Matthieu Vanhoutte1, Franck Semah, Renaud Lopes, Alice Jaillard, Grégory Petyt, Anne-Laure Aziz, Hélène Lahousse, Jérôme Declerck, Florence Pasquier, Bruce Spottiswoode, Rachid Fahmi.
Abstract
This study aims at assessing whether EANM harmonisation strategy combined with EQ·PET methodology could be successfully applied to harmonize brain 2-deoxy-2[18F]fluoro-D-glucose ([18F]FDG) positron emission tomography (PET) images. The NEMA NU 2 body phantom was prepared according to the EANM guidelines with an [18F]FDG solution. Raw PET phantom data were reconstructed with three different reconstruction protocols frequently used in clinical PET brain imaging: ([Formula: see text]) Ordered subset expectation maximization (OSEM) 3D with time of flight (TOF), 2 iterations and 21 subsets; ([Formula: see text]) OSEM 3D with TOF, 6 iterations and 21 subsets; and ([Formula: see text]) OSEM 3D with TOF, point spread function (PSF), and 8 iterations and 21 subsets. EQ·PET filters were computed as the Gaussian smoothing that best independently aligned the recovery coefficients (RCs) of reconstructions [Formula: see text] and [Formula: see text] with the RCs of the reference reconstruction, [Formula: see text]. The performance of the EQ·PET filter to reduce variations in quantification due to differences in reconstruction was investigated using clinical PET brain images of 35 early-onset Alzheimer's disease (EOAD) patients. Qualitative assessments and multiple quantitative metrics on the cortical surface at different scale levels with or without partial volume effect correction were evaluated on the [18F]FDG brain data before and after application of the EQ·PET filter. The EQ·PET methodology succeeded in finding the optimal smoothing that minimised root-mean-square error (RMSE) calculated using human brain [18F]FDG-PET datasets of EOAD patients, providing harmonized comparisons in the neurological context. Performance was superior for TOF than for TOF + PSF reconstructions. Results showed the capability of the EQ·PET methodology to minimize reconstruction-induced variabilities between brain [18F]FDG-PET images. However, moderate variabilities remained after harmonizing PSF reconstructions with standard non-PSF OSEM reconstructions, suggesting that precautions should be taken when using PSF modelling.Entities:
Mesh:
Substances:
Year: 2019 PMID: 31344691 DOI: 10.1088/1361-6560/ab35b4
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609