| Literature DB >> 31332108 |
Suneeta Dubey1, Kanika Jain2, T Nirmal Fredrick3.
Abstract
Quality assurance (QA) is the maintenance of a desired level of quality in a service, by means of attention to every stage of process of delivery. Correct image acquisition along with accurate and reproducible quantification of ophthalmic imaging is crucial for evaluating disease progression/stabilization, response to therapy, and planning proper management of these cases. QA includes development of standard operating procedures for the collection of data for ophthalmic imaging, proper functioning of the ophthalmic imaging equipment, and intensive training of technicians/doctors for the same. QA can be obtained during ophthalmic imaging by not only calibration and setting up of the instrument as per the manufacturer's specifications but also giving proper instructions to the patients in a language which they understand and by acquisition of good quality images. This review article will highlight on how to achieve QA in imaging which is commonly being used in ophthalmic practice.Entities:
Keywords: B-scan; Pentacam; ophthalmic imaging; optical coherence tomography; quality; quality assurance; slit-lamp photography
Mesh:
Year: 2019 PMID: 31332108 PMCID: PMC6677073 DOI: 10.4103/ijo.IJO_1959_18
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Common work instructions to be followed while acquiring images using imaging procedures
| Ensure calibration of the imaging units is as per the specification by the manufacturer. If any problem occurs with the equipment, notify the chief technician.[ |
| Ensure and encourage thorough reading and understanding of the operating manuals of the imaging units by technicians/doctors. They should be well versed with the machine and its operation. |
| Ensure maintenance of the front objective lens to prevent any imaging artifacts due to debris/scratches/fingerprints/eyelash/nose prints on the lens. |
| The measures to maintain the front objective lens are as follows: |
| Inspect lens with a penlight and an air bulb |
| Use brush to remove any debris |
| Clean the lens from the center outward using lens wipes wrapped around a cotton tipped applicator dampened with lens cleaner (diluted acetone or lens cleaning solution). Wipe the lens with one pass in one direction. Discard the used lens wipes. |
| Clean the lens daily or as per requirement even if required multiple number of times per day |
| Do not touch lens by hands/nose/forehead of the patient/technician/doctor[ |
| Cover imaging units by dust covers which are to be removed and replaced at the end of each session[ |
| Keep the lens cover for the objective lens in place until just prior to each examination[ |
| Position the patient comfortably by adequate adjustment of forehead and chin rests. Chin rest should be adjusted so that the eyes are approximately level with the height adjustment mark on the face rest pole. |
| Take assistant’s help with head fixation and support in cognitively impaired and physically disabled patients. Test can be performed with the patient in the wheelchair for physically disabled patients.[ |
| Ensure if the procedure is being performed on the correct patient and correct site by cross checking the file details with the patient/guardian |
| Ensure that the patient/guardian understands the procedure and signs on the consent form (if required as in before performing FFA) |
| Enter the data into the system in a prescribed format |
| Enter the details of the patient, procedure, result, and time are in the procedure register |
| Carry out the test as per SOP) |
| Undertake aseptic measures to prevent cross infection |
| Dispose biomedical waste properly as per Biomedical Waste disposal Guidelines (BMW) |
| Give proper instructions to the patients about the procedure in a language which they understand to increase their cooperation |
| Acquire images when the eyes are most still for the longest increment of time. In nystagmus patients, time acquisition of images to a null point in nystagmus can patch the fellow eye to improve patient fixation and reduce distraction in imaging techniques which acquire images of one eye at a time.[ |
| Instruct the patient to blink once/twice just before the image is taken to ensure a moist cornea and to safeguard unwanted blinks at the moment of image acquisition. Lids/lashes can be held open to prevent artifacts. Use lubricating eye drops in patients with dry eyes.[ |
| Store the images and data in a retrievable format |
| Ensure that the reported results are complete in all aspects including description, clinical correlation, probable diagnosis, laterality, date, and time. The report is to be signed by the technician and doctor on duty after verification. |
