Wilfred R Hagen1. 1. Department of Biotechnology , Delft University of Technology , Van der Maasweg 9 , 2629HZ Delft , The Netherlands.
Abstract
Analysis of the electron paramagnetic resonance (EPR) of transition ion complexes requires data taken at different microwave frequencies because the spin Hamiltonian contains operators linear in the frequency as well as operators independent of the frequency. In practice, data collection is hampered by the fact that conventional EPR spectrometers have always been designed to operate at a single frequency. Here, a broadband instrument is described and tested that operates from 0.5 to 12 GHz and whose sensitivity approaches that of single-frequency spectrometers. Multifrequency EPR from triclinic substitutional (0.5%) Cu(II) in ZnSO4 is globally analyzed to illustrate a novel approach to reliable determination of the molecular electronic structure of transition ion complexes from field-frequency 2D data sets.
Analysis of the electron paramagnetic resonance (EPR) of transition ion complexes requires data taken at different microwave frequencies because the spin Hamiltonian contains operators linear in the frequency as well as operators independent of the frequency. In practice, data collection is hampered by the fact that conventional EPR spectrometers have always been designed to operate at a single frequency. Here, a broadband instrument is described and tested that operates from 0.5 to 12 GHz and whose sensitivity approaches that of single-frequency spectrometers. Multifrequency EPR from triclinic substitutional (0.5%) Cu(II) in ZnSO4 is globally analyzed to illustrate a novel approach to reliable determination of the molecular electronic structure of transition ion complexes from field-frequency 2D data sets.
The technique of electron
paramagnetic resonance (EPR) is of considerable value for the determination
of structural and electronic properties of transition-metal complexes.
Ever since 1947[1] until this day the vast
majority of studies have been done with microwaves at a single frequency
in the X-band around 9.5 GHz. Since the spectra of these compounds
are often determined by a combination of frequency-dependent and frequency-independent
interactions, their deconvolution benefits from taking data at microwave
frequencies outside X-band. Typical problems whose solution requires
multifrequency data are, e.g., analysis of poorly resolved metal hyperfine
interaction, disentanglement of overlapping hyperfine spectra from
multiple centers, identification of low-symmetry (monoclinic, triclinic)
coordination, and determination of spatial and redox relations through
dipolar-interaction measurement in multicenter metalloproteins. In
practice, this raises an issue of availability of spectrometers operating
at other than X-band frequencies: conventional EPR spectrometers have
long been[2] single-frequency devices in
which a narrowband source of microwaves is combined with a cavity
resonator as sample holder. The high resonator quality factor of the
cavity ensures practical concentration sensitivity for chemical and
biochemical applications, but its single-frequency operation implies
that multifrequency spectroscopy requires the use of several spectrometers.I have earlier worked out the concept of replacing the resonator
with a coaxial transmission line in which the paramagnetic sample
acts as part of the dielectric that separates the inner and outer
conductor.[3] Unfortunately, in this transmission
EPR spectroscopy the advantage of broad frequency tunability is counteracted
by very low sensitivity of the transverse electromagnetic (TEM) transmission
line. It was, however, found that, in the very low gigahertz range,
the loss of sensitivity could be partially compensated by increasing
the length of the transmission line up to a limit determined by per-length
dielectric losses. The maximum length of the line was furthermore
limited by geometrical constraints of the magnet, and miniaturization
of the line’s axial dimensions was proposed as a future development
to lead to reduction in sample size, to increase in the TEM frequency
cutoff, and to yield more practical cell dimensions for cryogenic
experimentation.Since reduction of paramagnet-filled coaxial
line dimensions poses a considerable engineering challenge, I have
searched for possible alternative implementations of transmission
EPR that would be more suitable for miniaturization. Here I describe
development of such quasi-TEM cells using wire micro strip as the
transmission medium in which the paramagnetic sample is one of the
dielectric phases. It turns out that these cells, when combined with
transmission lines of significant lengths in combination with a tunable
phase shifter to adjust the optical-path length, actually form multimode
resonators with high Q-factors. These broadband-tunable
circuits form the basis for the construction of a single sensitive EPR spectrometer for the detection of transition-metal
complex spectra at an essentially unlimited number of frequencies
and illustrated here for the approximate range of 0.5–12 GHz
(i.e., UHF-, L-, S-, C-, and X-band). Below, I describe the theory,
engineering, and testing of a sensitive broadband spectrometer, and
I illustrate its applicability in a multifrequency study of a dilute
Cu(II) model coordination compound in a crystal of triclinic symmetry.
Results
Spectrometer
Design: The Sample Cell
At the heart of the broadband spectrometer
is a sensitive transmission/reflection device that must be tunable
to a large number of closely spaced resonance frequencies over a wide
frequency span. A key component in the construction of this cell is
the use, as microwave transmission medium, of a long stretch of “wire
micro strip” folded into a small geometric structure. The wire
micro strip principle is illustrated in Figure A: on top of a conducting base plate, covered
with a dielectric insulator, is a conducting wire of circular cross
section surrounded by air (b = 0) or by an additional
layer of the dielectric medium (b ≠ 0). When
current flows through the wire and returns through the ground plate,
electric field lines are set up between the wire and the plate. Magnetic
field lines develop perpendicular to the E lines, and
they form the alternating magnetic field that induces microwave transitions.
