Fadwa Damak1,2, Maki Asano3, Koji Baba4, Mohamed Ksibi5, Kenji Tamura3.
Abstract
Elemental analysis of olive oils by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) is challenging because of the high organic load in olive oil samples and the low analyte concentrations. However, conflicting operating procedures in the preparation of oils prior to analysis by ICP-MS have been reported to overcome these difficulties. This study compared three methods of inorganic elements' extraction from olive oils: The two commonly used microwave-assisted, acid digestion, and liquid-liquid, ultrasound-assisted extraction methods; and an optimized method: The combined microwave digestion-evaporation. Overall, microwave digestion-based methods did not compare opportunely, and ultrasound-assisted extraction was found to provide the best accord between simplicity of use, detection limits and precision improvement. The detection limits were in the range of 0.3-160 µg·kg-1, 0.012-190 µg·kg-1 and 0.00061-1.5 µg·kg-1, while repeatabilities were in the range of 5-21%, 5.4-99% and 5.1-40% for the microwave digestion, the combined digestion-evaporation and the ultrasound assisted extraction, respectively. The ultrasound-assisted extraction is therefore recommended as a preparation method for olive oils prior to analysis by ICP-MS. The broader range of elements that can be accurately detected is expected to help increase the discriminatory power and performance of geographical traceability models.Entities:
Keywords: ICP-MS; evaporation; microwave digestion; multielements; olive oil; ultrasound-assisted extraction
Year: 2019 PMID: 31430917 PMCID: PMC6789567 DOI: 10.3390/mps2030072
Source DB: PubMed Journal: Methods Protoc ISSN: 2409-9279
ICP-MS operating conditions.
| Parameter | Method | ||
|---|---|---|---|
| Microwave Digestion | Digestion-Evaporation | Ultrasonic Extraction | |
| Instrument | ELAN | NexION | NexION |
| ICP Rf power (W) | 1100 | 1600 | 1600 |
| Plasma Ar flow rate (L·min–1) | 15 | 18 | 18 |
| Auxiliary Ar flow rate (L·min–1) | 1.30 | 1.20 | 1.20 |
| Nebulizer (carrier gas) flow rate (L·min–1) | 0.77 | 0.98 | 0.98 |
| Sampler and skimmer cones | Nickel | Nickel | Nickel |
| lens voltage (Deflector voltage) (V) | 7.5 | −11.50 | −11.50 |
| Analog stage voltage (V) | −1700 | −1800 | −1800 |
| Pulse stage voltage (V) | 950 | 1100 | 1100 |
| Discriminator threshold (V) | 19 | 13 | 13 |
| Quadrupole rod offset (V) | −1.5(STD), | 0 or 0.50 (STD), | 0 or 0.50 (STD), |
| Detector | Pulse | Pulse | Pulse |
| Speed of peristaltic pump (rpm) | 20 | 20 | 20 |
| Sweeps/reading | 20 | 20 | 20 |
| Replicate/reading | 1 | 1 | 1 |
| Replicates | 3 | 3 | 3 |
| Dwell time (ms) | 50 | 50 | 50 |
| Scan mode | Peak hopping | Peak hopping | Peak hopping |
| STD and KED: rejection parameter a and rejection parameter q | 0, 0.25 | 0, 0.25 | 0, 0.25 |
| DRC mode: CH4 reaction gas flow (L·min–1) | 0.60 | 0.60 | 0.60 |
| DRC mode: rejection parameter a and Rejection parameter q | 0, 0.65 | 0, 0.65 | 0, 0.65 |
| KED mode He reaction gas flow (L·min–1) | - | 3.5 | 3.5 |
STD, standard mode; DRC, dynamic reaction cell mode; KED, kinetic energy discrimination mode.
Figure 1Scheme of optimization procedure for sample preparation prior to multielement determination by ICP-MS.
