| Literature DB >> 3145391 |
W R Lutz1, N Maleki, B E Bjärngard.
Abstract
A beam-spot camera for measurements of x-ray focal spots of accelerators has been evaluated. The device consists of closely packed 0.25-mm-thick lead and 0.25-mm-thick cardboard strips. It is placed on radiographic film with the lamellae parallel to the beam axis and an exposure made. The images were scanned with a microdensitometer. The results indicate that the broadening of the source intensity profile at half maximum is of the order 1 mm, which permits the use of the beam-spot camera for acceptance testing and quality control. Longer tails in the density profile limit the quantitative information that can be extracted from the images.Entities:
Mesh:
Substances:
Year: 1988 PMID: 3145391 DOI: 10.1118/1.596214
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071