This time-dependent field, that is, the microwave magnetic field incident
on the sample, is usually denoted as B1 to
discriminate it from the static field B0 of
the slowly scanning electromagnet that creates the magnetic ground
manifold of energy states of the compound under study. The field lines
of B1 are perfectly parallel to the base in
between the wire and the base, but they diverge on both sides of the
wire (Figure A). However,
when we orient the wire’s long axis along the B0 vector, then the B1 lines are
always perpendicular to B0 as they should
be for regular EPR spectroscopy.
Figure 1
Wire micro strip as a basis for multimode
broadband EPR resonators. (A) Schematic illustration (cross section)
of wire micro strip: a copper wire of diameter d passes
over a copper base plate at a height of h′
mm with a diamagnet ε1 (green) as insulator. The
wire can be surrounded by air (b = 0) or by a second
insulator ε2 (light green) of height b. Solid black lines are electric-field lines; broken red lines are
magnetic-field lines. (B) EPR reflection cell based on wire micro
strip in which some of the diamagnetic isolator is replaced with paramagnetic
sample (orange rectangle). A long optical path is created by having
the wire repeatedly pass over the sample in parallel turns. The shown
geometry ensures that the microwave B1 field
in the EPR sample is everywhere perpendicular to the external static
field B.
Wire micro strip as a basis for multimode
broadband EPR resonators. (A) Schematic illustration (cross section)
of wire micro strip: a copper wire of diameter d passes
over a copper base plate at a height of h′
mm with a diamagnet ε1 (green) as insulator. The
wire can be surrounded by air (b = 0) or by a second
insulator ε2 (light green) of height b. Solid black lines are electric-field lines; broken red lines are
magnetic-field lines. (B) EPR reflection cell based on wire micro
strip in which some of the diamagnetic isolator is replaced with paramagnetic
sample (orange rectangle). A long optical path is created by having
the wire repeatedly pass over the sample in parallel turns. The shown
geometry ensures that the microwave B1 field
in the EPR sample is everywhere perpendicular to the external static
field B.The cell is now constructed as follows (Figure B). A rectangular base plate
of pure coppermetal, with a thickness (i.e., smallest dimension)
of 1 mm to obtain a sturdy structure, is covered with layers of acrylic
tape as the diamagnetic medium. A rectangular volume is cut away from
the acrylic tape on one side, or on both sides, of the cell to create
a space to hold the EPR sample. In other words, the removed part of
the diamagnetic is to be replaced by paramagnetic material. Then a
lacquered copper wire of 0.2 mm diameter is regularly wound around
the cell, keeping sufficient space between windings (≥1.0 mm)
to avoid electronic cross talk. The ends of the wire are made bare
and are soldered to the inner conductor of coaxial connectors of the
SMA type, whose ground is soldered to the short ends of the rectangular
base plate. We have thus constructed a cell operating in transmission
mode. Leaving out one of the SMA connectors, and leaving one end of
the wire either open or soldered to the base plate, creates a cell
operating in reflection mode. The latter is particularly suited for
cryogenic applications: it can be hung from a connecting coax cable
into a conventional helium flow dewar between the poles of an electromagnet.
And when the geometry is chosen as in Figure B, then the wire segments over the paramagnetic
sample all run parallel to the B0 axis, and
thus all B1 lines that penetrate the paramagnetic
sample are in a plane perpendicular to B0.
A technical (3D isometric) drawing can be found in the Supporting Information, part A, Cell Design.A proper set of values for the geometric parameters in Figure A, i.e., the height of the
lower phase h′, the diameter of the wire, d, and the height of the upper phase b,
can be estimated from the universal impedance value of Z = 50 Ω for microwave components in combination with an empirical
approximative expression for the impedance of a square wire micro strip of width w over a dielectric of
height h:[4−7]in whichThis square-wire expression becomes applicable to round wire, with
diameter d, by the substitutionsHere Zair = 377
Ω is the characteristic impedance of air, εR is the relative electric permittivity of the isolator phase, and
εeff is its effective value for the given geometry.
An additional correction for a wire embedded in an extended isolator
phase (b ≠ 0) can be found in the Supporting Information, part B, Theory, where
the theoretical background for these equations is discussed. If we
now take a lacquered copper wire of d = 0.2 mm diameter,
which is thick enough to avoid regular wire breaking and at the same
time small enough to allow for several windings sufficiently separated
to avoid cross talk, then for an isolator thickness of h′= 0.2 mm (i.e., four covers of 0.05 mm acrylic tape) and
minor embedding (b = 0.1), we find a permittivity
εR ≈ 5, which is roughly an intermediate value
for inorganic salts, in order to end up with a characteristic impedance
of Z ≈ 50 Ω. We have herewith created
a low-loss, multimode (see Figure , below) resonator. In practice, there will be discontinuity
losses, e.g., at the boundaries between sample and acrylic, but also
dielectric attenuation, which is linear in the wire length and in
the microwave frequency (Supporting Information, part B, Theory). In addition, we want to purposefully create paramagnetic
loss by bringing the sample into resonance. This extra loss is measured
as the EPR signal; it is linear in the sample’s paramagnetism
(or imaginary relative magnetic permeability) and also in the length
of the wire over the sample and in the microwave frequency as detailed
in the Supporting Information.