FFA=Fundus fluorescein angiography; SOP=Standard operating procedure
Checklist for ascertaining quality for ophthalmic images
| Question | Answer |
|---|---|
| Is the calibration of the imaging unit up to date? | Yes/No |
| Is the technician/doctor thoroughly acquainted with the operating manual of the imaging unit? | Yes/No |
| Is the procedure properly explained to the patient/patient’s guardian in a language that he/she understands? | Yes/No |
| Is the consent taken? | Yes/No |
| Is the patient comfortably positioned for imaging? | Yes/No |
| Is the patient’s identity and intended procedure checked? | Yes/No |
| Are proper details of the patient entered in the software? | Yes/No |
| Is the laterality of the eye checked? | Yes/No |
| Is the test being carried out as per the SOP? | Yes/No |
| Is the report duly signed with the results by the technician/treating ophthalmologist? | Yes/No |
SOP=Standard operating procedure
Work instructions for ophthalmic images acquired by fundus camera
| Assure proper date and time on the imaging printouts. Change the date manually if the clock has failed or if the camera has been left unplugged for a long period of time. |
| Enter the laterality of the eye which is to be tested correctly |
| Darken the room to improve contrast and to allow pupils to dilate physiologically[ |
| Dilate the pupil in cases with media opacity and when peripheral fundus photographs are to be recorded. Photography through small (<4 mm) pupils will be difficult because some of the camera light does not enter the smaller pupil thus resulting in uneven illumination seen as dark shadows on the monitor.[ |
| Acquire multiple images until sharper, clearer images are acquired. |
| Take the images when centered on disc, macula, and peripheral images of any peripheral pathology |
| Ensure alignment, focus, and proper fixation of fundus photographs (there are fine and coarse adjustment knobs, coarse adjustment brings the image in focus, whereas fine adjustment fin focuses the image). Constantly adjust and position the camera to maintain alignment and fixation which may be required multiple times during the examination of a single patient.[ |
| Unacceptable images which warrant imaging to be repeated can be |
| Overexposed images (too light) |
| Underexposed images (too dark) |
| Obvious shadow on macula/disc |
| Shadows covering over 40% of the image |
| Out of focus/blurred image |
| Incorrect field of fundus which has been acquired |
| Image captured during blink |
| Incomplete acquisition of pathology |
| Green/white/partial halo[ |
| Various modes are available in fundus camera for photography: |
| Color, where the retina is illuminated by white light and examined in full color. |
| Red-free fundus photography uses a green filter of 540-570 nm to block out red wavelengths of light. This allows a better contrast for viewing retinal blood vessels and associated hemorrhages, pale lesions such as drusen and exudates, and subtle characteristics such as nerve fiber layer defects and epiretinal membranes. |
| Angiography records vascular flow within the retina and surrounding tissue by injecting a fluorescent dye into the blood stream (fundus fluorescein angiography and indocyanine green angiography) |
Figure 1Montage of the fundus of right eye acquired by a fundus camera after using all standard work instructions as highlighted in the text
Work instructions for ophthalmic images acquired by OCT
| Enter the correct date of birth of the patient |
| Select scan protocol or pattern as per the required image/area which is to be examined[ |
| Ensure proper alignment of the pupil with the center of scan. The centering line is helpful to center the scan over fovea, optic nerve head, or any other area of interest.[ |
| Select dark or light according to patient’s iris color to optimize illumination, brightness, and contrast settings. If neither eye color is selected, then the default protocol which is the average of the respective values of illumination, brightness, and contrast for the dark and light eye is selected.[ |
| Ensure polarization adjustment (P motor) to optimize the OCT image signal strength which results in a clearer image. It must be done before and repeated during a scanning session.[ |
| Adjust OCT image noise for best visualization of retinal tissue. For the line and cross line scan, select the option of averaging multiple scans to achieve final averaged image. To obtain the final averaged image in either the line or cross line scans, move the rectangle on or over the region of interest.[ |
