Dynamic beam current control for improved dose accuracy in PBS proton therapy.

C Bula1, M F Belosi, M Eichin, J Hrbacek, D Meer.   

Abstract

The step-and-shoot method of pencil beam scanning delivers the dose on a 3D grid in the target volume, with one dimension defined by the proton energy. While the dose per pencil beam may vary substantially within an iso-energy layer, the beam current typically remains constant. In this static operation mode, the inherent latency of the beam switch-off mechanism results in a lower limit for the deliverable spot dose, which may prevent the application of some of the low-weighted spots prescribed by the treatment planning system. To overcome this limitation, we introduced dynamic beam current control at the PSI Gantry 2, an innovative new approach successfully commissioned and in clinical operation since fall 2017. The control system was enhanced with a direct link to the vertical deflector located at the centre of the cyclotron. This connection allows much faster beam current changes (~0.1 ms) and hence opens up the possibility of dynamically reducing the current for individual low-dose spots. We demonstrate that with this new dynamic operation mode, all spots are delivered as planned without compromising treatment time. We show by two independent and complementary methods that the delivered dose distribution is improved.

Entities:  

Mesh:

Year:  2019        PMID: 31315104     DOI: 10.1088/1361-6560/ab3317

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Single-fraction 34 Gy Lung Stereotactic Body Radiation Therapy Using Proton Transmission Beams: FLASH-dose Calculations and the Influence of Different Dose-rate Methods and Dose/Dose-rate Thresholds.

Authors:  Patricia van Marlen; Wilko F A R Verbakel; Ben J Slotman; Max Dahele
Journal:  Adv Radiat Oncol       Date:  2022-04-10

2.  The potential of Gantry beamline large momentum acceptance for real time tumour tracking in pencil beam scanning proton therapy.

Authors:  Giovanni Fattori; Ye Zhang; David Meer; Damien Charles Weber; Antony John Lomax; Sairos Safai
Journal:  Sci Rep       Date:  2020-09-18       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.