Figure 2
Outline of
the microwave part of the broadband reflection spectrometer. Schematic
with components as indicated. (See main text and Supporting Information for details.) In the resonator circuit
that starts, and ends, at port 2 of the circulator, standing microwaves
are set up that become attenuated when the paramagnetic sample in
the broadband cell experiences the EPR effect.
Outline of
the microwave part of the broadband reflection spectrometer. Schematic
with components as indicated. (See main text and Supporting Information for details.) In the resonator circuit
that starts, and ends, at port 2 of the circulator, standing microwaves
are set up that become attenuated when the paramagnetic sample in
the broadband cell experiences the EPR effect.
Spectrometer Design: The Resonator Circuit and Its Operation
Connecting the two ends of the transmission cell with coax cables
to a source and a detector of microwaves and placing the cell in the
field of an electromagnet create a transmission EPR spectrometer.
A mode pattern is set up between the discontinuities created by the
cell’s SMA connectors; however, the pattern is very shallow
(Supporting Information, part C, Mode Patterns)
and the resonance peaks have poor loaded quality factors Q ≤ 10, where Q is defined as the resonance
frequency divided by the peak width at a detection power 3 dB above
that of the negative-peak power:The reflection
version of the cell, with one SMA connection omitted and with the
end of the wire of length L disconnected, behaves
like the microwave equivalent of an open organ pipe of length L in which well-defined resonances are set up at even integers
2n of a quarter wavelength, with wave velocity c0:With the cell connected to a source
and detector via a three-port circulator, as in Figure , the loaded Q factor now
increases by an order of magnitude. Since the only practical way of
further increasing this number is to increase the length of the cell
(at the expense of some line attenuation) we insert a long (up to
a few meters) piece of coax cable between the circulator and the cell,
and this results in an order of magnitude increase in the number of
tunable resonances and some 2 orders of magnitude increase in their Q factor to values of order 104. Finally, we
make the path of the resonator (and therefore its resonance frequencies)
adjustable by the insertion of a broadband, “trombone”
type phase shifter. The whole resonator structure now starts and ends
at port 2 of the circulator as outlined in Figure . Since high-quality microwave circulators
have a limited frequency span, a set of six different circulators
is used substitutionally to cover the range 0.5–12 GHz.A National Instruments vector signal transceiver (VST) is used as
a programmable, solid-state source and detector of microwaves in the
range from 65 MHz to 6 GHz. Maximum power output of 12 dBm is boosted
with a broadband amplifier to some 25 dBm or 250 mW. Between the VST
and the circulator, a home-built frequency multiplier structure is
placed to extend the frequency range to at least 12 GHz. Details of
this multiplier and the other microwave equipment, the electromagnet,
and the helium flow system are given in the Supporting
Information, part D, Hardware.A typical EPR experiment
proceeds as follows. With the microwave source set to an intermediate
output level, e.g., 0 dBm, i.e., 1 mW, a scan is made of the power
reflected by the resonator circuit as a function of the frequency
over the range available for the used circulator. Inspection identifies
a resonance peak, or “dip”, near the frequency of interest.
In a subsequent, real-time procedure (described in detail in the Supporting Information, part E, Software) this
dip is optimized for maximal loaded Q value. Then,
complex IQ (in-phase and quadrature) data are collected at high rate,
typically some 106 signals per second, online converted
to real values of reflected power, and online processed to a numerically
filtered standard size of 1024 field points. Field scans are made,
e.g., in 20 s, and no electronic RC filter is employed. This setup
implies a fast data acquisition rate on the order of 104 averages per spectral point, which is essential for the attained
signal-to-noise ratio of the final spectra. Data from forward and
reverse scans are continuously collected and averaged to a final EPR
absorption spectrum, which is numerically differentiated for ease
of comparison with conventional EPR data. In case the spectrum is
partially dispersive, it is turned into a pure absorption by means
of a Hilbert transform according to Ernst’s procedure.[8] All data acquisition and manipulation software
has been written in LabVIEW with some of the more computationally
demanding routines inserted as dynamic link libraries written in Intel
Visual FORTRAN (see the Supporting Information, part E, Software).
Spectrometer Testing on Dpph Model Spectra
As a test of monochromaticity, stability, and resolution, the single-line
spectrum of solid 2,2-diphenyl-1-picrylhydrazyl (dpph) powder has
been measured at ambient temperature over a wide field scan as a function
of the frequency (Figure ).
