pressure profile obtained at three locations in the shoe sole could not only differentiate between different kinds of human gaits including walking and running but also identify possible fall conditions. This work also demonstrates the capability of the sensors to differentiate between foot anatomies, such as a flat foot (low central arch) and a medium arch foot, which is biomechanically more efficient. Furthermore, the sensors were able to sense various basic joint movement responses demonstrating their suitability for personalized health-care applications.
The growing demand for flexible, ultrasensitive, squeezable, skin-mountable, and wearable sensors tailored to the requirements of personalized health-care monitoring has fueled the necessity to explore novel nanomaterial-polymer composite-based sensors. Herein, we report a sensitive, 3D squeezable graphene-polydimethylsiloxane (PDMS) foam-based piezoresistive sensor realized by infusing multilayered graphene nanoparticles into a sugar-scaffolded porous PDMS foam structure. Static and dynamic compressive strain testing of the resulting piezoresistive foam sensors revealed two linear response regions with an average gauge factor of 2.87-8.77 over a strain range of 0-50%. Furthermore, the dynamic stimulus-response revealed the ability of the sensors to effectively track dynamic pressure up to a frequency of 70 Hz. In addition, the sensors displayed a high stability over 36000 cycles of cyclic compressive loading and 100 cycles of complete human gait motion. The 3D sensing foams were applied to experimentally demonstrate accurate human gait monitoring through both simulated gait models and real-time gait characterization experiments. The real-time gait experiments conducted demonstrate that the information of the pressure profile obtained at three locations in the shoe sole could not only differentiate between different kinds of human gaits including walking and running but also identify possible fall conditions. This work also demonstrates the capability of the sensors to differentiate between foot anatomies, such as a flat foot (low central arch) and a medium arch foot, which is biomechanically more efficient. Furthermore, the sensors were able to sense various basic joint movement responses demonstrating their suitability for personalized health-care applications.
The growing demand for
flexible, ultrasensitive wearable sensors
for myoelectric prosthesis, soft robotics, and personalized health
monitoring applications has been the main driving factor behind the
rapid advancement in soft and flexible material processing technologies.
In particular, human motion monitoring devices are emerging to be
the most sought-after wearable devices as they can potentially provide
a host of valuable information regarding the health and well-being
of an individual. For example, gait monitoring in individuals suffering
from Parkinson’s disease, stroke, multiple sclerosis, and other
neurological conditions can provide valuable information regarding
the progression of the diseases, and hence, continuous monitoring
of gait characteristics can enable early diagnosis, subsequently enabling
personalized treatment plans for patients.[1] Accurate monitoring of human gait and other human motion parameters
like joint and limb movements necessitates the availability of squeezable,
skin-mountable, low-cost, and durable sensors.Traditionally,
piezoelectric, piezoresistive, and capacitive transduction
mechanisms have mostly been exploited in the development of flexible
sensors to convert strain stimuli into electrical signals. Inorganic
piezoelectric materials such as lead zirconate titanate, zinc oxide
(ZnO), barium titanate (BaTiO3),[2−7] and soft piezoelectric polymers like polyvinylidene fluoride (PVDF)
and polyvinylidene fluoride-trifluoroethylene (PVDF-TrFE)[8−12] have been explored extensively for developing various flexible sensors
in the past. On the other hand, though piezoresistive microelectromechanical
system sensors using metallic- and semiconductor-based strain gauges
have been quite popular for strain sensing applications owing to their
well-established fabrication processes and large measurement range,
their application as wearable sensors is limited due to their high
stiffness and low stretchability.[13,14] Flexible and
squeezable sensors utilizing the piezoresistive property of nanomaterial-elastomer
composites are relatively new, and researchers across the globe have
been exploring various combinations of novel nanomaterials and suitable
elastomers for developing a new generation of innovative, flexible
piezoresistive sensors.[15−21] Polymer materials like polydimethylsiloxane (PDMS), ecoflex, polyimide
(PI), rubber, and polyurethane (PU) have been commonly used as flexible
polymer substrates due to their superior flexibility compressibility
and excellent responsiveness to torsion, tension, and compression.[16,18,22−24] To make the
sensors low cost, renewable, and biodegradable, a recent work reported
the use of a printing paper substrate as a novel alternative to traditional
elastomers.[25] For the conductive nanomaterials,
silver nanowires (AgNWs) and various types of carbon-based materials
like carbon nanotubes (CNTs), carbon nanofibers (CNFs), carbon blacks,
and graphene have been explored by researchers.[18,19,26−32] Of all the conductive carbon-based nanomaterials, recently, graphene
has been exploited the most for developing nanomaterial-polymer composite-based
piezoresistive sensors mainly because of its excellent conductivity,
stiffness, and elastic properties.[16,22,33] The most common method of developing graphene-based
piezoresistive sensors has been the use of nanoporous graphene foams
synthesized by chemical vapor deposition. Due to the fragility of
freestanding 3D graphene foams, they are infiltrated with elastomers
like PDMS to enhance their mechanical properties like elasticity and
durability. Pang et al. proposed a novel method involving infiltration
of PDMS in a graphene-coated nickel foam template and subsequent etching
to preserve the graphene nanoporous structure with the PDMS scaffold
for developing a highly sensitive piezoresistive sensor.[16] Recently, researchers have also reported graphene-based
fiber sensors for various strain monitoring applications.[31,34] A recently published review article presents a detailed overview
of the developments in the field of flexible polymer-based strain
sensors and discusses the recent developments in the field of electrically
conductive polymer composites.[35]Most of the graphene-elastomer foam/spongy materials reported so
far have either employed fragile freestanding foamy graphene structures
or involved sophisticated multistep fabrication methods involving
polymer infiltration in a graphene-coated template and subsequent
etching. In addition, most of these works focused on the synthesis
of the spongy nanomaterials; however, their applications as sensing
materials to develop 3D squeezable sensors were not investigated.[36−38] In this work, we propose a facile process of developing an ultralightweight
and highly squeezable 3D microporous graphene-PDMS foam-based sensor.