Operating program for the microwave system used for the three batches.
| Batch N° | Step | Time (mn) | Power (W) | Temp 1 (°C) |
|---|---|---|---|---|
| 1 | 1 | 2 | 1000 | 50 |
| 2 | 1 | 0 | 30 | |
| 3 | 31 | 1000 | 190 | |
| 4 | 1 | 0 | 160 | |
| 5 | 6 | 1000 | 190 | |
| 6 | 13 | 1000 | 190 | |
| 2 | 1 | 2 | 1000 | 70 |
| 2 | 1 | 0 | 50 | |
| 3 | 31 | 1000 | 210 | |
| 4 | 1 | 0 | 180 | |
| 5 | 6 | 1000 | 210 | |
| 6 | 13 | 1000 | 210 | |
| 3 | 1 | 2 | 1000 | 90 |
| 2 | 1 | 0 | 70 | |
| 3 | 31 | 1000 | 230 | |
| 4 | 1 | 0 | 200 | |
| 5 | 6 | 1000 | 230 | |
| 6 | 13 | 1000 | 230 |
Calibration range of elements in the three analytical methods’ analysis by ICP-MS.
| Solution Name (Concentrations in µg·L−1) | ||||||
|---|---|---|---|---|---|---|
| Method | Na | Ca | Mg | Fe | XSTC-622B | N * |
| Digestion | - | - | - | - | 0.01–50 | 8 |
| Combined digestion-evaporation | 0.05–100 | 0.025–50 | 0.01–20 | 0.01–20 | 0.005–10 | 11 |
| Ultrasound-assisted extraction | 0.05–50 | 0.025–25 | 0.01–10 | 0.01–10 | 0.005–5 | 12 |
* number of calibration points evenly distributed across the corresponding range.
Results of determination of analytical procedure parameters (Microwave digestion at 200 °C, according to Llorent-Martinez et al., 2014 with minor modifications).
| Element | Isotope | Operation Mode | LOD (µg·kg−1) | LOQ (µg ·kg −1) | Repeatability (RSD %) * | Linearity (R2) | Accuracy (%) |
|---|---|---|---|---|---|---|---|
| Na | 23 | Standard | 120 | 350 | 9.9 | 0.9998 | 84 |
| Mg | 24 | Standard | 160 | 470 | 12 | 0.9991 | 66 |
| Fe | 56 | DRC (CH4) | 120 | 390 | 13 | 0.9999 | 88 |
| Zn | 66 | DRC (CH4) | 110 | 360 | 16 | 0.9997 | 97 |
| V | 51 | Standard | 1.7 | 5.6 | 14 | 1.0000 | 97 |
| Mn | 55 | Standard | 6.0 | 20 | 20 | 1.0000 | 103 |
| As | 75 | Standard | 0.73 | 2.4 | 20 | 0.9990 | |
| Rb | 85 | Standard | 0.30 | 1.0 | 17 | 0.9996 | |
| Sr | 88 | Standard | 5.1 | 17 | 5 | 0.9999 | |
| Ba | 138 | Standard | 4.6 | 15 | 15 | 1.0000 | 102 |
| Pb | 208 | Standard | 6.9 | 23 | 21 | 1.0000 | 97 |
* based on the mean RSD of three replicates of 21 olive oil samples.
Comparison of the limits of detection of the microwave digestion method between our results and those of the original research (for elements in common).
| LOD (µg·Kg−1) | ||
|---|---|---|
| Element | Our Study | Original Study |
| Fe | 120 | 600 |
| V | 1.7 | 15 |
| As | 0.73 | 15 |
| Pb | 6.9 | 4 |
Counts at mass number 12 corresponding to the most abundant isotopes of carbon: A comparative method of residual carbon content estimation under different decomposition conditions.
| One Step | Two Steps | |||||||
|---|---|---|---|---|---|---|---|---|
| 190 °C | 210 °C | 230 °C | 210 °C then 230 °C | |||||
| Mass number | NE* | E** | NE | E | NE | E | NE | E |
| 12 | 172,855,743 | 13,216,179 | 217,036,881 | 77,300,752 | 219,254,250 | 191,976,927 | over range | 76,931,314 |
NE, no evaporation; E, with evaporation.
