The ageing kinetics of amorphous atactic (a-PS), isotactic (i-PS), and syndiotactic (s-PS) polystyrene were studied by means of flash-differential scanning calorimetry. The specimens were aged for up to 2 h at six different ageing temperatures: the optimum ageing temperature, that is, the temperature at which the enthalpy overshoot at the glass transition is maximal for the given elapsed time, and five ageing temperatures ranging from 20 to 80 K below the optimum ageing temperature. A logarithmic increase of the enthalpy overshoot with ageing time is observed for specimens at their optimum ageing temperatures. For temperatures significantly lower than the optimum, there is a range where the enthalpy overshoot is constant, but for higher temperatures (still below the optimum), a logarithmic increase is also observed. Moreover, the ageing kinetics appear to depend on tacticity, with s-PS and i-PS exhibiting the slowest and fastest ageing kinetics, respectively, and a-PS exhibiting ageing kinetics between these two extremes.
The ageing kinetics of amorphous atactic (a-PS), isotactic (i-PS), and syndiotactic (s-PS) polystyrene were studied by means of flash-differential scanning calorimetry. The specimens were aged for up to 2 h at six different ageing temperatures: the optimum ageing temperature, that is, the temperature at which the enthalpy overshoot at the glass transition is maximal for the given elapsed time, and five ageing temperatures ranging from 20 to 80 K below the optimum ageing temperature. A logarithmic increase of the enthalpy overshoot with ageing time is observed for specimens at their optimum ageing temperatures. For temperatures significantly lower than the optimum, there is a range where the enthalpy overshoot is constant, but for higher temperatures (still below the optimum), a logarithmic increase is also observed. Moreover, the ageing kinetics appear to depend on tacticity, with s-PS and i-PS exhibiting the slowest and fastest ageing kinetics, respectively, and a-PS exhibiting ageing kinetics between these two extremes.
Glasses are known to be far from their
thermodynamic equilibrium,
and their structural relaxation toward equilibrium, via a sequence
of molecular rearrangements, is known as physical ageing.[1] Physical ageing of polymeric glasses is known
to affect the material properties, like the mechanical[2,3] or gas-transport properties,[4] posing
limitations on their applications. A lot of studies have been conducted
on the influence of physical ageing of glassy polymers on yielding,[1−3,5,6] creep,[7−9] volume relaxation,[8,10−14] enthalpy changes,[9,11,15−17] molecular-scale dynamics,[18−21] and structural changes.[22−24] However, the connection between
the molecular structure and the ageing kinetics has not yet been established.Despite the fact that numerous techniques are employed to study
the large number of properties affected by physical ageing in glassy
polymers, only one of them can cool and heat the material rapidly
enough to allow probing of the ageing kinetics by starting from a
“true” rejuvenated amorphous state. This technique is
flash-differential scanning calorimetry (flash-DSC) in which ultra-high
heating and cooling rates (>100 000 K/min) can be applied.[25] The majority of the ageing studies with flash-DSC
involve measurements concerning a single material such as atactic
polystyrene (a-PS),[26−30] polycarbonate (PC),[29] poly lactic-acid,[31,32] and others.[33,34] Until now, there are only very
few studies investigating the effect of tacticity on the ageing kinetics[35] of glassy amorphous polymers. The main reason
for the lack of studies is that syndiotactic and isotactic polymers
crystallize under normal cooling rates, resulting in a semicrystalline
material. For polystyrenes, there is another limiting factor next
to the high crystallization rate[36] of s-PS,
namely, its melting temperature is very close to its degradation temperature,
which makes the preparation of an amorphous sample in oxygen atmosphere
impossible because the material starts to degrade before the crystallinity
is erased. The high crystallization rate can be tackled by the high
cooling rate that flash-DSC offers, and the complete melting of crystals
before the material degrades has been taken care of by using a nitrogen
atmosphere during the measurement.In this work, we study the
effect of tacticity on the ageing kinetics
at high and low ageing temperatures for atactic, isotactic, and syndiotactic
polystyrene using flash-DSC in their glassy state.
Methodology
Materials
Atactic and syndiotactic PS (Figure ) with syndio contents ≥90%
were bought in powder and pellet forms, respectively, from Polymer
Source.[37] Isotactic PS (Figure ) with 90% isotactic content
was bought in the powder form from Scientific Polymer Products.[38] The corresponding molecular weights and polydispersity
indices (PDIs) are listed in Table .
Figure 1
Stereochemistry of (a) atactic, (b) syndiotactic, and
(c) isotactic
polystyrene.
