Head-to-head comparison of a CZT-based all-purpose SPECT camera and a dedicated CZT cardiac device for myocardial perfusion and functional analysis.

Alessia Gimelli1, Riccardo Liga2, Michelangelo Bertasi3, Annette Kusch3, Paolo Marzullo3,4.   

Abstract

PURPOSE: To compare the outputs of a novel all-purpose SPECT camera equipped with CZT detectors (Discovery NM/CT 670) with the state-of-the-art represented by a dedicated CZT (Alcyone, Discovery 530c) cardiac camera in patients submitted to myocardial perfusion imaging (MPI).
METHODS: We included 19 patients that underwent sequential low-dose 99mTc-tetrofosmin (148-185 MBq during stress and 296-370 MBq at rest) MPI with Alcyone and Discovery 670 cameras. Quantitative (% tracer's uptake) and semi-quantitative analyses of perfusion data were performed for each scan. Moreover, major left ventricular (LV) functional and structural parameters were derived from each camera and compared.
RESULTS: The two cameras showed excellent correlation for segmental myocardial % uptake at stress (R = 0.90; P < 0.001) and at rest (R = 0.88; P < 0.001) with narrow Bland-Altman limits of agreement. The level of diagnostic agreement of Discovery 670 and Alcyone cameras regarding perfusion analysis was excellent (Cohen's κ 0.85). Similarly, the two cameras showed excellent correlation in the evaluation of LV ejection fraction (R = 0.95), peak filling rate (R = 0.97), and mass (R = 0.98).
CONCLUSIONS: Our preliminary results suggest that MPI with an all-purpose Discovery 670 CZT-SPECT camera is feasible, comparing well with the current state-of-the-art technology.
© 2019. American Society of Nuclear Cardiology.

Entities:  

Keywords:  Gated SPECT; MPI; SPECT; instrumentation

Mesh:

Substances:

Year:  2019        PMID: 31385223     DOI: 10.1007/s12350-019-01835-4

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  2 in total

1.  First comparison of performances between the new whole-body cadmium-zinc-telluride SPECT-CT camera and a dedicated cardiac CZT camera for myocardial perfusion imaging: Analysis of phantom and patients.

Authors:  Maxime Morelle; Dimitri Bellevre; Claude Hossein-Foucher; Alain Manrique; Alban Bailliez
Journal:  J Nucl Cardiol       Date:  2019-04-08       Impact factor: 5.952

2.  Automatic evaluation of myocardial perfusion on SPECT: Need for "Normality".

Authors:  Riccardo Liga; Alessia Gimelli
Journal:  J Nucl Cardiol       Date:  2017-10-25       Impact factor: 5.952

  2 in total
  1 in total

1.  Myocardial blood flow reserve and absolute myocardial blood flow for the assessment of patients with coronary artery disease with or without microvascular dysfunction.

Authors:  Raffaele Giubbini; Elisabetta Cerudelli; Luca Camoni
Journal:  J Nucl Cardiol       Date:  2020-08-04       Impact factor: 5.952

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.