| Ensure optimum placement of the scan between the red dashed lines. If the lines are horizontal, the scan should be in the upper part of the target area (3 diopter disc area will have some of the scan image which will fall below the lower line which is acceptable).[ |
| Ensure an even signal strength across the entire scan with no section of any weak signal (signal strength assessment tool along with their recommended values is enumerated in Table 5) |
| Minimize missed or dropped scans which are due to blinking/weak signals/eyelash obstruction/iris clipping. The missed or dropped scans are indicated in gray instead of white scan indicator lines. |
| More missing scans equate to more interpolation in the map values. These scans should be repeated.[ |
| Select the option of follow-up if this visit is a follow-up scan to allow the comparison in between the two visits[ |
OCT=Optical coherence tomography
Figure 2Optical coherence tomography (OCT) of the patient acquired after using all standard work instructions as highlighted in the text showing good signal strength (red box)
Signal strength assessment tools in various machines of OCT
| OCT machine | Cirrus | RTvue | Spectralis | 3D-OCT 1000 |
|---|---|---|---|---|
| Signal strength assessment tool | SS | SSI | Quality | Image quality metric |
| Range of indicator | 0-10 | 0-100 | 0-40 | 0-100 |
| Recommended threshold value | 6 | 39 | 15 | 45 |
OCT=Optical coherence tomography; SS=Signal strength; SSI=Signal strength index. Signal strength assessment score is based on the total amount of the retinal signal received by OCT
Work instructions for ophthalmic images acquired by FFA
| Ensure adequate pupillary dilatation before starting procedure |
| Ensure proper positioning of illumination beam within pupil using joystick |
| Focus on the area of interest and maintain fixation by repeated and precise instructions to patients |
| Acquire good quality control images of both eyes before injection of fluorescein dye |
| Follow preplanned photographic sequence to prevent any missing of relevant photographs in between |
| Ensure adequate illumination of photographs to prevent too bright/too dim image quality |
FFA=Fundus fluorescein angiography
Figure 3Fundus fluoroscein angiography (FFA) of the patient acquired after using all standard work instructions as highlighted in the text
Work instructions for ophthalmic images acquired by Pentacam
| Ensure that QS, which specifies the quality of the topographic capture, is displayed as “OK.” If not done, the software tends to extrapolate the missing information[ | |
| Check corneal form factor asphericity (Q) | |
| Q <0 | Untreated prolate cornea |
| Q >1 | Treated oblate cornea (post LASIK/PRK/PK) |
| Q=0 | Spherical cornea |
| Normal cornea is prolate and has a Q-value of −0.26[ | |
| Avoid the following artifacts: | |
| Shadows on cornea from long eye lashes/trichiatic lashes | |
| Ptosis or insufficient eye opening | |
| Irregularities of tear film (dry eye/mucinous/greasy film) | |
| Incomplete/distorted image (due to corneal pathology)[ | |
| Ensure proper centering of the scan. Can use automatic release option which decreases the chances of the patient blinking during the scan.[ | |
| No eye drops should be applied to the patient’s eye prior to examination which may interfere with tear film and affect the measurements in corneal topography scans[ | |
QS=Quality specification
Figure 4Pentacam of the patient acquired after using all standard work instructions as highlighted in the text showing quality specification (QS) to be OK (red box)
The area of the retina screened on the basis of the position of the B-scan probe
| Clock hour - probe position | Clock hour - area screened |
|---|---|
| 3-Limbus | 9-Posterior |
| 3-Equator | 9-Equator |
| 3-Fornix | 9-Anterior |
| 6-Limbus | 12-Posterior |
| 6-Equator | 12-Equator |
| 6-Fornix | 12-Anterior |
Work instructions for ophthalmic images acquired by ultrasound B-scan
| Ensure lesions are placed in the center of the scanning beam |
| Instruct the patient to fix the gaze so that the probe or beam is perpendicular to the interfaces at the area of interest |
| Ensure the lowest possible decibel gain should be used to optimize the resolution of images[ |
| Higher gain increases the sensitivity of the instrument in displaying weak echoes such as vitreous opacities. |
| Lower gain only allows display of strong echoes such as retina and sclera |
Figure 5Ultrasound B-scan along with A-scan (below) of the patient acquired after using all standard work instructions as highlighted in the text