Figure 3
Broadband system test on a sharp line over a wide field range.
The single, sharp line (<3 G peak-to-peak) of solid dpph was measured
in single, slow (180 s) scans of 4500 G in 6000-point resolution over
the approximate frequency range 0.5–12 GHz in frequency steps
of ca. 0.5 GHz (i.e., 24 frequencies). The baseline of each spectrum
has been y-axis positioned to the employed frequency
value. Spectral color code: blue, nonmultiplied frequencies; red,
frequency-doubled.
Broadband system test on a sharp line over a wide field range.
The single, sharp line (<3 G peak-to-peak) of solid dpph was measured
in single, slow (180 s) scans of 4500 G in 6000-point resolution over
the approximate frequency range 0.5–12 GHz in frequency steps
of ca. 0.5 GHz (i.e., 24 frequencies). The baseline of each spectrum
has been y-axis positioned to the employed frequency
value. Spectral color code: blue, nonmultiplied frequencies; red,
frequency-doubled.The reflection cell has
a base plate of 6.5 mm width and an extended length of 400 mm. The
goal of the latter is to have the SMA connector well outside the magnetic
field since in a previous work I found SMA connectors to contain an S = 1 paramagnet whose very broad EPR signal shows up as
a sloping baseline.[3] The sample compartment
is 80 × 4 × 0.2 mm, i.e., a volume of 64 μL. With
80 windings, the total wire length (i.e., optical path) from the SMA
connector to the open end is some 130 cm. Slow single forward scans
were recorded over 180 s over a field range of 0–4500 G. Resonance
frequencies (dips) in the range 0.5–12 GHz were chosen to be
approximately separated by 0.5 GHz. A 2D stack plot of the results
is presented in Figure . The separations between baseline levels (vertical axis) and the
separations between dpph signals (horizontal axis) are not constant
because the dips are not exactly spaced at 0.5 GHz intervals. Note
that all zero crossings connect through a straight line diagonal over
the figure, attesting to the linearity of the electronic Zeeman interaction
in the microwave frequency.Then the response of the spectrometer
to changes in microwave power level was studied with the same reflection
cell by measuring the dpph spectrum now over a narrow field range
of 80 G and at a single frequency of 2.6 GHz. The latter value was
chosen to be in the base range of the VST (i.e., <6 GHz) and within
the range of an available high-power (output: 43 dBm or 15 W) broadband
amplifier (0.8–2.7 GHz). Field-forward and field-reverse scans
each of 20 s were collected and combined for increased signal-to-noise
ratio. The spectrometer can be operated over at least 13 decades of
microwave power (Figure A).
Figure 4
Power-response tests of the wire micro strip cell. (A) Integrated
absorption intensity of solid dpph is linear in the incident microwave
power over 13 decades. (B) Spectra at lowest and highest power employed.
(C) Hilbert transformation to remove dispersion, as a function of
incident power. (D) Example of spectrum before (red) and after (blue)
dispersion correction. (E) Loaded Q factor of resonator-circuit
dip near 2.62 GHz versus incident power. Color code (from high to
low power): blue, data obtained with 43 dB amplifier; red, no amplifier
used; black, span to determine dip Q factor extended
from 8 to 45 MHz; cyan, span extended to 70 MHz. (F) Examples of resonator
dip pattern for different incident powers.
Power-response tests of the wire micro strip cell. (A) Integrated
absorption intensity of solid dpph is linear in the incident microwave
power over 13 decades. (B) Spectra at lowest and highest power employed.
(C) Hilbert transformation to remove dispersion, as a function of
incident power. (D) Example of spectrum before (red) and after (blue)
dispersion correction. (E) Loaded Q factor of resonator-circuit
dip near 2.62 GHz versus incident power. Color code (from high to
low power): blue, data obtained with 43 dB amplifier; red, no amplifier
used; black, span to determine dip Q factor extended
from 8 to 45 MHz; cyan, span extended to 70 MHz. (F) Examples of resonator
dip pattern for different incident powers.The integrated intensity of the EPR spectrum is essentially
linear in the applied power. At power levels above the highest value
reported here (34 dBm or 1.56 W, ambient temperature, no cooling),
the system becomes increasingly unstable by produced heat. A pure
absorption spectrum is obtained over a power range of some 60 dB,
namely, from circa −50 to +10 dBm. Outside these limits, increasingly
a Hilbert rotation correction[8] is required
to transform partially dispersive spectra into pure absorptions (Figure C,D). Implementation
of the correction of experimental spectrum Sexp to absorption spectrum Sabsorption is as follows:in which
α is the rotation factor in the range −1 to +1 (0 for
pure absorption and ±1 for pure dispersion), sig is the sign
of α, and H is the fast Hilbert transform routine
from LabVIEW’s Signal Processing Virtual Instruments library.