Sugar cubes were used as templates for developing microporous PDMS
foams followed by dip coating of the foams in conductive multilayered
graphene (MLG) suspension to infiltrate them with graphene nanoflakes.
The porous graphene-infiltrated foams were studied employing a scanning
electron microscope (SEM) to understand its strain-induced resistance
modulation mechanism, and the conductive domain disconnection mechanism
was invoked to explain its piezoresistive property. The density of
the graphene-PDMS foam sensors was calculated to be 0.305 g cm–3 owing to the porous structure. The response of the
graphene-PDMS sensors developed with the proposed method was characterized
for static and dynamic pressure stimuli. From the static and dynamic
compressive strain tests, the response of the sensor was found to
have two linear regions with an average gauge factor lying in the
range of 2.87–8.77. Accelerated lifetime tests were conducted
on the spongy sensor through cyclic compressive loading involving
36000 load cycles to demonstrate its overall reliability and durability.
The responses of the sensors to dynamic loading were characterized
to observe their sensing performances at high-frequency strain loading.
Finally, the application of these sensors in monitoring both simulated
and real-time human gaits and other body motion parameters was validated
through experiments that demonstrate the broad applicability of such
sensors in various applications including personalized health monitoring,
soft robotics, myoelectric prosthesis, and other wearable devices.
Three identical sensors were assembled on a soft shoe sole and used
in synchronization to differentiate between the pressure profiles
under a low arch/flat foot and a medium arch foot. The sensor assembly
was also used for demonstrating the capability for real-time gait
characteristic acquisition and differentiation between different kinds
of human movements, including walking, running, periodic leaning,
and standing. The simple method of sensor development demonstrated
in this work will guide the development of a future generation of
3D squeezable and highly sensitive pressure/strain sensors suitable
for various high-performance, wearable, and flexible devices.
Results and Discussion
3D Squeezable Graphene-PDMS
Foam Sensor
Sensor Fabrication and
Morphological Study
The process steps involved in fabrication
of the graphene-PDMS
squeezable strain sensor are schematically illustrated in Figure a. The experimental
details of the fabrication process are described in the Experimental Methods section. The electrical connections for
acquiring electrical outputs from the sensor were made by smearing
and subsequently curing a thin layer of conductive silver epoxy on
the two sides of the graphene-PDMS foam as shown in Figure b. The optical images shown
in Figure c,d demonstrate
the compressibility and ultralightweight of the developed sensor,
respectively. Furthermore, the squeezability of the sensor was demonstrated
by holding it between two fingers and applying a series of squeeze–release
cycles (shown in Video S1).
Figure 1
Fabrication of the squeezable
graphene-PDMS foam sensor: (a) schematic
representation of infiltrating PDMS foam with multilayered graphene
nanoflakes (MLG); (b) schematic representation of a single PDMS-MLG
foam sensor with electrical contacts; (c) squeezability and (d) ultralight
nature of the graphene-PDMS foam sensor.