Table 1
Number-Average
Molecular Weight Mn and PDI of the Polystyrene
Samples
sample
Mn (kg/mol)
PDI (-)
a-PS
412
1.05
s-PS
250
3.5
i-PS
400
2
Stereochemistry of (a) atactic, (b) syndiotactic, and
(c) isotactic
polystyrene.
Differential Scanning Calorimetry
Flash-Differential Scanning
Calorimetry
Experiments
were carried out in Mettler Toledo Flash-DSC 1 equipped with a Huber
TC100 intracooler under a constant flow of dry nitrogen using UFS1
sensors.
Differential Scanning Calorimetry
Experiments were
carried out in a Mettler-Toledo 823e/700 module with a Cryostat intracooler
using a 50 μL aluminum pan under a constant flow of dry nitrogen.
Sample Preparation
For the flash-DSC measurements,
very small specimens of material of approximately 30–170 ng
were cut and put onto the sensor with the use of an eyelash. For the
conventional DSC, the powder-form samples were put in an aluminum
pan with a mass between 5 and 10 mg.
Estimation of Sample Mass
In the case of the flash-DSC
measurements, the sample mass cannot be measured with a conventional
scale because the weight is too small (nanogram). For semicrystalline
materials, the mass can be estimated as follows. A sample of known
mass is cooled and heated in conventional DSC at 20 K/min, and the
specific melting enthalpy (Δhm)
is calculated by peak integration of the heating curve. Afterward,
a sample of unknown mass is cooled at 20 K/min and heated at 1000
K/s in the flash-DSC equipment. The melting enthalpy (ΔHm) is calculated from the heating curve, and
the mass of the flash-DSC sample can be determined byIn the case of amorphous polymers,
the sample mass can be estimated by using the method of Cebe et al.,[39] employing the liquid heat capacity from flash-DSC
(cooling rate 1000 K/s) and the specific heat-capacity from the ATHAS
bank.[40] The sample mass mam is determined bywhere C is the heat
capacity obtained from the flash-DSC
experiments far above the glass transition temperature and c is the specific
liquid heat capacity taken from the ATHAS bank. The validity of the
method is confirmed by applying both eqs and 2 to a semicrystalline sample.
The error between the two mass-estimation methods is found to be about
5%.
Methods
In order to study the ageing kinetics of the
selected materials, the temperature protocol shown in Figure was used. The following steps
can be identified: Step i: the sample was heated well above Tg (a-PS) or Tm (s-PS
and i-PS) at a rejuvenation temperature Trej (Table and Figure ) in order to erase
its thermomechanical history (i.e., thermal rejuvenation). Step ii:
The sample was rapidly quenched at a cooling rate of 1000 K/s to an
end-quenching temperature Tq well below Tg, and subsequently heated rapidly (heating
rate: 1000 K/s) above Tg to obtain the
heating curve which serves as a reference state. Step iii: The sample
is rapidly quenched (1000 K/s) and then heated (1000 K/s) to the chosen
ageing temperature, Ta, and held at that
temperature for a specific ageing time ta. Step iv: The sample was quenched rapidly (1000 K/s) at 25 °C
for ageing temperatures higher than 25 and 0 °C for ageing temperatures
below 25 °C to capture the aged state and heated rapidly (1000
K/s) above Tg (or Tm) to obtain the endothermic peak caused by ageing, and to
thermally rejuvenate the sample for the next measurement. Step v:
One more cycle of quenching and heating was performed to validate
the rejuvenated state of the polymer (heating curve).
Figure 2
Schematic representation
of the flash-DSC heating protocol: (i)
Heating above Tg or Tm, (ii) reference-state heating curve, (iii) isothermal
ageing for ageing time ta and at temperature Ta, (iv) aged-state heating curve, and (v) reference-state
heating curve. The values of rejuvenation, Trej, and end-quenching temperature, Tq, are provided in Table .
Table 2
Glass-Transition
Temperature, Tg, and Melting Temperature, Tm, for a-PS, s-PS, and i-PS, Measured during
Heating with
a Heating Rate of 1000 K/s, and Rejuvenation, Trej, and End-Quenching Temperatures, Tqa
sample
Tg (°C)
Tm (°C)
Trej (°C)
Tq (°C)
a-PS
128
190
25 and 0
s-PS
124
270
310
25 and 0
i-PS
126
240
270
25 and 0
The end-quenching temperature is
25 °C for ageing temperatures higher than 25 and 0 °C for
ageing temperatures below 25 °C.
