High resolution time-of-flight MR-angiography at 7 T exploiting VERSE saturation, compressed sensing and segmentation.

Christian R Meixner1, Patrick Liebig2, Peter Speier3, Christoph Forman3, Bernhard Hensel4, Manuel Schmidt5, Marc Saake6, Michael Uder6, Arnd Doerfler5, Robin M Heidemann3, Sebastian Schmitter7, Armin M Nagel8.   

Abstract

BACKGROUND: 3D Time-of-Flight (TOF) MR-angiography (MRA) substantially benefits from ultra-high magnetic field strengths (≥7 T) due to increased Signal-to-Noise ratio and improved contrast. However, high-resolution TOF-MRA usually requires long acquisition times. In addition, specific absorption rate constraints limit the choice of optimal pulse sequence parameters, especially if venous saturation is employed.
PURPOSE: To implement and evaluate an arterial TOF-MRA for accelerated high-resolution angiography at ultra-high magnetic field strength. FIELD STRENGTHS/SEQUENCE: 7 T modified gradient-echo TOF sequence including venous saturation using Variable-Rate Selective Excitation (VERSE), Compressed Sensing (CS) and sparse application of saturation pulses, called segmentation, were included for acceleration. ASSESSMENT: To analyze the acceleration techniques all volunteers were examined with the same protocols. CS with different sampling patterns and regularization factors as well as segmentation were applied for acceleration. For comparison, conventional acceleration techniques were applied (GRAPPA PAT 3 and Partial Fourier (6/8 in slice/phase encoding)). Images were co-registered and 40 mm thick transversal maximum intensity projections were created to calculate the relative number of vessels. To analyze the visibility of small vessels, the lenticulostriate arteries (LSA) were examined. This was done via multiscale vessel enhancement filtering in a VOI and quantification via Fiji ImageJ as well as qualitatively evaluation by two radiologists. Additionally, the venous/arterial vessel-to-background ratios (vVBR/aVBR) were calculated for chosen protocols.
RESULTS: For the acceleration of a high resolution TOF-MRA (0.31 mm isotropic), under-sampling of 9.6 showed aliasing artifacts, whereas 7.2 showed no aliasing. The regularization factor R had a strong impact on the image quality according to smoothing (R = 0.01 to R = 0.005) and noise (R = 0.0005 to R = 0.00005). With the alternating sampling patterns it was shown that the k-space center should not be under-sampled too much. Additionally segmentation could be verified to be feasible for stronger acceleration with sufficient venous suppression.
CONCLUSION: The combination of several independent techniques (VERSE, CS with acceleration factor 7.2, R = 0.001, Poisson disc radius of 80%, 3 segments) enables the application of high-resolution (0.31 mm isotropic) TOF-MRA with venous saturation at 7 T in clinical time settings (TA ≈ 5 min) and within the SAR limits.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  7 Tesla; Compressed sensing; MR-angiography; Time-of-flight; Ultra-high field; VERSE

Mesh:

Year:  2019        PMID: 31434005     DOI: 10.1016/j.mri.2019.08.014

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  3 in total

1.  Added diagnostic values of three-dimensional high-resolution proton density-weighted magnetic resonance imaging for unruptured intracranial aneurysms in the circle-of-Willis: Comparison with time-of-flight magnetic resonance angiography.

Authors:  Younghee Yim; Seung Chai Jung; Jung Youn Kim; Seon-Ok Kim; Byung Jun Kim; Deok Hee Lee; Wonhyoung Park; Jung Cheol Park; Jae Sung Ahn
Journal:  PLoS One       Date:  2020-12-03       Impact factor: 3.240

2.  Imaging of the pial arterial vasculature of the human brain in vivo using high-resolution 7T time-of-flight angiography.

Authors:  Saskia Bollmann; Hendrik Mattern; Michaël Bernier; Simon D Robinson; Daniel Park; Oliver Speck; Jonathan R Polimeni
Journal:  Elife       Date:  2022-04-29       Impact factor: 8.713

3.  Optimization of undersampling parameters for 3D intracranial compressed sensing MR angiography at 7 T.

Authors:  Matthijs H S de Buck; Peter Jezzard; Aaron T Hess
Journal:  Magn Reson Med       Date:  2022-03-28       Impact factor: 3.737

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.