Literature DB >> 31328560

Squeaking and other noises in patients with ceramic-on-ceramic total hip arthroplasty.

Fabio Baruffaldi1, Rosaria Mecca1, Susanna Stea1, Alina Beraudi1, Barbara Bordini1, Marilina Amabile1, Alessandra Sudanese2, Aldo Toni2.   

Abstract

INTRODUCTION: Noise in ceramic-on-ceramic (CoC) total hip arthroplasty (THA) is a potential symptom of abnormal bearing wear. Squeaking and other prosthetic hip noises are multi-factorial phenomena that can be analysed and may provide prognostic information.
METHODS: 46 patients with noisy CoC bearings were investigated using X-ray, computed tomography and joint fluid analysis, and classified into either high or low risk of ceramic liner fracture groups according to previously published guidelines. Noise events from the bearings of 16 high risk cases which were subsequently revised were compared with 30 patients in the low risk control group who did not undergo revision. Noise events were analysed for their physical characteristics using a standardised protocol and classified as either low frequency and short duration 'clicking' or long duration and high frequency 'squeaking'.
RESULTS: The peak frequency of squeaking during forward walking was significantly higher for patients in the case group who were revised, compared with the control group. The patient-reported onset of squeaking (46 months postoperatively) was earlier than short-noise emissions (82 months). In the standardised sequence of movements, short-noise always occurred more frequently than squeaking. Small heads (28 mm) were more likely to develop short-noise, while large heads (⩾32 mm) were more likely to develop squeaking. DISCUSSION: Noise evaluation may provide additional value for predicting failure of CoC bearings, though some questions should be better investigated in a dedicated prospective trial.

Entities:  

Keywords:  Ceramic-on-ceramic bearing; noise; squeaking; total hip arthroplasty

Mesh:

Year:  2019        PMID: 31328560     DOI: 10.1177/1120700019864233

Source DB:  PubMed          Journal:  Hip Int        ISSN: 1120-7000            Impact factor:   2.135


  1 in total

1.  Arthroscopic evaluation and treatment of a squeaking hip. A case report.

Authors:  Jonathan Bellity; Marc Elkaïm; Didier Hannouche; Rémy Nizard
Journal:  BMC Musculoskelet Disord       Date:  2020-12-03       Impact factor: 2.362

  1 in total