Literature DB >> 31307812

Clinical evaluation of two-piece zirconia abutments with bonded titanium inserts for implant-supported restorations.

Yu-Ting Lin1, Yu-Fu Shen2, Pein-Chi Wei3, Kuang-Wei Hsu4.   

Abstract

STATEMENT OF PROBLEM: Information about the long-term evaluation of 2-piece zirconia abutments (TPZAs) with bonded titanium inserts for implant-supported restorations is sparse.
PURPOSE: The purpose of this clinical study was to evaluate the clinical outcome of TPZAs with bonded titanium inserts for implant-supported restorations.
MATERIAL AND METHODS: This study retrospectively evaluated the records of patients who received implant restorations supported by TPZAs with bonded titanium inserts. The patients were recalled between January 2017 and December 2017 so that the restorations could be examined clinically and radiographically for technical and biological complications.
RESULTS: A total of 27 patients with 30 restorations and 42 zirconia abutments (ZAs) were included in the study. The patient records revealed that all implants remained integrated. However, 5 restorations supported by seven ZAs in 4 patients had to be refabricated because of abutment failure before the recall. Six of the failed abutments were in the molar area, with a combination of zirconia and titanium as the connection platform. Additional complications included screw loosening (1 patient), ceramic chipping (1 patient), and peri-implantitis (1 patient). The mean observation period was 6.6 ±1.5 years (range: 3.6 to 9.2 years).
CONCLUSIONS: After a mean observation period of 6 years after insertion, TPZAs with bonded titanium inserts appear to be a suitable treatment option in the anterior and premolar regions. However, in the molar area, the use of the same type of ZA without a complete metal-to-metal connection platform to support the restoration led to a high incidence of fracture.
Copyright © 2019 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2019        PMID: 31307812     DOI: 10.1016/j.prosdent.2019.01.006

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  6 in total

1.  Fractographic and Microhardness Evaluation of All-Ceramic Hot-Pressed and CAD/CAM Restorations after Hydrothermal Aging.

Authors:  Roxana Diana Vasiliu; Ion-Dragoș Uțu; Lucian Rusu; Adrian Boloș; Liliana Porojan
Journal:  Materials (Basel)       Date:  2022-06-03       Impact factor: 3.748

2.  Fracture strength analysis of titanium insert-reinforced zirconia abutments according to the axial height of the titanium insert with an internal connection.

Authors:  Seung-Rye Song; Kyeong-Mee Park; Bock-Young Jung
Journal:  PLoS One       Date:  2021-04-01       Impact factor: 3.240

3.  Comparative finite element analysis of mandibular posterior single zirconia and titanium implants: a 3-dimensional finite element analysis.

Authors:  Sung-Min Choi; Hyunsuk Choi; Du-Hyeong Lee; Min-Ho Hong
Journal:  J Adv Prosthodont       Date:  2021-12-22       Impact factor: 1.904

4.  Titanium Base Abutments in Implant Prosthodontics: A Literature Review.

Authors:  Ahmad M Al-Thobity
Journal:  Eur J Dent       Date:  2021-11-18

5.  Comparing the Fracture Resistance and Modes of Failure in Different Types of CAD/CAM Zirconia Abutments with Internal Hexagonal Implants: An In Vitro Study.

Authors:  Yu-Tsen Chang; Yu-Ling Wu; Hung-Shyong Chen; Ming-Hsu Tsai; Chia-Chen Chang; Aaron Yu-Jen Wu
Journal:  Materials (Basel)       Date:  2022-04-04       Impact factor: 3.623

6.  Clinical and Laboratory Outcomes of Angled Screw Channel Implant Prostheses: A Systematic Review.

Authors:  Vanya Rasaie; Jaafar Abduo; Mehran Falahchai
Journal:  Eur J Dent       Date:  2022-02-21
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.