Fabrication of the squeezable
graphene-PDMS foam sensor: (a) schematic
representation of infiltrating PDMS foam with multilayered graphene
nanoflakes (MLG); (b) schematic representation of a single PDMS-MLG
foam sensor with electrical contacts; (c) squeezability and (d) ultralight
nature of the graphene-PDMS foam sensor.Multiple dip coating and drying cycles of the PDMS sponge in homogeneous
graphene/N,N-dimethylformamide (DMF)
suspension led to the attachment of multilayered graphene nanoflakes
(MLG) in the inner pore walls of the microporous PDMS substrate. Figure a,b shows the SEM
micrographs of the unloaded PDMS foam. An average pore diameter of
386 μm was observed for the developed PDMS foam (averaged over
eight measurements on different sponges). Figure c,d shows the SEM micrographs after loading
the PDMS with MLG. As the SEM micrographs clearly depict, the graphene
nanoflakes penetrated into the porous structure of the PDMS foam and
attached themselves onto the inner walls of the microporous structure.
Figure 2
SEM micrographs
of the PDMS foam before and after MLG loading:
(a,b) porous structure of the unloaded PDMS foam at different magnifications;
(c,d) MLG-loaded PDMS foam at two different magnifications with the
pore walls covered with MLG nanoflakes forming a nanomaterial percolation
network.
SEM micrographs
of the PDMS foam before and after MLG loading:
(a,b) porous structure of the unloaded PDMS foam at different magnifications;
(c,d) MLG-loaded PDMS foam at two different magnifications with the
pore walls covered with MLG nanoflakes forming a nanomaterial percolation
network.The strain-responsive resistance
change mechanism in the graphene-PDMS
foam could arise from the conductive domain disconnection mechanism,
which was also reported in the past for other types of thin films
made of nanomaterials.[19,24,39−41] Within the MLG nanomaterial flake network, electrons
pass through the overlapping network of conductive MLG flakes. Application
of external force/stress causes a change in the overlapping area between
the conductive MLG flakes, thus leading to a change in resistance
as schematically explained in Figure a. Also, electrons can tunnel across a thin polymer
barrier separating two adjacent nanomaterial domains, thus forming
quantum tunneling junctions. The tunneling resistance between two
adjacent graphene nanoflakes separated by a polymer layer can be predicted
using Simmon’s tunneling resistance theory.[42] In the past, researchers have reported a tunneling cutoff
distance of 2–3 nm between two parallel graphene sheets separated
by polymer insulation.[43,44] Upon application of pressure,
a pore wall may get compressed enough so that the effective distance
between two graphene nanoflakes adhering to the opposite sides of
the wall may reduce to 2 nm or less in which case electrons will be
able to tunnel across the PDMS wall barrier. Figure b schematically explains the stress-induced
tunneling resistance modulation. Given the relatively large size of
the pores, it can be safely assumed that the strain-responsive piezoresistivity
in the graphene-PDMS foams reported in this paper originates mostly
due to the conductive domain disconnection mechanism (while the stress-induced
tunneling resistance modulation mechanism plays a minimal to no role
at all) wherein an external force/stress causes the graphene nanoflakes
to slide against each other, thus leading to a large change in overall
resistance of the graphene-PDMS foam sensor.
Figure 3
Schematic diagram explaining
the possible strain-induced resistance
modulation mechanisms in graphene-PDMS foam sensor: (a) schematic
explaining the conductive domain disconnection mechanism explaining
strain-induced resistance modulation observed in the sensor; (b) schematic
representation of stress-induced tunneling resistance modulation.
Schematic diagram explaining
the possible strain-induced resistance
modulation mechanisms in graphene-PDMS foam sensor: (a) schematic
explaining the conductive domain disconnection mechanism explaining
strain-induced resistance modulation observed in the sensor; (b) schematic
representation of stress-induced tunneling resistance modulation.
Strain Sensor Characterization
The graphene-PDMS foam sensors were characterized under various
strain
loading conditions including static and dynamic loading in order to
demonstrate its applications for flexible and wearable sensors. Figure a shows the schematic
representation of the experimental setup used for the characterization
of the sensors (details provided in the Experimental
Methods section). An initial precompression of 1% was applied
to avoid problems related to initial sliding/settling of graphene
nanoflakes, which are observed in similar types of squeezable sensors
developed in the past.[33] A compressive
strain was applied in steps of 0.5% all the way up to 9.5%. After
attaining a peak compressive strain of 9.5%, the strain was released
in steps of 0.5% to return to the starting position. The experiment
was repeated three times, and no noticeable delay was observed between
the piston extension and the sensor response throughout the duration
of the experiment. With increasing compressive strain, the resistance
of the sensor was observed to decrease linearly for strains up to
9.5%. Figure b shows
the plot of the modulus of normalized resistance change calculated
from the data acquired from the Wheatstone bridge circuit versus the
compressive strain. The output of the sensor was observed to increase
linearly with increasing strain for compressive strains up to 9.5%.
