Literature DB >> 31344248

First Simultaneous Explantation and Re-Implantation of an Argus II Retinal Prosthesis System.

Anna-Maria Seuthe, Arno Haus, Kai Januschowski, Peter Szurman.   

Abstract

A 59-year-old blind woman with retinitis pigmentosa had undergone implantation of an Argus II system without complications and took advantage of the system for 7 months. Due to a technical error of the system, the glasses of the prosthesis developed increased heat generation, making the use of the system unbearable. Because of the initial satisfaction and high motivation of the patient, it was decided to strive for an implant exchange. In a 4-hour surgery session, the implant was removed and a new system was implanted. Hereby, several critical surgery steps could be identified that are important to consider for future surgeries. Besides a strong fibrotic scarring of conjunctiva and tenon aggravating the extraction of the extraocular components, the preparation of the sclerotomy proved to be difficult for achieving water tightness. Furthermore, the extraction of the retinal tack and the repositioning of the new array were challenging steps. During the postoperative course, a severe hemorrhagic choroidal detachment occurred but resolved without surgical intervention. The implant was well-placed and showed good function. The exchange of an Argus II is feasible, however, with the challenge of hypotony and hemorrhage as well as possible fibrosis complicating the repositioning of the chip. [Ophthalmic Surg Lasers Imaging Retina. 2019;50:462-465.]. Copyright 2019, SLACK Incorporated.

Entities:  

Mesh:

Year:  2019        PMID: 31344248     DOI: 10.3928/23258160-20190703-10

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging Retina        ISSN: 2325-8160            Impact factor:   1.300


  3 in total

Review 1.  Clinical Progress and Optimization of Information Processing in Artificial Visual Prostheses.

Authors:  Jing Wang; Rongfeng Zhao; Peitong Li; Zhiqiang Fang; Qianqian Li; Yanling Han; Ruyan Zhou; Yun Zhang
Journal:  Sensors (Basel)       Date:  2022-08-30       Impact factor: 3.847

2.  Optoelectronic Devices for Vision Restoration.

Authors:  Victor Wang; Ajay E Kuriyan
Journal:  Curr Ophthalmol Rep       Date:  2020-04-20

3.  INTRAOPERATIVE OPTICAL COHERENCE TOMOGRAPHY AND ENDOSCOPY-GUIDED EXPLANTATION OF ARGUS II DEVICE.

Authors:  Cindy X Cai; Lejla Vajzovic
Journal:  Retin Cases Brief Rep       Date:  2020-03-17
  3 in total