| Literature DB >> 31412774 |
Manik Chopra1, Yam-Hong Wong2,3, Lars Søndergaard1, Ole De Backer1.
Abstract
BACKGROUND: Left atrial appendage closure is increasingly used. LAA closure procedures can rarely be complicated by device embolization, typically towards the aorta. In this case report, we describe the successful percutaneous retrieval of an embolized Amulet device from the left atrium in a patient with previous MitraClips. CASEEntities:
Keywords: Device embolization; LAAC; Left atrial appendage closure
Mesh:
Year: 2019 PMID: 31412774 PMCID: PMC6693090 DOI: 10.1186/s12872-019-1170-8
Source DB: PubMed Journal: BMC Cardiovasc Disord ISSN: 1471-2261 Impact factor: 2.298
Fig. 1a-b CT measurements of the usual landing zone, and c-d CT measurements of a more distal alternative landing zone. e Amulet LA closure device released with signs of adequate stability and minimal leakage. CT, computed tomography
Fig. 2a Embolized Amulet device whirling in the left atrium. b Approach with two transseptal accesses for stabilization and retrieval of the embolized Amulet device, respectively. c Amulet device elongated and stabilized in the inferior vena cava before final retrieval into the MitraClip steerable guiding catheter and out of the patient. d Minimal shunt over the interatrial septum after retrieval of the embolized Amulet device