The data was treated with a linear regression fit to estimate the
compressive gauge factor () of the graphene-PDMS
foam. The compressive
gauge factor of the sensor was determined to be 8.77 from the slope
of the linear regression. To assess the strain sensing performance
of the graphene-PDMS foam sensor for larger strains, the sensor was
subjected to five different maximum compressive cyclic strains (10,
20, 30, 40, and 50%) at a constant frequency of 5 Hz using a minishaker
setup as depicted in the schematic in Figure c (details of the experimental setup are
provided in the Experimental Methods section).
The minishaker was driven with a square-wave stimulus at a constant
frequency of 5 Hz, which resulted in cyclic compressive stains in
the sponge. The sensor outputs were recorded for at least 50 cycles
at each of the aforementioned maximum strain levels, and the normalized
resistance changes were calculated and plotted for the individual
compressive strain cycles as shown in Figure d. Figure e shows the superimposed plots of the normalized resistance
change demonstrated by the sensor in response to the five maximum
compressive strain cycles (appropriate band-pass filter was used to
eliminate the 50 Hz power supply interference). The data from the
cyclic compressive strain characterization experiment were analyzed,
and the mean normalized resistance change values were determined individually
for each of the five different maximum compressive strain levels.
Furthermore, the gauge factor of the sponge was determined individually
for each of the aforementioned compressive strain values and plotted
in the form of a bar graph, as shown in Figure f. From the plots in Figure b,f, two distinct regions of operations (of
the graphene-PDMS foam sensor) can be identified. The sensor demonstrated
a reasonably linear response for strain levels up to 9.5%. A sharp
decrease in the gauge factor was observed for strains exceeding 10%.
The decrease in the gauge factor at higher strain rates can be attributed
to the fact that, at higher strain values, the internanoparticle distances
are continuously bridged, and hence, the sensor reaches near network
saturation.[45] Thus, saturation in the percentage
decrease in resistance for higher compressive strain levels is observed,
which is clearly reflected by the reduced gauge factor values (at
higher strain levels exceeding 10%). Similar response characteristics
have been observed for similar nanoparticle-elastomer composite foam-based
sensors in the past.[16,45]Table compares the gauge factor of our graphene-PDMS
foam sensor with some other sensors reported by various researchers
in the past.
Figure 4
Graphene-PDMS foam sensor characterization: (a) schematic
representation
of the setup used for conducting the piezoresistivity characterization
experiments; (b) plot of normalized resistance change in the graphene-PDMS
foam sensor versus applied compressive strain up to 9.5%; (c) schematic
representation of the setup used for conducting the piezoresistivity
characterization experiments for larger compressive strains in the
range of about 10–50%; (d) plot showing the sensor response
in terms of normalized resistance change when subjected to five different
compressive loading at different strains (between 10 and 50%); (e)
superimposition plot showing the normalized resistance change of the
sensor for five different compressive strains between 10 and 50%;
(f) bar chart showing the calculated gauge factor for the graphene-PDMS
foam sensor for the five different compressive strains between 10
and 50%.
Table 1
Summarizing the Gauge
Factors of Various
Flexible Strain Sensors Reported in the Past
material
gauge
factor
linearity
AgNWs-PDMS[19]
2–14
linear up to 40%
CNTs-Ecoflex[18]
1–2.5
linear
aligned SWCNTs-PDMS[26]
0.82
two linear
regions
carbon black-PDMS[27]
1.8–5.5
two linear regions
carbon black-EcoFlex[46]
3.8
nonlinear
single CNF strain sensor[47]
1.96–2.55
linear
GPN-PDMS[16]
2.6–8.5
two linear regions
graphene-rubber[24]
35
linear and
exponential regions
graphene ink-on-PDMS[48]
37
linear
graphene-PDMS foam (this work)
2.87–8.77
two linear regions
Graphene-PDMS foam sensor characterization: (a) schematic
representation
of the setup used for conducting the piezoresistivity characterization
experiments; (b) plot of normalized resistance change in the graphene-PDMS
foam sensor versus applied compressive strain up to 9.5%; (c) schematic
representation of the setup used for conducting the piezoresistivity
characterization experiments for larger compressive strains in the
range of about 10–50%; (d) plot showing the sensor response
in terms of normalized resistance change when subjected to five different
compressive loading at different strains (between 10 and 50%); (e)
superimposition plot showing the normalized resistance change of the
sensor for five different compressive strains between 10 and 50%;
(f) bar chart showing the calculated gauge factor for the graphene-PDMS
foam sensor for the five different compressive strains between 10
and 50%.