Schematic representation
of the flash-DSC heating protocol: (i)
Heating above Tg or Tm, (ii) reference-state heating curve, (iii) isothermal
ageing for ageing time ta and at temperature Ta, (iv) aged-state heating curve, and (v) reference-state
heating curve. The values of rejuvenation, Trej, and end-quenching temperature, Tq, are provided in Table .The end-quenching temperature is
25 °C for ageing temperatures higher than 25 and 0 °C for
ageing temperatures below 25 °C.
Analysis
The analysis of experimental data concerning
the cooling-rate dependence of Tg and
physical ageing can be analyzed in terms of enthalpy loss (ΔHa) or the fictive temperature (Tf).[29] Tool[41] defined Tf as a measure of
the glass structure. The intersection of the extrapolated glass line
and the extrapolated liquid line determines Tf. The Tf of an aged glass is related
to the change in enthalpy with ageing[29]where ΔHa is the enthalpy loss upon ageing, Tf is the fictive temperature of the aged glass, Tf0 is the fictive temperature of the unaged
glass, and
ΔC is the step
change in heat capacity. During physical ageing, Tf0 decreases to an equilibrium value, while ΔHa increases from a zero value to the equilibrium
enthalpy value.[29] We chose to analyze our
data using the ΔHa changes. Both
types of analysis will yield identical conclusions/observations. The
analysis procedure is described below.
Fitting a Straight Line
to the Liquid Regime[29]
The calculation
of the enthalpy overshoot of the
aged specimen (i.e., the integral under the C(T)-curve) relies on drawing
an unbiased liquid line, especially when a large overshoot is present.
Thus, each heating scan was superposed with the subsequent scan of
a freshly quenched specimen. The liquid line was determined as the
best linear fit of the C(T)-curve at high temperatures, away from the enthalpy
overshoot and the transition region. A consistent liquid line is crucial
for calculating the integral under the C(T)-curve. As shown in Figure , the enthalpy overshoot
of the aged specimen is calculated in between the point where the C(T)-curve
departs from the linear liquid line till the point where it intersects
again. This integral is considered positive. Then, we apply the same
procedure to the rejuvenated sample and subtract its integral from
that of the aged one.
Figure 3
Representative curve for the calculation of the enthalpy
loss ΔHa upon ageing, according
to the Petrie method.[42]
Representative curve for the calculation of the enthalpy
loss ΔHa upon ageing, according
to the Petrie method.[42]
Determination of Excess Enthalpy
The excess enthalpy
was determined using the method of Petrie[42] in which the enthalpy loss ΔHa upon ageing is calculated by the difference of the integrated C(T)-curves
of the aged and reference (rejuvenated) material (Figure )Two measurements per sample are conducted
for each ageing time and temperature.
Sensor Corrections
Conditioning and correction were
performed on the flash-DSC chip sensors before the sample is positioned
on the sensor according to the manufacturers recommended protocol.
No additional temperature corrections were applied because the calibration
by the manufacturer was found to have an error of about 0.6–0.7
K.[43]
Results
In Figure , the
flash-DSC heating curves for atactic, syndiotactic, and isotactic
amorphous PS annealed at various ageing temperatures and ageing times
are shown. It can be seen that the area of the endothermic peak as
well as the temperature where the endothermic peak appears are increasing
with the ageing time. Also, the absence of a melting peak, as a result of the high cooling
rate (1000 K/s) in the case of semicrystalline polymers, validates
the fact that the samples are in an amorphous state (Figure ).
Figure 4
Heat flow Q versus temperature T for (a) a-PS, (b) s-PS, and
(c) i-PS, for various ageing times.
Heat flow Q versus temperature T for (a) a-PS, (b) s-PS, and
(c) i-PS, for various ageing times.We study the effect of ageing temperature Ta on the excess enthalpy ΔHtot – ΔHrej for an
annealing
time of 10 min for atactic, syndiotactic, and isotactic amorphous
PS (Figure ). Figure a shows that for
a-PS, the excess enthalpy increases linearly with temperature where
it reaches a maximum at around 102 °C (≈26 K below the
glass-transition temperature). After this temperature, the enthalpy
overshoot decreases linearly. Similarly, the s-PS (Figure b) excess enthalpy follows
a linear increase with an optimum at ≈96 °C, (≈28
K below Tg) and a decrease after this
optimum as well. i-PS (Figure c) has a similar behavior with an optimum enthalpy overshoot
at ≈94 °C (32 K below Tg).
Based on these results, we choose the temperature in which each polymer
shows the largest overshoot as the optimum ageing temperature (Topt), determined by fitting a polynomial of
the fifth order to the ΔHtot –
ΔHrej versus Ta data (Figure ). We perform measurements also at 20 K (Topt-20K), 30 K (Topt-30K), 40 K (Topt-40K), 60 K (Topt-60K), and 80 K (Topt-80K) below Topt for all specimens. It is
also noted that the behavior displayed in Figure on the left and right sides of the maximum, Topt, is asymmetric; similar observations have
been made earlier.[44] The fact that the
left side changes only smoothly with temperature is indicative of
non-α-processes being relevant in this regime.
