Literature DB >> 31295707

Wireless power transfer analysis of circular and spherical coils under misalignment conditions for biomedical implants.

Yellappa Palagani1, Krithikaa Mohanarangam1, Jae Hoon Shim1, Jun Rim Choi2.   

Abstract

In this paper, a 3-coil inductive WPT system having circular shaped external coil and miniaturized spherical shaped implantable coil is designed and optimized at 13.56 MHz to overcome reduced PTE under misalignment conditions. The external coil is placed at a distance of 10 mm from the implantable coil. The proposed design is referred to as C-S WPT system. To evaluate the performance of the C-S WPT system, a 3-coil inductive WPT system having circular shaped external coil and circular shaped implantable coil is designed and is referred to as C-C WPT system. The impacts of perfect alignment, lateral and/or angular misalignments on the overall PTE of the systems are analyzed and compared. At perfect alignment, the PTE of C-C and C-S WPT systems are 59.8% and 52.6%, respectively. Though the PTE of C-C WPT system is slightly higher than C-S WPT system, there is a sharp decrease in PTE of C-C system for angles greater than 40°. At 90° of angular misalignment, the PTE's of C-C and C-S WPT systems are 0.4% and 5.6% respectively. When the coils are laterally displaced by a distance of 10 mm, the C-C and C-S WPT systems produce PTE of 33.4% and 27.1% respectively. The C-C and C-S WPT systems attain PTE of 1.6% and 5.5% respectively at 10 mm displacement and 50° rotation. The simulated systems are fabricated to analyze the result in real-environment. The measurement results show that the C-S WPT system provides better PTE when compared to C-C WPT system under angular and a combination of angular and lateral misalignments.
Copyright © 2019. Published by Elsevier B.V.

Entities:  

Keywords:  3-Coil inductive WPT system; Circular; IMD'S; Misalignment issue; PTE; Spherical; Three-layered human tissue

Year:  2019        PMID: 31295707     DOI: 10.1016/j.bios.2019.04.051

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  1 in total

1.  Simplified Mutual Inductance Calculation of Planar Spiral Coil for Wireless Power Applications.

Authors:  Iftikhar Hussain; Dong-Kyun Woo
Journal:  Sensors (Basel)       Date:  2022-02-16       Impact factor: 3.576