Catheter entrapment in Chiari network: Extraction with radiofrequency.

Marina Vieira Nagahama1, Marcel Henrique Sakai2, Maria Cecília Xavier Souto3, Elano Sousa da Frota4, Claudio Cirenza5, Angelo Amato Vincenzo de Paola6.   

Abstract

We report a challenging case of a duodecapolar mapping catheter entrapment in Chiari network and its release by radiofrequency energy application with an ablation catheter.
Copyright © 2019 Indian Heart Rhythm Society. Production and hosting by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Ablation; Catheter; Chiari network; Entrapment; Radiofrequency

Year:  2019        PMID: 31419491      PMCID: PMC6823705          DOI: 10.1016/j.ipej.2019.08.001

Source DB:  PubMed          Journal:  Indian Pacing Electrophysiol J        ISSN: 0972-6292


Case report

We describe the case of a 42-year-old male patient with atrial flutter who was referred for ablation. While placing duodecapolar mapping catheter around the tricuspid valve annulus (7 French, type Halo, Biosense Webster Inc.), its tip got trapped at the right atrium junction with inferior vena cava, in the Chiari network region (Fig. 1A). Transthoracic echocardiography confirmed the diagnosis. Initially we attempted traction and rotation movements of the catheter without success. The patient presented severe right hypochondrium pain during traction of the catheter. We introduced a sheath (Agilis, St. Jude Medical, 8Fr) over the catheter to aid removal, also unsuccessfully; contrast injection through sheath was realized to confirm location (Fig. 1B). Then we performed radiofrequency energy application at entrapment site (Fig. 1C) with a therapeutic catheter (7 French, curve D, tip 4mm, Biosense Webster Inc.) for 10 seconds, power of 30 watts, temperature of 50 °C, with catheter release (Fig. 1D). Echocardiogram after procedure did not demonstrate complications.
Fig. 1

Fluoroscopic images (30° left anterior oblique view). a) Halo catheter tip trapped in the Chiari network region (arrow). b) Introduction of sheath over Halo catheter and contrast injection. c) Radiofrequency application. d) Removal of catheter through sheath.

Fluoroscopic images (30° left anterior oblique view). a) Halo catheter tip trapped in the Chiari network region (arrow). b) Introduction of sheath over Halo catheter and contrast injection. c) Radiofrequency application. d) Removal of catheter through sheath. Chiari network is considered as embryologic remnants of right valve of the sinus venosus and septum spurium, consisting of fenestrated membranes at the junction of the right atrium and venae cavae, with a prevalence of approximately 2%. In 1897, Hans Chiari described the intricate fenestrated reticulum containing multiple threads inserting on the anterior surface of Eustachian valve, the posterior wall of the ostial inferior vena cava, crista terminalis, and the intervenous tubercle. Many reports have since been published concerning Chiari network, describing its attachment to multiple intracardiac devices [1]. Literature reports show use of traction and rotation movements, laser sheath, snare catheter, endomyocardial biopsy forceps, radiofrequency energy delivery and cardiac surgery for release. Shimoike et al. related the release of a guide wire entangled by Chiari network with 4 seconds of radiofrequency delivery in the proximal end of the guide wire [2,3]. To the best of our knowledge, our report is the first case of radiofrequency catheter ablation to release another catheter trapped in the Chiari network. We believe that cautious radiofrequency delivery is an efficient and safe approach to release devices from Chiari network.

Conflicts of interest

None declared.
  2 in total

1.  Entrapment of a guide wire by the Chiari network in a patient with ablated idiopathic ventricular tachycardia.

Authors:  E Shimoike; N Ueda; T Maruyama; Y Kaji; Y Niho
Journal:  J Interv Card Electrophysiol       Date:  2001-06       Impact factor: 1.900

2.  Accidental Entrapment of Electrical Mapping Catheter by Chiari's Network in Right Atrium during Catheter Ablation Procedure.

Authors:  Atsushi Sakamoto; Tsuyoshi Urushida; Tomoaki Sakakibara; Makoto Sano; Kenichiro Suwa; Takeji Saitoh; Masao Saotome; Hideki Katoh; Hiroshi Satoh; Hideharu Hayashi
Journal:  Case Rep Cardiol       Date:  2016-06-05
  2 in total
  1 in total

1.  Catheter entrapment in the Chiari network during catheter ablation.

Authors:  Luis Quininir; Peter P Luk; Mark A McGuire
Journal:  HeartRhythm Case Rep       Date:  2020-08-21
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.