Figure 5
Excess enthalpy, ΔHtot –
ΔHrej, vs annealing temperature, Ta, for (a) a-PS, (b) s-PS, and (c) i-PS. The
ageing time is ta = 10 min. Lines represent
the polynomial fits (see main text).
Excess enthalpy, ΔHtot –
ΔHrej, vs annealing temperature, Ta, for (a) a-PS, (b) s-PS, and (c) i-PS. The
ageing time is ta = 10 min. Lines represent
the polynomial fits (see main text).In Figure , the
excess enthalpy (ΔHtot –
ΔHrej) as a function of the ageing
time for the various ageing temperatures for atactic, syndiotactic,
and isotactic PS is shown. It can be seen that ΔHtot – ΔHrej follows
a logarithmic increase with time for the optimum ageing temperatures.
For very short ageing times (≤0.1 s), the ΔHtot – ΔHrej for
the three PS specimens is close to zero. For longer ageing times (≥10
s), the excess enthalpy shows a logarithmic increase for all specimens.
At ageing temperatures below the optimum (Figure ), it can be observed that the increase of
ΔHtot – ΔHrej does not follow a logarithmic increase and is reduced
compared to that for long ageing times. Similar to the observations
for maximum ageing temperature for short ageing times, the enthalpy
overshoot observed for the three PS specimens is close to zero.
Figure 6
Excess enthalpy,
ΔHtot –
ΔHrej, vs ageing time, ta, for (a) a-PS, (b) s-PS, and (c) i-PS.
Excess enthalpy,
ΔHtot –
ΔHrej, vs ageing time, ta, for (a) a-PS, (b) s-PS, and (c) i-PS.Using time–temperature superposition, an
activation energy
can be determined. In Figure , the ΔHtot – ΔHrej curves are shifted horizontally along the
logarithmic ageing-time axis, taking Topt as a reference. The good superposition of lines allows calculating the activation
energy by plotting the logarithmic shift-factor (a) as a function of temperature and obtaining
an Arrhenius plot for temperatures from Topt-40K to Topt-20K. The activation energies Ea for a-PS, s-PS, and i-PS calculated from the
slope (Ea/R, where R is the gas constant) in Figure are found to be (166.6 ± 4.0) kJ/mol
for a-PS, (163.8 ± 6.0) kJ/mol for s-PS and (144.4 ± 9.7)
kJ/mol for i-PS. Even if we take the lowest values for Ea/R for a-PS and s-PS, we cannot reach
the highest value for i-PS. This indicates that despite the differences
being subtle, they are by no means negligible.
Figure 7
Excess enthalpy obtained
from time–temperature superposition
using the optimum ageing temperature as the reference, Tref = Topt, for (a) a-PS,
(b) s-PS, and (c) i-PS.
Figure 8
Arrhenius plot of the shift factor, aT, for i-PS (half-filled circles), s-PS (filled circles), and a-PS
(empty circles). The activation energy is represented by the slope
of the lines.
Excess enthalpy obtained
from time–temperature superposition
using the optimum ageing temperature as the reference, Tref = Topt, for (a) a-PS,
(b) s-PS, and (c) i-PS.Arrhenius plot of the shift factor, aT, for i-PS (half-filled circles), s-PS (filled circles), and a-PS
(empty circles). The activation energy is represented by the slope
of the lines.The slope in the Arrhenius
plot (Figure ) for
a-PS calculated in this work has a
value of around 20 kK, while the slope calculated in the work of Koh
and Simon[43] has a value of 13 kK. Considering
that in this work, we used a limited temperature range for the construction
of the Arrhenius plot in Figure compared to the work of Koh and Simon,[43] errors may have been introduced because of the
errors in the shift factors. Despite these differences, the activation
energies measured by flash-DSC are significantly smaller than the
ones typically reported in the literature for the α-process,[45] a fact that was also noted earlier.[46] The relatively small activation energies are,
again, supporting the idea that non-α-processes are relevant
for the low-temperature behavior of the glass.Furthermore,
we quantify the ageing rate from the slope of ΔHtot – ΔHrej in
the range where it exhibits a linear increase with respect to
the logarithm of timeIn Figure , the
ageing rate, Rag, is plotted as a function
of the ageing temperature Ta. At Topt-80K, Rag is close to zero, increases slightly with ageing temperature at Topt-60K, and has a large increase at Topt-40K. The ageing rates for i-PS, a-PS,
and s-PS are comparable for temperatures from Topt-80K to Topt-40K. In addition, a-PS exhibits higher ageing rates than s-PS for temperatures
above Topt-40K, where i-PS has
the highest ageing rate of all specimens at the optimum Ta.
