Literature DB >> 3136967

Measurement of oxygen and carbon dioxide partial pressures in synovial fluid after tonometry.

P Geborek1, B Lindoff, S O Valind.   

Abstract

A commercially available gas analyser was used to measure tension of oxygen (PO2) and carbon dioxide (PCO2) in synovial fluid samples after tonometry. Measured values of PCO2 were close to the expected (median difference 0.2 kPa, range -0.4 to 0.4) within the analysed concentration range of 4-10 kPa. No consistent difference between measured and expected values of PO2 were found for oxygen in the range 3-11 kPa (median difference 0.1 kPa, range -0.3 to 1.2). For oxygen tensions below 3 kPa, however, the measured values invariably overestimated the actual PO2, the errors ranging from 0.3 to 1.9 kPa, median 1.1. The importance of proper handling of samples was investigated and storage for 1 h at 0 degrees C in plastic syringes resulted in elevation of the PO2 levels measured (range of elevation 0.2 to 3.6 kPa, median 1.15), whilst no significant differences were found when stored in glass syringes. Within the limits stated, commercially available gas analysers may thus be used to investigate these parameters related to local tissue metabolism in effusive joint conditions.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3136967     DOI: 10.1111/j.1475-097x.1988.tb00286.x

Source DB:  PubMed          Journal:  Clin Physiol        ISSN: 0144-5979


  2 in total

1.  Direct assessment of synovial blood flow and its relation to induced hydrostatic pressure changes.

Authors:  P Geborek; K Forslind; F A Wollheim
Journal:  Ann Rheum Dis       Date:  1989-04       Impact factor: 19.103

2.  Factors influencing the acid-base changes in the air-pouch exudate following carrageenan induced inflammation in rats.

Authors:  V Alfaro; J Ródenas; J Pesquero; M T Mitjavila; L Palacios; T Carbonell
Journal:  Inflamm Res       Date:  1996-08       Impact factor: 4.575

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.