Reliability Test and
Dynamic Response Characterization
To demonstrate the long-time
reliability of the graphene-PDMS foam
sensor, an accelerated lifetime testing was conducted on the sensor
by subjecting it to a series of 36000 cyclic compressive loading and
unloading at 5% compressive strain using the same experimental setup
previously shown in Figurec. The minishaker was driven at a constant frequency of 10
Hz, and the power amplifier driving the setup was set such that the
compressive strain generated was approximately 5%. The sensor response
was acquired from the Wheatstone bridge circuit, and an appropriate
band-pass filter was applied to eliminate the 50 Hz power supply interference. Figure a shows the normalized
resistance change plots acquired from the reliability tests. A zoomed-in
plot placed on the right-hand side of the main plot shows the consistency
of the sensor response cycles for the applied cyclic compressive strains.
Figure 5
Cyclic
compressive loading and dynamic strain sensing characterization
of the graphene-PDMS foam sensor: (a) response of the sensor to cyclic
loading and unloading at 5% compressive strain. The figure on the
right shows the zoomed-in version of the plot; (b) plot showing the
sensor response at 10 Hz in time domain; zoomed in plot on the right
showing the sensor response at 10 Hz in the interval of 1.3–2.3
s; (c) FFT amplitude plots showing the sensor responses at 35 and
70 Hz.
Cyclic
compressive loading and dynamic strain sensing characterization
of the graphene-PDMS foam sensor: (a) response of the sensor to cyclic
loading and unloading at 5% compressive strain. The figure on the
right shows the zoomed-in version of the plot; (b) plot showing the
sensor response at 10 Hz in time domain; zoomed in plot on the right
showing the sensor response at 10 Hz in the interval of 1.3–2.3
s; (c) FFT amplitude plots showing the sensor responses at 35 and
70 Hz.To further evaluate the strain
sensing performance of the sensor
under dynamic loading conditions, the graphene-PDMS foam sensor was
subjected to compressive strains at three different frequencies using
the same minishaker setup described previously. The minishaker was
driven with square-wave stimuli at three different frequencies (10,
35, and 70 Hz), which caused compression of the sponge at those frequencies. Figure b shows the as-acquired
sensor response for the oscillatory test conducted at 10 Hz. For the
35 and 70 Hz stimuli, the sensor responses were acquired and treated
with appropriate low-pass filters in order to eliminate the 50 Hz
power supply interference. Fast Fourier transform (FFT) was carried
out on the individual responses to determine the average amplitude
of the sensor response as shown in Figure c. The sensor amplitude responses were observed
to have increased with the applied stimulus frequency.
Sensor Applications
Application of the Graphene-PDMS
Foam for
Human Motion Monitoring
Continuous monitoring of gait characteristics
can enable early diagnosis of diseases like stroke, multiple sclerosis,
and Parkinson’s disease, thus enabling personalized treatment
plans for patients.[1] To demonstrate the
applicability of the graphene-PDMS foam sensors in gait monitoring,
a gait simulation response experiment was conducted using the Instron
5940 UTS (details of the experimental setup are provided in the Experimental Methods section). The pressure behavior
under the heel of a walking person was appropriately mimicked employing
simulated gait models applied to the movable piston of the test system.
The pressure pattern under the heel of a walking individual comprises
a gradual ramping up to the maximum pressure (body weight divided
by the area of the heel pad) followed by a partial pressure release
and finally ramping down to a complete pressure release when the heel
is lifted off the ground.[49] Due to limitations
of the test setup used, the force ramp up and ramp down rate was slow
(20 mm/min movement of piston) due to which each gait cycle lasted
30 s unlike in real human being where each gait cycle lasts 1.08 ±
0.11 s.[50] The experiment was carried out
for 45 gait cycle repetitions to demonstrate the consistency in sensor
response. Figure a
shows the sensor response for the gait simulation experiment. The
zoomed-in version of Figure a (right) shows the sensor response for four complete gait
cycles. The schematic diagrams in the figure inset explain the sensor
response by comparing it to the heel movement. Overall, a good consistency
was observed in the sensor response throughout the gait simulation
experiment.
Figure 6
Application of the graphene-PDMS foam sensor for human gait monitoring:
(a) response plot of the sensor to simulated gait; plot on the right
shows the zoomed-in version of the sensor response over four gait
cycles with schematics explaining heel positioning; (b) schematic
representation of the soft shoe sole sensor assembly (SSA); (c) plot
showing the sensor responses from the toe ball and heel regions while
walking; (d) plot showing the sensor responses from the toe ball,
foot arch, and heel regions while running; (e) plot showing sensor
responses from the toe ball, foot arch, and heel regions while leaning
forward and backward in a periodic fashion.
