| Literature DB >> 3129630 |
N Scher1, D Poe, F Kuchnir, C Reft, R Weichselbaum, W R Panje.
Abstract
There is general concern among otolaryngologists that irradiation of a stainless steel prosthesis used in mandibular reconstruction may cause irradiation overdosage to adjacent tissues. A tissue-equivalent plastic/steel model, simulating the characteristics of a stainless steel, reconstructed mandible, was irradiated and measurements were made with a parallel plate ionization chamber. The results of our measurements show that irradiation of an implanted steel plate results in an overdosage (120%) "in front" and an underdosage (80%) "behind" the steel plate. The regions of overdosage and underdosage are 2 to 3 mm thick. The overall dose modification is greatly reduced when two opposing fields are used. We conclude that irradiation of a stainless steel, reconstructed mandible with a 6-mV photon beam through opposing fields does not significantly alter the amount of radiation delivered to surrounding tissues.Entities:
Mesh:
Substances:
Year: 1988 PMID: 3129630 DOI: 10.1288/00005537-198805000-00017