At the higher end of the power spectrum (microwatt to watt) the loaded
quality factor Q of the resonator circuit, as measured
by fitting a Lorenzian to the dip pattern, is very high, namely, on
the order of (1–2) × 104. Below an incident
power of 1 μW Q decreases linearly and extrapolates
to unity at ca. 100 fW (Figure E,F). Note that the here determined loaded Q factor is a characteristic of the complete system, including the
microwave source and detector, and is not an intrinsic property of
the broadband cell.
Sensitivity Compared to Conventional EPR
Spectrometers
To get an impression of the sensitivity of
the instrument in comparison with conventional single-frequency spectrometers
I measured the signal of a small-volume wire microstrip reflection
cell (4 μL) filled with dpph at dip frequencies close to those
of our S-band and X-band spectrometers and then used an equivalent
sample in our split-ring S-band cavity and our standard rectangular
X-band cavity to measure the signal either by direct detection (i.e.,
cavity connected to the VST via an appropriate circulator but with
mode tuning using the cavity iris in the absence of a tuning phase
shifter) or by conventional 100 kHz field-modulation spectroscopy
using a modulation amplitude of 0.5 G. In all six cases the noise
level was monitored at constant, off-resonance magnetic field for
the same scan time as used for the signal detection. Root-mean-square
noise amplitude was determined by numerical analysis using the Basic
RMS subroutine available in LabVIEW’s Waveform Measurements
palette. Signal intensity was measured as peak-to-peak amplitude of
the unfiltered first-derivative spectra. Resulting S/N levels are
presented in Table .
Table 1
Sensitivity Comparison for Constant Sample Sizea
S-band
X-band
mode of detection
freq (MHz)
S/N
freq (MHz)
S/N
regular EPR with field modulation
4037
978
9871
1704
regular cavity with VST
4038
714
9871
1250
wire micro strip
resonator with VST
3683
237
9829
321
The signal of a
dpph powder sample of 4 μL volume was measured under standardized
conditions of 100 G scan range, −20 dBm microwave power, no
RC-filtering, 120 s acquisition time, and (when applicable) 0.5 G
modulation, with the indicated setups. All resulting 1024-point spectra
were subjected to 6-point Savitzky–Golay filtering. Signal-to-noise
(S/N) values are reported as ratios of the peak-to-peak amplitudes
of the signal and the noise in the first-derivative spectrum.
The signal of a
dpph powder sample of 4 μL volume was measured under standardized
conditions of 100 G scan range, −20 dBm microwave power, no
RC-filtering, 120 s acquisition time, and (when applicable) 0.5 G
modulation, with the indicated setups. All resulting 1024-point spectra
were subjected to 6-point Savitzky–Golay filtering. Signal-to-noise
(S/N) values are reported as ratios of the peak-to-peak amplitudes
of the signal and the noise in the first-derivative spectrum.Under the employed instrument settings
direct detection with the VST using a regular cavity is only mildly
less sensitive than conventional field-modulation detection with the
same cavity. More remarkably, with the present state-of-the-art of
the broadband wire micro strip reflection cell, the S/N levels obtained
in this particular case are only an approximate factor of 5 below
the levels of the single-frequency cavity-based experiments. Generalization
of this condition would imply that, where the recording of a good
quality spectrum with an X-band spectrometer typically takes some
3 min, obtaining comparable quality data anywhere in the 0.5–12
GHz frequency range with the present setup would require some 3 ×
52 = 75 min.
Broadband EPR of Low-Symmetry Cu(II) Complex
A preparation of 0.5% Cu(II) in diamagnetic ZnSO4·1H2O finely ground powder was taken as a realistic spectroscopic
test system. The ZnSO4·1H2O crystal is
monoclinic[9] and CuSO4·1H2O is triclinic.[10] The all-oxygen
ligation excludes a superhyperfine interaction, and the 1:200 Cu:Zn
dilution minimizes dipolar interaction between copper centers. Furthermore,
since the sample is a dry powder, the strong variation of line width
with nuclear quantum number mI and with
microwave frequency that is typically observed in frozen aqueous solutions
of copper complexes as a consequence of strain,[11] was anticipated to be absent or at least inextensive. Thus,
the spin Hamiltonian for this low-symmetry 3d9 system with S = 1/2, I = 3/2 is one of considerable,
but perhaps not overwhelming complexity:in which g, A, and P are diagonal, rhombic matrices/tensors, R is the Euler rotation matrix that relates the Cartesian
axis system that diagonalizes the hyperfine matrix A with
that in which g is diagonal, and R2 is the rotation matrix that diagonalizes the quadrupole tensor P (Supporting Information, part
F, Hamiltonian).Low-temperature data were collected at multiple
frequencies from 0.8 to 11 GHz. Data taken with conventional spectrometers
at the X- and Q-bands were added as benchmarks of sensitivity and
resolution. Figure is an overview of the resulting spectra plotted on a reciprocal g value scale (or linear energy scale) as the abscissa and
the microwave frequency as the ordinate for the spectral baselines,
which is one out of several possible 2D presentations, this one emphasizing
the nonlinear reduction of the hyperfine structure with increasing
frequency. The complete set exhibits an unprecedented richness of
fine spectral details: at higher frequencies a hyperfine splitting
of considerable magnitude along g is clearly resolved. The g four-line hyperfine pattern at higher frequencies is straightforward
with equidistant peaks of equal width except for a slight broadening
at the extremes due to the presence of the two natural abundant isotopes 63Cu (∼69%) and 65Cu (∼31%).