Application of the graphene-PDMS foam sensor for human gait monitoring:
(a) response plot of the sensor to simulated gait; plot on the right
shows the zoomed-in version of the sensor response over four gait
cycles with schematics explaining heel positioning; (b) schematic
representation of the soft shoe sole sensor assembly (SSA); (c) plot
showing the sensor responses from the toe ball and heel regions while
walking; (d) plot showing the sensor responses from the toe ball,
foot arch, and heel regions while running; (e) plot showing sensor
responses from the toe ball, foot arch, and heel regions while leaning
forward and backward in a periodic fashion.To demonstrate the capability of the sensor for real-time gait
and foot pressure monitoring, three identical graphene-PDMS sensors
were attached and secured on a soft flat shoe sole with the intention
of acquiring the sensor response from three distinct pressure points
(toe ball, foot arch, and heel) of the right foot as shown in Figure b. The shoe sole-sensor
assembly (SSA) was placed inside a shoe and worn by a person with
a medium arch foot. Figure c shows the response of the sensor (acquired in real time)
while the person was walking slowly. For this work, the sensor responses
from the toe ball and the heel regions are shown as these are the
two most intense pressure regions in the medium arch biomechanically
efficient foot. The phase lag between sensors from the toe ball and
heel region demonstrates the walking behavior of the person. While
walking, when the heel is placed down, the pressure increases to a
maximum value followed by subsequent relaxation while the whole foot
is placed down on the floor. At the point where the foot is completely
down on the floor, the two sensor response curves intersect each other
indicating equal pressure distribution. As the heel is lifted slowly
while placing the toe ball down on the floor, the pressure of the
toe ball increases up to a maximum and the pressure of the heel decreases
to a minimum value. This behavior is repeated throughout the entire
duration of walking, as shown by the sensor response plot in Figure c. The SSA was also
applied for real-time running pressure variation monitoring. As shown
by the plot in Figure d, the pressure response is very different from walking. In the case
of running or jogging, most of the impact is absorbed by the toe ball
followed by the middle arch, which is reflected clearly by the sensor
response plot. Furthermore, the phase lag characteristics differ significantly
from normal walking. To demonstrate the capability of the SSA in detecting
pressure variation, the person wearing it leaned forward and backward
in a periodic fashion leading to a periodic pressure distribution
variation between the toe ball and the heel. As expected, the sensor
response plot in Figure e clearly shows the phase lag between the toe ball and heel sensor
pressure response. Interestingly, the pressure variation from the
foot arch (middle sensor) is relatively less than those in the other
two regions, which can be attributed to the fact that the maximum
share of the weight of a human body is borne by the toe ball and heel,
which leads to larger pressure concentration in those two regions
in comparison to the foot arch region.The fact that the maximum
share of the weight of a human body is
borne by the toe ball and heel (which leads to larger pressure concentration
in those two regions in comparison to the foot arch region) was utilized
to differentiate between a low arch (flat) foot and a medium arch
foot using the SSA. Figure a compares a low arch/flat foot with a medium arch foot. Due
to the difference in the anatomies of the two feet types, their pressure
profiles are distinct and different. As seen in the figure, low arch-type
foot typically has a foot arch sitting low to the ground, and hence,
it has significantly more pressure concentration in the middle foot
arch region in comparison to the medium arch foot. To demonstrate
the capability of the SSA in distinguishing between the two different
foot types, the setup was worn by a person with a flat foot, and the
pressure response was recorded while the foot was placed down. The
experiment was repeated on a person with a medium arch foot. The plots
in Figure b show the
SSA responses acquired from the persons with the two different foot
types. As expected, the SSA response from the person with low arch/flat
foot indicates a more even pressure distribution between the three
pressure regions. Meanwhile, the SSA response acquired from the person
with medium arch foot shows a more skewed pressure distribution with
the toe ball and heel sharing the maximum share of the load in comparison
to the middle arch region. The experiments demonstrate the capability
of the SSA to distinguish between the different feet anatomies.
Figure 7
Application
of the graphene-PDMS foam sensor for distinguishing
between foot anatomies and human motion monitoring: (a) images comparing
a low arch/flat foot with a medium arch biomechanically efficient
foot; (b) plot comparing the SSA pressure responses acquired from
the two different foot types. The pressure distribution in the case
of the low arch foot is more even than that in the medium arch foot;
(c) response of the sensor to index finger flick; (d) response of
the sensor to wrist flick.
Application
of the graphene-PDMS foam sensor for distinguishing
between foot anatomies and human motion monitoring: (a) images comparing
a low arch/flat foot with a medium arch biomechanically efficient
foot; (b) plot comparing the SSA pressure responses acquired from
the two different foot types. The pressure distribution in the case
of the low arch foot is more even than that in the medium arch foot;
(c) response of the sensor to index finger flick; (d) response of
the sensor to wrist flick.Furthermore, to demonstrate the applicability of the sensor in
sensing finger and wrist joint movements, the sensor was secured on
a wearable nitrile glove, which was then worn to demonstrate working
on the sensor. Five cycles of finger and wrist flicking were carried
out, and the output from the balanced Wheatstone bridge circuit to
which the sensor was connected was recorded as shown in Figure c,d. The experiments conducted
demonstrate the feasibility of using such sensors for developing wearable
biomedical devices for health monitoring applications.