Figure 9
Ageing rate (Rag) as a function
of
ageing temperature for i-PS (half-filled circles), s-PS (filled circles),
and a-PS (empty circles).
Ageing rate (Rag) as a function
of
ageing temperature for i-PS (half-filled circles), s-PS (filled circles),
and a-PS (empty circles).
Discussion
PS[15,47] as well as other glassy
polymers such as
PC[48,49] and PVC[50] have
been reported to have logarithmic enthalpy overshoot increase at high-temperature
ageing. For ageing at the optimum ageing temperature, the enthalpy
overshoot increases logarithmically with time. For ageing temperatures
below the optimum ageing temperatures, ageing shows a non-logarithmic
dependence on time. The farther the ageing temperature of the material
is from its optimum ageing temperature, the longer the material stays
in an arrested state until it shows a logarithmic increase in excess
enthalpy with time (Figure ), similar to the enthalpy increase which has been reported
for PC.[48]While all materials exhibit
the same ageing mechanisms for all
temperatures (logarithmic increase of excess enthalpy with Topt and delayed increase of the excess enthalpy
with ageing at temperatures far below optimum), it is worth noting
the effect of tacticity on the ageing rate (Figure ). Specifically, a-PS and i-PS show a similar
ageing rate for temperatures from Topt-40K to Topt-20K, while at the optimum
ageing temperature, the fastest ageing kinetics are observed for i-PS.
s-PS shows the slowest ageing kinetics between the three polymers,
specifically for temperatures from Topt-40K to Topt.We believe that the reason
behind the difference of the rate at
which a-PS, s-PS, and i-PS age is related to the difference in chain
flexibility of those polymers. Nakaoki and Kobayashi[51] studied, by solid-state high-resolution 13C
NMR, the gauche content of amorphous atactic, syndiotactic, and isotactic
PS, finding 27.9, 25.0, and 34.3%, respectively. Based on their results,
they concluded that i-PS is more likely to have fewer trans sequences,
which leads to a shorter random coil and to a smaller characteristic
ratio (C∞). This has been confirmed
by small-angle neutron scattering studies which have revealed that
the characteristic ratio is C∞s-PS > C∞a-PS > C∞i-PS, that points out that the
stiffest
chain is the one of s-PS.[52] Specifically,
the C∞s-PS value is 7.9 to 13 for increasing Mw from 214 000 to 380 000 g/mol.[53] The C∞a-PS value is 6.7 to 9.6 for increasing Mw from 325 000 to 500 000 g/mol,
and the C∞i-PS values are 4.9 and 5 for Mw = 254 000 and 500 000 g/mol,
respectively.[53] Syndiotactic sequences
make the polymer chain rigid[54] and slow
down the reorganization of the chains, which consequently leads to
slower ageing kinetics, as it has been shown in our measurements (Figure ). Tensile creep
measurements on polyimide Kapton-H and cellulose acetate butyrate
ester have also suggested that the ageing kinetics are slower for
the more rigid polymers.[55]As shown
above, the rate of ageing decreases with increasing chain
stiffness, s-PS ageing significantly slower than i-PS. This is in
stark contrast to the behavior of these two materials in crystallization;
s-PS crystallizes about hundred times faster than i-PS because of
differences in the conformation of the C–C backbone and the
work of chain folding.[56] This contrast
shows that ageing and crystallization are clearly distinct phenomena.
Conclusions
The effect of tacticity on the ageing kinetics of glassy amorphous
PS was investigated by flash-DSC. Our results show that the tacticity
does have an effect on the physical ageing kinetics, with syndiotactic
and isotactic PS showing the two extremes, slow and fast ageing kinetics,
respectively, and atactic PS showing ageing kinetics between these
two extremes.The enthalpy overshoot of all PS specimens follows
a logarithmic
increase with ageing time at their optimum ageing temperature. For
temperatures far below its optimum, there is a range, different for
each specimen, where the enthalpy overshoot is constant, whereas at
higher temperatures (but still below the optimum temperature), it
also shows a logarithmic increase. In addition, the activation energy
for i-PS is found to be the lowest, while the activation energies
of a-PS and s-PS have the same values within experimental error. The
results reported in this paper indicate that non-α-processes
are relevant for the low-temperature behavior of the glass.