Literature DB >> 31370617

Compensation of array lens effects for improved co-registration of passive acoustic mapping and B-mode images for cavitation monitoring.

Michael D Gray1, Constantin C Coussios1.   

Abstract

Passive acoustic mapping (PAM) techniques offer a simple means of spatio-temporal cavitation monitoring during therapeutic ultrasound procedures. Implementation with a conventional diagnostic ultrasound system allows natural integration of PAM with B-mode imaging. However, the refracting properties of diagnostic array lenses may introduce PAM image registration errors that could lead to inaccuracies in treatment monitoring and guidance. To address these concerns, this paper presents lens characterization of two different array designs, analytical estimation of lens-induced source mapping errors in simple media, and experimental demonstration and correction of lens effects, reducing the depth-averaged image co-registration errors to no more than 0.52 mm.

Year:  2019        PMID: 31370617     DOI: 10.1121/1.5118238

Source DB:  PubMed          Journal:  J Acoust Soc Am        ISSN: 0001-4966            Impact factor:   1.840


  1 in total

1.  Passive Cavitation Mapping by Cavitation Source Localization From Aperture-Domain Signals-Part I: Theory and Validation Through Simulations.

Authors:  Arsenii V Telichko; Taehwa Lee; Marko Jakovljevic; Jeremy J Dahl
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2021-03-26       Impact factor: 2.725

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.