Literature DB >> 31424123

Pulmonary vein isolation using a higher power shorter duration CLOSE protocol with a surround flow ablation catheter.

Benjamin Berte1, Gabi Hilfiker1, Ian Russi1, Federico Moccetti1, Florim Cuculi1, Stefan Toggweiler1, Frank Ruschitzka2, Richard Kobza1.   

Abstract

INTRODUCTION: The CLOSE protocol combines ablation index (AI) and ≤6 mm interlesion distance using standard power settings for the treatment of atrial fibrillation (AF). The purpose was to compare the safety and efficacy of a conventional CLOSE and a higher power shorter duration (HPSD)-CLOSE pulmonary vein isolation (PVI) strategy. METHODS AND
RESULTS: All consecutive patients referred for PVI were included after informed consent was obtained from them. Group 1 was treated with a standard CLOSE protocol and group 2 with a HPSD-CLOSE protocol (45 W anterior and 35 W posterior). Procedural parameters and 6-month follow-up were analyzed. In total, 174 patients (group 1: n = 94 [paroxysmal: n = 74]; group 2: n = 80 [paroxysmal: n = 65], similar baseline characteristics) were included. PVI was reached in all, but procedure duration (82 ± 18 minutes vs 100 ± 22 minutes; P < .0001) and radiofrequency (RF) time (23 ± 5 minutes vs 36 ± 11 minutes; P < .0001) was shorter in group 2. First pass isolation was similar in groups 2 and 1 (left veins: 94% vs 90%; P = .42 and right veins: 83% vs 84%; P = .79, respectively). Six-month off- antiarrhythmic drugs freedom of AF/AT was similar in groups 2 and 1 (82% [paroxysmal: 86%] vs 83% [paroxysmal: 88%]; P = .93, respectively). Major complications were similar (group 2: 1% vs group 1: 3%; P = .39).
CONCLUSION: A higher-power short duration approach can shorten a CLOSE procedure and reduce ablation time without having a negative impact on safety or efficiency.
© 2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  CLOSE protocol; ablation; atrial fibrillation; efficiency; high power short duration

Mesh:

Substances:

Year:  2019        PMID: 31424123     DOI: 10.1111/jce.14122

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  14 in total

1.  Meta-analysis comparing outcomes of high-power short-duration and low-power long-duration radiofrequency ablation for atrial fibrillation.

Authors:  Dibbendhu Khanra; Abdul Hamid; Saurabh Deshpande; Anindya Mukherjee; Sanjiv Petkar; Mohammad Saeed; Indranill Basu-Ray
Journal:  Anatol J Cardiol       Date:  2022-01       Impact factor: 1.596

2.  Two years after pulmonary vein isolation guided by ablation index-a multicenter study.

Authors:  Pedro A Sousa; Luís Puga; Luís Adão; João Primo; Ziad Khoueiry; Ana Lebreiro; Paulo Fonseca; Philippe Lagrange; Luís Elvas; Lino Gonçalves
Journal:  J Arrhythm       Date:  2022-03-15

Review 3.  Efficiency, Safety, and Efficacy of High-Power Short-Duration Radiofrequency Ablation in Patients with Atrial Fibrillation.

Authors:  Xuerong Sun; Jiang Lu; Jinxuan Lin; Tianjie Feng; Ni Suo; Lihui Zheng; Zhimin Liu; Gang Chen; Xiaohan Fan; Shu Zhang; Guodong Niu
Journal:  Cardiol Res Pract       Date:  2021-02-15       Impact factor: 1.866

4.  Regional differences in the predictors of acute electrical reconnection following high-power pulmonary vein isolation for paroxysmal atrial fibrillation.

Authors:  Kyoichiro Yazaki; Koichiro Ejima; Shohei Kataoka; Miwa Kanai; Satoshi Higuchi; Daigo Yagishita; Morio Shoda; Nobuhisa Hagiwara
Journal:  J Arrhythm       Date:  2021-07-23

5.  The impact of current strategy using intracardiac echocardiography, lesion index, and minimum substrate ablation on clinical outcomes after catheter ablation procedure for atrial fibrillation.

Authors:  Tetsuma Kawaji; Takanori Aizawa; Shun Hojo; Akihiro Kushiyama; Hidenori Yaku; Kenji Nakatsuma; Kazuhisa Kaneda; Masashi Kato; Takafumi Yokomatsu; Shinji Miki
Journal:  J Arrhythm       Date:  2021-08-05

6.  Acute and long-term efficacy of ablation index-guided higher power shorter duration ablation in patients with atrial fibrillation: A prospective registry.

Authors:  So-Ryoung Lee; Hyoung-Seob Park; Eue-Keun Choi; Euijae Lee; Seil Oh
Journal:  J Arrhythm       Date:  2021-07-21

7.  Safety and Efficacy of High Power Shorter Duration Ablation Guided by Ablation Index or Lesion Size Index in Atrial Fibrillation Ablation: A Systematic Review and Meta-Analysis.

Authors:  Xing Liu; Chun Gui; Weiming Wen; Yan He; Weiran Dai; Guoqiang Zhong
Journal:  J Interv Cardiol       Date:  2021-06-02       Impact factor: 2.279

8.  Comparison of high-power and conventional-power radiofrequency energy deliveries in pulmonary vein isolation using unipolar signal modification as a local endpoint.

Authors:  Koichiro Ejima; Satoshi Higuchi; Kyoichiro Yazaki; Shohei Kataoka; Daigo Yagishita; Miwa Kanai; Morio Shoda; Nobuhisa Hagiwara
Journal:  J Cardiovasc Electrophysiol       Date:  2020-05-15

9.  Safety and efficacy of the ThermoCool SmartTouch SurroundFlow catheter for atrial fibrillation ablation: A meta-analysis.

Authors:  Chao-Feng Chen; Xiao-Fei Gao; Mei-Jun Liu; Chao-Lun Jin; Yi-Zhou Xu
Journal:  Clin Cardiol       Date:  2019-11-19       Impact factor: 2.882

10.  Comparison of high-power short-duration and low-power long-duration radiofrequency ablation for treating atrial fibrillation: Systematic review and meta-analysis.

Authors:  Chao-Feng Chen; Jing Wu; Chao-Lun Jin; Mei-Jun Liu; Yi-Zhou Xu
Journal:  Clin Cardiol       Date:  2020-10-27       Impact factor: 2.882

View more