Literature DB >> 3136465

Skin dose near compensating filters in radiotherapy.

S J Thomas1, G Bruce.   

Abstract

In radiotherapy treatments with MV beams, the use of tissue compensators affects the dose to the skin. Methods of calculating the relative skin dose (RSD) are described and a formula is derived to predict the contribution of a tissue compensator to the RSD. Measurements of RSD for various field sizes and distances from the compensator are presented.

Entities:  

Mesh:

Year:  1988        PMID: 3136465     DOI: 10.1088/0031-9155/33/6/004

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

1.  Comparing dose in the build-up region between compensator- and MLC-based IMRT.

Authors:  Khosrow Javedan; Geoffrey G Zhang; Sarah Hoffe; Vladimir Feygelman; Kenneth Forster
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102