Results of determination of analytical procedure parameters (Microwave digestion at 190 °C followed by evaporation).
| Element | Isotope | Mode | LOD (µg·kg−1) | LOQ (µg· kg−1) | Repeatability (RSD %) | Linearity (R2) |
|---|---|---|---|---|---|---|
| Na | 23 | He | 13 | 45 | 5.4 | 0.9985 |
| Mg | 24 | He | 3.5 | 11 | 27 | 0.9991 |
| Ca | 44 | CH4 | 190 | 640 | 6.6 | 0.9990 |
| V | 51 | He | 0.012 | 0.041 | 53 | 0.9996 |
| Cr | 52 | CH4 | 0.47 | 1.6 | 25 | 0.9996 |
| Mn | 55 | He | 0.49 | 1.6 | 52 | 0.9989 |
| Fe | 56 | He | 5.8 | 19 | 45 | 0.9957 |
| Ni | 60 | CH4 | 2.5 | 8.2 | 26 | 0.9998 |
| Cu | 63 | He | 2.3 | 7.7 | 16 | 0.9989 |
| Zn | 64 | CH4 | 13 | 42 | 12 | 0.9997 |
| As | 75 | He | 0.093 | 0.31 | 99 | 0.9993 |
| Rb | 85 | STD | 0.036 | 0.12 | 38 | 1.000 |
| Sr | 88 | STD | 0.098 | 0.33 | 15 | 1.000 |
| Mo | 98 | CH4 | 0.15 | 0.51 | 46 | 0.9999 |
| Pb | 208 | STD | 2.4 | 8.1 | 22 | 0.9999 |
He, Helium collision mode; CH4, CH4 dynamic reaction cell mode; STD, no gas standard mode; R2, linear regression coefficient of calibration for LOD and LOQ determinations.
Results of determination of analytical procedure parameters for ultrasonic extraction of elements.
| Element | Isotope | Mode | LOD (µg·kg−1) | LOQ (µg·kg−1) | Repeatability (RSD %) | Linearity (R2) | Accuracy (%) |
|---|---|---|---|---|---|---|---|
| Na | 23 | He | 0.42 | 1.4 | 11 | 0.9985 | 136 |
| Mg | 24 | He | 0.11 | 0.35 | 11 | 0.9986 | 73 |
| Ca | 44 | He | 1.5 | 4.9 | 13 | 0.9811 | |
| Ti | 47 | He | 0.16 | 0.45 | 11 | 0.9977 | |
| V | 51 | He | 0.069 | 0.23 | 1.9 | 0.9975 | |
| Cr | 52 | He | 0.035 | 0.12 | 12 | 0.9987 | 64 |
| Mn | 55 | He | 0.0060 | 0.021 | 10 | 0.9979 | 67 |
| Fe | 56 | He | 0.14 | 0.47 | 5.1 | 0.9984 | |
| Ni | 60 | CH4 | 0.18 | 0.60 | 40 | 0.9974 | 81 |
| Cu | 63 | He | 0.20 | 0.66 | 30 | 0.9931 | 80 |
| Zn | 66 | He | 0.077 | 0.26 | 18 | 0.9988 | 84 |
| As | 75 | CH4 | 0.036 | 0.12 | 6.2 | 0.9987 | |
| Rb | 85 | STD | 0.00061 | 0.0021 | 8.9 | 0.9993 | 79 |
| Sr | 88 | STD | 0.0025 | 0.0085 | 9.1 | 0.9993 | 77 |
| Mo | 98 | STD | 0.0047 | 0.016 | 8.6 | 0.9970 | 78 |
| Ba | 138 | STD | 0.0014 | 0.0049 | 13 | 0.9990 | 133 |
| Pb | 208 | STD | 0.0035 | 0.012 | 13 | 0.9989 |
He, Helium collision mode; CH4, CH4 dynamic reaction cell mode; STD, no gas standard mode; R2, linear regression coefficient of calibration for LOD and LOQ determinations.