| Literature DB >> 31339492 |
Benoit Nespola1, Hélène Comitogianni1, Isabelle Jahn1, Joëlle Goetz1.
Abstract
The complement system is composed of a set of plasma or membrane proteins. Complement protein deficiencies can be inherited or acquired, through the presence of autoantibodies or through consumption. We evaluated the analytical performance of the Optilite® analyser for the determination of the C3 and C4 levels and for the evaluation of the total complement activity. The intra- and inter-series CVs were evaluated and have showed satisfactory results, the concordances with analysers currently used in the laboratory (BNII® and BCT®, Siemens) are very good, as is the agreement between the serum and plasma samples. We also determined the reference values for the different parameters tested in view of a routine use of Optilite® analyser in the laboratory.Entities:
Keywords: C3; C4; Optilite® analyser; analytical performances; total complement activity
Mesh:
Substances:
Year: 2019 PMID: 31339492 DOI: 10.1684/abc.2019.1453
Source DB: PubMed Journal: Ann Biol Clin (Paris) ISSN: 0003-3898 Impact factor: 0.459