Conclusions
In conclusion, this work presented a facile
method for developing
a graphene-PDMS foam-based ultralightweight (having a density of 0.31
g cm–3), squeezable, linear, and highly sensitive
sensor. The sensor demonstrated in this work utilized a microporous
PDMS substrate with graphene nanoflakes attached to its inner pore
walls forming an MLG percolation network, which responds to pressure/strain
by virtue of the conductive domain disconnection mechanism. To support
the theory of conductive domain disconnection mechanism (which explains
the strain-responsive resistance change property demonstrated by the
sensor), SEM microscopic studies were conducted, which revealed the
attachments of graphene nanoflakes on the inner pore walls of the
PDMS foam substrate, thus backing up our hypothesis. The sensor was
subjected to a series of static and dynamic strain stimuli response
tests to evaluate its sensing performance and repeatability. The sensor
responses were found to be linear, and the average gauge factor was
determined to be 8.77 for compressive strains up to 9.5%. For compressive
strains exceeding 10%, the gauge factor was found to vary between
2.87 and 8.77 (in the strain range of 10–50%). To demonstrate
the feasibility of applying the sensor for various wearable devices
and personalized health monitoring applications, both simulated and
real-time gait responses and other human monitoring experiments were
conducted. A soft shoe sole sensor assembly was fabricated and demonstrated
to identify various gait characteristics, including walking, running,
periodic leaning, and standing. The sensor assembly was also found
to be capable of differentiating the foot types based on their middle
arch architecture. The simple method for developing highly sensitive,
lightweight, and squeezable piezoresistive sensors demonstrated in
this work will inspire a future generation of inexpensive and highly
efficient pressure and strain sensors suitable for human motion detection
and personalized health monitoring applications.
Experimental Methods
Preparation
of PDMS Foam Base
Sundale
extra-large sugar cubes were used as templates for fabricating PDMS
foams. PDMS (Sylgard 184) acquired from Dow Corning was mixed in a
ratio of 10:1 (base curing) to prepare the PDMS liquid and then degassed
in a vacuum desiccator for 90 min to remove any unwanted trapped air
bubbles. The sugar cubes were then dipped in the degassed liquid PDMS
and placed in the vacuum desiccator for 90 min to let the liquid PDMS
seep into the pores of the sugar cube template by capillary action.
At the end of the desiccation procedure, the PDMS-loaded sugar cubes
were cleaned to remove excess PDMS from their surfaces to avoid the
formation of a skin layer, which could hinder the dissolution of sugar
to release the PDMS foam. The cleaned PDMS-loaded sugar cubes were
placed in an oven at 120 °C for 1 h to cure the PDMS. The cured
PDMS-loaded sugar cubes were then placed in a sonicator bath at 40
°C for 1 h to dissolve the sugar template and release the PDMS
foam structure.
Loading of the PDMS Foams
with Graphene Nanoflakes
To fabricate piezoresistive graphene-PDMS
foam sensors, the PDMS
foams were loaded with graphene nanoflakes by immersing them in a
sonicated homogeneous graphene suspension solution prepared by mixing
200 mg of 1.6 nm thick graphene nanoflakes (AO-1) acquired from Graphene
Supermarket in 100 mL of N,N-dimethylformamide. Figure a shows the process
steps involved in the loading procedure. In total, six dip coating
cycles were conducted where each dip coating cycle comprised immersion
of the PDMS foams in the graphene suspension followed by air-drying
in an oven at 60 °C for 1 h.
Preparation
of the Graphene-PDMS Foam Sensor
Epotek H20E conductive epoxy
was used for making the electrical
contacts for acquiring signal from the sensor. The parts A and B of
the epoxy kit were mixed in a ratio of 1:1, smeared on two opposite
faces of the foam, and subsequently cured at 120 °C for 20 min. Figure b shows the schematic
representation of the graphene-PDMS foam sensor. Thin multistrand
electrical wires were used for connecting the sensors to an appropriate
Wheatstone bridge circuit.
Morphological Study
A Philips FEI
XL30 environmental scanning electron microscope was employed to study
the morphological properties of both the unloaded and graphene-loaded
PDMS foams. Samples having dimensions of 2 cm × 2 cm × 2
cm were sputtered with gold and placed on an appropriate SEM stub.
An acceleration voltage of 15 kV and a spot size of 3.0 were used
while maintaining a working distance of 7 mm for carrying out the
imaging studies.