Figure 5
Field-frequency 2D plot of 0.5% Cu(II)
in ZnSO4 from 0.8 to 35.1 GHz. The signal was measured
at the nine indicated frequencies, and the set was augmented with
data obtained with regular X-band and Q-band spectrometers. For the
broadband machine a wire micro strip resonator was used with a sample
volume of 70 × 27 × 0.2 mm on both sides, i.e., a total
volume of 0.75 mL. Spectra were collected by averaging over 30–80
min at an incident power level 1 mW and a sample temperature of 16
K. The X-band spectrum was taken in 4 min with 0.2 mW, 2 G modulation,
and 16 K. The Q-band spectrum was collected in 12 min with 20 mW,
10 G modulation, and 34 K. The magnetic-field scale of all spectra
has been replotted on a reciprocal g value scale, 1/g = βB/hν. Vertical
broken blue lines correspond to the first hyperfine line with B = B(g) – 1.5A, the g value (straight
line), and the last hyperfine line with B = B(g) + 1.5A.
Field-frequency 2D plot of 0.5% Cu(II)
in ZnSO4 from 0.8 to 35.1 GHz. The signal was measured
at the nine indicated frequencies, and the set was augmented with
data obtained with regular X-band and Q-band spectrometers. For the
broadband machine a wire micro strip resonator was used with a sample
volume of 70 × 27 × 0.2 mm on both sides, i.e., a total
volume of 0.75 mL. Spectra were collected by averaging over 30–80
min at an incident power level 1 mW and a sample temperature of 16
K. The X-band spectrum was taken in 4 min with 0.2 mW, 2 G modulation,
and 16 K. The Q-band spectrum was collected in 12 min with 20 mW,
10 G modulation, and 34 K. The magnetic-field scale of all spectra
has been replotted on a reciprocal g value scale, 1/g = βB/hν. Vertical
broken blue lines correspond to the first hyperfine line with B = B(g) – 1.5A, the g value (straight
line), and the last hyperfine line with B = B(g) + 1.5A.As an illustration of how we can go about analyzing such
a data set Figure shows simulations at high (Q-band: 35.088 GHz), intermediate (C-band:
5.772 MHz), and low (L-band: 1.221 GHz) frequency. In the left-hand
panels the Cu(II) electronic structure is modeled as being of orthorhombic
symmetry, in the right-hand panels triclinic symmetry is assumed congruent
with crystal structure.[9,10] Spectroscopically, the key difference
between the two sets is that in the lower symmetry the g- and A-matrices in the spin Hamiltonian are not diagonal
in the same axes system. Attempts to explore the occurrence of lower
than orthorhombic symmetry in the EPR of randomly oriented coordination
complexes are rare.[12,13] In spectral analysis by simulation
it is common to probe only isotropic, axial, and rhombic powder patterns.
The customary approach is that of using an analytical expression deduced
with perturbation theory in which the metal hyperfine interaction
is taken as a perturbation, up to second order, of the electronic
Zeeman interaction. For orthorhombic symmetry the relevant expression
was deduced for frequency-swept spectra by Pake and Estle in 1971
(eqs B15–B17 in ref (14)), and I have rewritten their result for the more practical
situation of field-swept spectra (eqs 5.18–5.33 in ref (15)). Analytical perturbation
expressions for power patterns of lower than orthorhombic symmetry
have not been published. For the present situation the perturbation
approach will not suffice because the assumption that a copper hyperfine
interaction is a perturbation to a copper Zeeman interaction breaks
down at, and below, L-band frequencies.[15] I therefore used a simulator based on full-blown energy matrix diagonalization
for position and intensity computation which is valid for any ratio
of hyperfine over Zeeman interaction. This approach has the added
advantage that extension to more complex Hamiltonians, in particular
to triclinic symmetry, and also to inclusion of copper quadrupolar
interaction does not significantly increase the required computational
time per simulation, and this makes a rigorous test on metal-site
symmetry practical. Figure is the result of such a test.
Figure 6
Orthorohombic and triclinic
symmetry based simulations of the Cu site in Q-, C-, and L-bands.
The experimental data are from Figure . All simulations are based on a single set of g values: g = 2.5115, 2.1235, 2.0675.
The orthorhombic A values (gauss) are A = 69, 9, 35; the triclinic values
are A = 75, 0, 0 with
rotation (degrees) α = +24 and β = +20. The numbers apply
to 63Cu; the simulations are for natural abundance 63Cu + 65Cu. In all cases quadrupolar interaction
is not resolved; i.e., the P matrix is zero. Gaussian
line widths (standard deviation in gauss) are isotropic 5 in the L-band
and 6 in the C-band, and W = 12, 23, 14 (orthorhombic) or 13, 22, 14 (triclinic) in the
Q-band. The orthorhombic symmetry based simulation is consistent over
the full frequency range (ignoring a variation in line width with
nuclear orientation), and the triclinic symmetry based simulation
is not.
