Technical Note: Assessment of scatter originating from the x-ray tube collimator assembly of modern angiography systems.

Ashley T Tao1, Daniel Miller1, Mark Hindal1, Kenneth A Fetterly2.   

Abstract

PURPOSE: While scatter from the patient is assumed to be the primary source of occupational radiation dose associated with fluoroscopically guided interventional procedures, the potential contribution of scatter from the x-ray collimator assembly is unknown. The purpose of this work was to survey clinical x-ray angiography systems to assess the potential contribution of collimator assembly scatter on occupational radiation dose.
METHODS: Experimental methods were designed to measure the relative contributions of scatter originating from within the collimator assembly of the x-ray tube to total scatter, which included scatter from a patient-simulating phantom. Measurements were acquired as a function of lateral distance from the x-ray beam center using a posterior anterior (PA) projection and at a fixed location for variable right anterior oblique to left anterior oblique projections in the range -90º to 90º. For one system, the collimator assembly was partially disassembled to assess the scatter contribution of individual components. For two systems, 0.5 mm Pb was added to the inner surface of the collimator assembly cover and tested for efficacy to block collimator assembly scatter.
RESULTS: Considering all x-ray systems and only the PA projection, collimator assembly scatter contributed 20-50% to total scatter. For x-ray projection angles of -90º to 90º, the relative contribution of collimator assembly to total scatter was dependent on projection angle and ranged from 5% to 56%. X-ray systems with kerma-area product meters demonstrated higher collimator assembly scatter than those without. Considering all projection angles, the addition of 0.5 mm Pb to the inside of the collimator assembly cover reduced collimator assembly scatter from 28% to 16% of total scatter for both systems.
CONCLUSION: Findings from this work suggest that contemporary radiation safety practices and guidelines should be revised to account for scatter originating from the collimator assembly of angiographic x-ray tubes.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  fluoroscopy; occupational dose; x-ray collimator assembly scatter

Mesh:

Year:  2019        PMID: 31310325     DOI: 10.1002/mp.13720

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071