To characterize the performance
of the graphene-PDMS foam sensor
for pressure/strain sensing applications, Instron 5940 Universal Testing
Systems with modified sample holders (appropriate for compressive
tests) having a maximum force capacity of 2 kN was employed. To determine
the compressive gauge factor of the sensor, a program was developed
in the accompanying BlueHill software whereby the movable piston of
the instrument was programed to move in steps of 50 μm to achieve
a total compression of 0.95 mm in the foam sensor followed by stepwise
relaxation (step size of 50 μm) all the way back to the original
starting position. A compression step of 50 μm led to a compressive
strain of 0.5% in the cubic sponge having an edge length 10 mm (after
applying a 1% precompression). The copper pistons of the instrument
doubled as the contact electrodes for acquiring electrical signals
as shown in Figure a. The foam was connected to an appropriate Wheatstone bridge circuit,
which converted the resistance change in the sponge to a voltage signal
output. The signal output from the balanced Wheatstone bridge circuit
was continuously acquired using a National Instruments data acquisition
system (DAQ, NI USB-6009) and logged using National Instruments Signal
express software as schematically shown in Figure a. The experiment was repeated thrice to
observe consistency in the measurements.To assess the strain
sensing performance of the graphene-PDMS foam sensor for larger strain
percentages, the sensor was subjected to five different maximum compressive
cyclic strains (10, 20, 30, 40, and 50%) at a constant frequency of
5 Hz using a minishaker setup as schematically presented in Figurec. An appropriate
Wheatstone bridge circuit was designed to which the strain sensor
was connected as an arm. The sensor was placed and secured between
two glass plates, and a series of cyclic compressive stimuli were
applied to the sensor employing a 120 mm long steel rod having a diameter
of 2 mm connected to a Brüel & Kjær permanent magnet
minishaker as shown in Figure c. The minishaker was driven at 5 Hz frequency using a Rigol
function generator connected to a Brüel & Kjær (model
number 2718) power amplifier, which generated a cyclic compressive
strain in the graphene-PDMS sensor. The experiments were repeated
for five different maximum strain percentages, namely, 10, 20, 30,
40, and 50%. The voltage output from the Wheatstone bridge circuit
was acquired by employing the National Instruments data acquisition
system (DAQ, NI USB-6009) and recorded using National Instruments
Signal express software at a sampling rate of 1 kHz.The reliability
of the sensor was studied by subjecting it to a
series of 36000 cyclic compressive loading and unloading at 5% compressive
strain using the same experimental setup previously shown in Figurec. The minishaker
was driven at a constant frequency of 10 Hz, and the power amplifier
driving the setup was set such that the compressive strain generated
was approximately 5%. Like in the previous case, data was logged continuously
by connecting the sensor to a Wheatstone bridge circuit to generate
a voltage signal output and recording by employing the same data acquisition
setup described previously.
For evaluating the dynamic pressure/strain sensing
performance of the graphene-PDMS foam sensor, the sensor was subjected
to compressive strains at three different frequencies using the same
minishaker setup described previously. The sensor was driven at 10,
35, and 70 Hz frequencies using a Rigol function generator connected
to a Brüel & Kjær (model number 2718) power amplifier.
The voltage output from the Wheatstone bridge circuit was acquired
by employing the National Instruments data acquisition system (DAQ,
NI USB-6009) and recorded using National Instruments Signal express
software at a sampling rate of 1 kHz.
Sensor
Applications in Gait Monitoring and
Human Motion Detection
For the gait simulation experiment
to demonstrate the gait monitoring capability of the graphene-PDMS
foam sensor, the same setup described previously in the case of the
pressure/strain sensor characterization experiment was employed. A
program was developed in the Bluehill software whereby the piston
moved with a constant speed of 20 mm/min to achieve a peak compressive
force of 20 N for 10 s followed by a ramp down at the same speed to
an intermediate compressive force of 3 N for 5 s before finally ramping
all the way down to 0 N. The test was repeated for 45 cycles to observe
the overall consistency in sensor response. For real-time gait and
walking monitoring applications, three identical sensors were placed
on a flat soft shoe sole and secured properly as schematically represented
in Figure b. The sensors
were connected to appropriate Wheatstone bridge circuits for continuous
data logging. Experiments involving walking, running, and leaning
in a periodic fashion were conducted on the shoe sole sensor assembly.
The setup was also used for differentiating between a low arch/flat
foot and a medium arch foot. For demonstrating the capability of the
foam sensor for wearable applications, the electrically bonded sensors
were secured on nitrile gloves as shown in the insets of Figure c,d. The gloves were
worn, five cycles of index finger and wrist flicking were conducted,
and the sensor responses were acquired. For all the experiments, data
was logged continuously using the data acquisition setup described
previously.