Orthorohombic and triclinic
symmetry based simulations of the Cu site in Q-, C-, and L-bands.
The experimental data are from Figure . All simulations are based on a single set of g values: g = 2.5115, 2.1235, 2.0675.
The orthorhombic A values (gauss) are A = 69, 9, 35; the triclinic values
are A = 75, 0, 0 with
rotation (degrees) α = +24 and β = +20. The numbers apply
to 63Cu; the simulations are for natural abundance 63Cu + 65Cu. In all cases quadrupolar interaction
is not resolved; i.e., the P matrix is zero. Gaussian
line widths (standard deviation in gauss) are isotropic 5 in the L-band
and 6 in the C-band, and W = 12, 23, 14 (orthorhombic) or 13, 22, 14 (triclinic) in the
Q-band. The orthorhombic symmetry based simulation is consistent over
the full frequency range (ignoring a variation in line width with
nuclear orientation), and the triclinic symmetry based simulation
is not.It is found that the high-frequency,
Q-band data are equally well fit with either an orthorhombic Hamiltonian
or a triclinic one. However, the underlying models are very different:
orthorhombic symmetry requires the assumption of a rather unusually
anisotropic A-matrix, while a triclinic-symmetry fit
of equal quality is obtained assuming an axial A-matrix
with dominant A term
and small splittings (0–10 G) in the perpendicular direction
(a pattern that abounds in the copper EPR literature over the last
half a century) but here with rotation required between g and A. On the basis of this analysis of single-frequency
data, one would be tempted to conclude that an appealing congruence
exists indeed between electronic structure (triclinic) and crystal
structure (triclinic). The multifrequency data now afford a rigorous
test between the two models. Upon a 6-fold reduction in frequency
to the C-band (the middle panels of Figure ) the triclinic simulation, based on the
parameters established in the Q-band, exhibits significant misfit
in the xy region, while the orthorhombic simulation
rather faithfully reproduces the C-band spectrum. With another 5-fold
reduction in frequency to the L-band, the difference in quality of
the two fits is even more pronounced. We are thus forced to revisit
our conclusion that was based on the Q-band data alone: fits of the
multifrequency low-temperature EPR data are consistent with orthorhombic
symmetry and not with the room-temperature crystallographic triclinic
symmetry.As an aside, the fittings in Figure led to some additional conclusions: introduction
of finite quadrupolar terms in the Hamiltonian, be the Q-tensor collinear with g, or with A, or
subject to a general rotation, never led to a quality increase, of
any significance, of the fits. Apparently, copper quadrupolar interaction
is not resolved in this multifrequency data set. Also, with this powder
sample a variation of line width, W, with nuclear
orientation, i.e., as a function of quantum number mI, is in fact observed (contrast to initial expectation)
in all frequency bands especially in the x direction.
Such a pattern has been included in the past in copper simulations
in the form of ad hoc expressions of the form W = f(mI), for example, W = A + BmI + CmI2, but since its interpretation
has until now not moved beyond a qualitative assignment to “strains”.[15] I have not pursued this subject here. Also,
the experimental data exhibit minor details that are not reproduced
in the fits. This is particularly clear in the spectra in the 5–11
GHz range (Figure ) where in the A pattern
we consistently observe line asymmetries and also variation in the
three splitting between the four hyperfine lines. A possible source
for these small effects might be dipolar interaction between different
copper sites of the 0.5% Cu-doped ZnSO4 microcrystals.
Inclusion of these effects in the energy-matrix diagonalization routine
would be theoretically trivial but practically quite taxing: even
inclusion of pairwise Cu–Cu interaction (dimer formation) only
would increase the size of the energy matrix from 8 × 8 to 64
× 64. That dimerization indeed occurs is evident from X-band
data taken at ambient temperature showing a half-field transition
plus a broad spectrum underlying the monomeric copper spectrum (Supporting Information, part G, Dimerization).
Finally, as a control experiment on the methodology, it was ascertained
that, for the C- and Q-band spectra orthorhombic simulations based
on the perturbation-theory expression gave fits identical to those
from energy-matrix diagonalization.
Discussion
The
resonant cavity has been at the heart of EPR spectrometers from early
on in the history of the spectroscopy. The first explicit reference
to a microwave cavity, of undefined dimensions, for S-band EPR was
in 1946 by Cummerow and Halliday.[2] Possibly
even earlier, the discoverer of EPR spectroscopy, E. K. Zavoisky,
may have used what appears to be a split-ring cavity also for the
S-band.[16] Shortly thereafter, Bagguley
and Griffiths reported an EPR study in which use was made of an X-band
cavity, again of unspecified dimensions and geometry.[1] Since then, the single-mode (i.e., always operating at
its fundamental frequency) either rectangular or cylindrical X-band
cavity (9–10 GHz) became the de facto standard, which it still
is today, as the result of a combination of practical determinants,
in particular: relative ease of construction, availability of X-band
sources, and favorable noise characteristics of detection diodes.
