Claudio Reverberi1, Granit Rabia2, Fabrizio De Rosa2, Davide Bosi2, Andrea Botti2, Giorgio Benatti2.
Abstract
BACKGROUND: The RITMIA™ app (Heart Sentinel™, Parma, Italy) is a novel application that combined with a wearable consumer-grade chest-strap Bluetooth heart rate monitor, provides automated detection of atrial fibrillation (AF), and may be promising for sustainable AF screening programs, since it is known that prolonged monitoring leads to increased AF diagnosis.Entities:
Mesh:
Year: 2019 PMID: 31355264 PMCID: PMC6634013 DOI: 10.1155/2019/4861951
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Flowchart of the study design. Figure also shows system components and in particular an example of a chest-belt HR monitor sensor similar to the ones used in the study and an example of RITMIA™ App interface. AF: atrial fibrillation, SR: sinus rhythm.
Figure 2Examples of 12-lead ECG (left side) with the cloud-based interface of the RITMIA™ App showing the paired recording (right side). RR intervals are plotted on a graph showing RR interval on the y-axis and time on the x-axis. A colour-code is automatically attributed to each dot, corresponding to the automated diagnosis as “probable AF”=red, “unclassified non-AF arrhythmia”=yellow and “normal rhythm”=blue. (a) pre-ECV 12-lead ECG showing atrial fibrillation and paired App recording automatically labelled as “probable AF”. (b) post-ECV 12-lead ECG showing restoration of sinus rhythm and paired App recording automatically labelled as “normal rhythm” and “unclassified non-AF arrhythmia”. ECV: electrical cardioversion. AF: atrial fibrillation.
Study Population Characteristics.
| STUDY POPULATION CHARACTERISTICS (N=95) | |
|---|---|
| Age (years) | 66.2±10.7 |
| Male/Female | 75/20 |
| AF | 87 (91.6%) |
| AFL | 8 (8.4%) |
| CHA2DS2-VASC score | 2.3±1.5 |
| Preprocedural TEE | 59 (62.1%) |
|
| |
| ANTICOAGULANT | |
|
| |
| Warfarin | 32 (33.7%) |
| Dabigatran | 18 (18.9%) |
| Rivaroxaban | 22 (23.2%) |
| Apixaban | 20 (21.1%) |
| Edoxaban | 3 (3.2%) |
|
| |
| Successful ECV | 83 (87.4%) |
Study population characteristics. AF=atrial fibrillation; AFL=atrial flutter; TEE=transesophageal echocardiography; ECV=electrical cardioversion.
Ritmia Automated Diagnosis Compared To 12-Lead ECG Diagnosis.
| 12-Lead ECG | APP AF | APP NO-AF | Total |
|---|---|---|---|
| AF | 96 | 3 | 99 |
| Flutter | 0 | 8 | 8 |
| SR | 4 | 79 | 83 |
| Total | 100 | 90 |
|
Contingency table showing comparison between 12-lead ECG diagnosis and RITMIA™ automated diagnosis.