Pulmonary versus aortic pressure behavior of a bovine pericardial valve.

Hanna Pragt1, Joost P van Melle2, Gijsbertus J Verkerke3, Massimo A Mariani4, Tjark Ebels5.   

Abstract

BACKGROUND: The Carpentier Edwards Perimount Magna Ease aortic valvular prosthesis (Edwards Lifesciences, Irvine, Calif) has been among the most frequently and successfully used tissue prosthetic cardiac valves. Furthermore, this prosthesis has been used off-label in the pulmonary position. Until now, there has been a paucity of data regarding the functioning of tissue prosthetic valves under pulmonary conditions.
METHODS: Using a pulse duplicator, hydrodynamic characteristics of a 21-mm and 25-mm Magna Ease valve were evaluated. Among parameters evaluated were leakage orifice area, closing time (ie, time required to close), and leakage duration. This procedure was performed under different pulmonic pressure conditions (15/5 mm Hg, 28/11 mm Hg, 73/32 mm Hg) and normal aortic pressure (120/80 mm Hg) as a reference. Moving images were obtained using a Phantom MIRO M320S high-speed camera (Vision Research Inc, Wayne, NJ) at 600 frames per second and used to analyze valve area in closed position.
RESULTS: Under normal pulmonic conditions (28/11 mm Hg) the leakage orifice area was 0.020 ± 0.012 mm2 for the 21-mm valve and 0.054 ± 0.041 mm2 for the 25-mm valve (P = .03). Hydrodynamic characteristics of the valves differed between pulmonary and aortic testing condition. Valve closing volumes were significantly lower under pulmonary hypotension and normal pulmonary conditions than under normal aortic conditions (P < .05).
CONCLUSIONS: Under normal pulmonary pressure conditions, the hydrodynamic characteristics of Magna Ease valves are significantly different compared with aortic conditions. Further research is needed to determine whether these results are associated with prosthetic valve failure.
Copyright © 2019 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Pulomnary valve replacement; prosthetic heart valve

Year:  2019        PMID: 31383558     DOI: 10.1016/j.jtcvs.2019.05.084

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  2 in total

1.  Beyond the Valve: Lifelong Management of Right Ventricular Outflow Tract Lesion in Adult Congenital Heart Disease.

Authors:  Chang-Ha Lee
Journal:  Korean Circ J       Date:  2022-08       Impact factor: 3.101

2.  A Bicentric Propensity Matched Analysis of 158 Patients Comparing Porcine Versus Bovine Stented Bioprosthetic Valves in Pulmonary Position.

Authors:  Bunty Ramchandani; Raúl Sánchez; Juvenal Rey; Luz Polo; Álvaro Gonzalez; Maria-Jesús Lamas; Tomasa Centella; Jesús Díez; Ángel Aroca
Journal:  Korean Circ J       Date:  2022-04-21       Impact factor: 3.101

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.