Gilbert L Nelson1, Amanda M Lines2, Job M Bello3, Samuel A Bryan2.
Abstract
Microfluidics is an appealing analytical tool in the global effort to close the nuclear fuel cycle. Using a microfluidic chip permits the analysis of greatly reduced sample volumes compared to what is necessary for traditional analytical methods. There is a commensurate reduction in disposal volume and cost. The development of novel sensors is necessary to take full advantage of the microchip configuration, where optical-spectroscopy-based approaches offer a powerful route to characterize chemical composition. This study uses simultaneously applied UV-vis and micro-Raman spectroscopies adapted to function on the microscale to analyze in situ both the Nd3+ (UV-vis-active) and HNO3 (Raman-active) concentrations in the same sample. An adjustable translation platform was designed to hold the micro-Raman probe above and perpendicular to the chip face and the UV-vis probe in the plane of the chip. These complimentary spectral techniques when processed through multivariate partial least-squares (PLS) models gave an accurate picture of the widely varying solution concentrations as a function of time for each solution component. Solution matrix effects can drastically alter analyte signatures as measured by both UV-vis absorbance and Raman spectroscopy. PLS methods successfully modeled these spectral changes and accurately measured concentrations of components of interest within the microfluidic chip.Entities:
Keywords: Partial Least Squares; Raman; UV−vis; chemometric; microfluidic; online monitoring
Mesh:
Substances:
Year: 2019 PMID: 31434479 DOI: 10.1021/acssensors.9b00736
Source DB: PubMed Journal: ACS Sens ISSN: 2379-3694 Impact factor: 7.711