Literature DB >> 31313811

Effects of Intraoperative Magnetic Resonance Thermal Imaging Signal Artifact During Laser Interstitial Thermal Therapy on Thermal Damage Estimate and Postoperative Magnetic Resonance Imaging Ablative Area Concordance.

Sean M Munier1, Akshay N Desai1, Nitesh V Patel1, Shabbar F Danish1.   

Abstract

BACKGROUND: Magnetic resonance-guided laser interstitial thermal therapy (MRgLITT) is a minimally invasive procedure that utilizes intraoperative magnetic resonance thermal imaging (MRTI) to generate a thermal damage estimate (TDE) of the ablative area. In select cases, the MRTI contains a signal artifact or defect that distorts the ablative region. No study has considered the impact of this artifact on TDE accuracy.
OBJECTIVE: To determine the effect of intraoperative MRTI signal artifact on postoperative magnetic resonance imaging (MRI)-predicted ablative area.
METHODS: All ablations were performed using the Visualase MRI-Guided Laser Ablation System (Medtronic). Patients were grouped based on whether the intraoperative MRTI contained signal artifact that distorted the ablative region. Cross-sectional area of the ablative lesion from the MRI image was measured, and the difference between intraoperative TDE and postoperative MRI cross-sectional area was calculated and compared between groups with and without intraoperative MRTI artifact.
RESULTS: A total of 91 patients undergoing MRgLITT for various surgical indications were examined. MRTI artifact was observed in 43.9% of cases overall. The mean absolute difference between TDE and the postoperative MRI cross-sectional area was 94.8 mm2 (SEM = 11.6) in the group with intraoperative MRTI artifact and 54.4 mm2 (SEM = 5.5) in the nonartifact group.
CONCLUSION: MRTI signal artifact is common during LITT. The presence of signal artifact during intraoperative MRTI results in higher variation between intraoperative TDE and postoperative MRI cross-sectional ablative area. In cases in which intraoperative MRTI artifact is observed, there may be a larger degree of variation between observed intraoperative TDE and measured postoperative MRTI ablative area.
Copyright © 2019 by the Congress of Neurological Surgeons.

Entities:  

Keywords:  LITT; Laser ablation; Laser interstitial thermal therapy; MRTI; Magnetic resonance imaging; Magnetic resonance thermometry; Thermal ablation parameters

Mesh:

Year:  2020        PMID: 31313811     DOI: 10.1093/ons/opz182

Source DB:  PubMed          Journal:  Oper Neurosurg (Hagerstown)        ISSN: 2332-4252            Impact factor:   2.703


  1 in total

1.  Biopsy Artifact in Laser Interstitial Thermal Therapy: A Technical Note.

Authors:  Thomas Noh; Parikshit Juvekar; Raymond Huang; Gunnar Lee; Christian T Ogasawara; Alexandra J Golby
Journal:  Front Oncol       Date:  2021-11-18       Impact factor: 6.244

  1 in total