Later, in response to the need of routinely collecting data at least
at one other frequency for, e.g., systems with convoluted hyperfine
and Zeeman interactions, a commercial Q-band (ca. 35 GHz) spectrometer
was produced and was quickly adopted for studies of, for example,
Cu(II)[17] or Mo(V)[18] proteins. In more recent times commercial continuous-wave (cw) spectrometers
have become available from the L-band (ca. 1 GHz) to the J-band (ca.
263 GHz), and outside this range machines have been laboratory-built
for subgigahertz[19] to terahertz[20] frequencies. In spite of these developments,
multifrequency EPR has not become a routine practice in the study
of transition ion complexes, presumably due to the excessive cost
of investment and operational complexity. Furthermore, X-band flanking
studies in the approximate 1–20 GHz range are relatively rare,
which is remarkable in view of the fact that pronounced spectral changes
with frequency are to be expected here, in particular for the many
systems with significant central hyperfine interaction or with multicenter
dipolar interactions such as powders of synthetic coordination compounds
as well as complex metalloproteins. These problems are also not well
addressed using the many special forms of advanced EPR based on double
resonance and/or pulse techniques, which find their main field of
application in the resolution of small superhyperfine interactions
from ligands. A paradigm-changing approach would appear to be desirable:
broadly frequency tunable measurements with a single spectrometer
of sufficient sensitivity for the collection of high-quality cw EPR
spectra from dilute, randomly oriented transition ion complexes.This quest for a fundamental change has two key technical aspects:
a broadly tunable source and direct-detection system and a broadly
tunable resonator. For the frequency range of interest, e.g., 1–20
GHz, only a small number of relevant papers have appeared in the last
half a century of EPR literature. In an early study (1962) use of
a nonresonant, frequency-independent helix was proposed as a microwave
delay line.[21] With a sample placed inside
the helix, this structure is similar to the probe described here,
but with the ground plate absent. A microcoil version with inner diameter
of ca. 0.1 mm was later used to build a frequency-scanning (0.1–8.5
GHz) spectrometer with constant field, which was applied to strong
samples of dpph and AgMn spin glass.[22] In
a different vein, a broadly tunable (0.34–4.0 GHz) “cavity”
was proposed with the helix surrounded by a ground plate in the form
of a rectangular box.[23] No practical EPR
applications were described. An early (1966) study on mechanically
adjustable cavities in the 1.0–4.6 GHz range found no follow
up, possibly because of limited stability of the tunable source, a
modulated traveling wave tube.[24] A more
recent attempt describes a similar resonator now for 4–40 GHz
combined with a vector network analyzer with direct detection, but
the sensitivity of the setup is not clear, as only single crystal
samples of pure molecular magnets were measured.[25] An early (1974) attempt to use strip line as the basis
for a broadband resonator was tested, using conventional S-band and
X-band equipment, and was found to give reasonably good-quality spectra
of Mn(II) contamination in MgO,[26] but again
no follow up was reported. Several more recent papers describe the
testing of sample cells based on either meandering micro strips[27,28] or coplanar thin films[29,30] with direct detection
with a VNA, but more often with less sensitive diode-based power meters.
Resolution appears to be oftentimes problematic, e.g., when a dpph
line is found with a severely broadened fwhh line width of ca. 22
G[27] or when a nitronyl-nitroxide radical
is found to give a single line[29,30] where regular EPR of
these compounds is typically rich in hyperfine structure.[31] Also, detection of 0.2 μg of dpph with
a signal-to-noise ratio of 121 may suggest an impressive absolute
sensitivity,[28] but realizing that the sample
is the dried out result of a 5 mM stock solution makes one aware that
the concentration sensitivity is in fact rather poor; in chemical
and biochemical applications of dilute transition ion EPR the concentration sensitivity is of far greater importance than
the absolute sensitivity. Other studies have reported
the use of analog or digital, scalar or vector network analyzers as
source/detector of EPR[32−35] including my own work.[3] None of the cited
papers provide examples of multifrequency data of high quality from
randomly oriented, dilute transition ion complexes.To my knowledge Figure of the present paper
is the first instance of such a milestone. It makes a mark of technical
feasibility: the required equipment is not more complex and not more
expensive than a standard commercial single-frequency spectrometer;
in particular, the resonator is much simpler and cheaper than a standard
cavity. It is also a mark of technical potential: concentration sensitivity
is less than an order of magnitude below the present optimum of standard
X-band equipment, and improvements can be envisioned (e.g., modulation
of field or frequency). First and foremost, however, it outlines a
new chemical goal for EPR spectroscopy of randomly oriented systems:
a methodology for the reliable determination of molecular electronic
structure based on analysis of field-frequency 2D data sets.