Determination of Phosphate in Human Serum with Zirconium/Reduced Graphene Oxide Modified Electrode.

Sima Pourbeyram1, Maryam Soltanpour1, Soghra Fathalipour1.   

Abstract

An electrochemical phosphate sensor based on zirconium and reduced graphene oxide modified pencil graphite electrode (Zr/rGO-PGE) is proposed. The XRD, CV and EIS confirmed that GO was partially reduced on the PGE. Scanning electron microscopy (SEM) exhibited the layered and wrinkled structures for the rGO-PGE and Zr/rGO-PGE, respectively. Cyclic voltammetry showed the immobilized rGO was highly stable and had high activity toward zirconium adsorption. The prepared electrode was used for the electrochemical determination of phosphate. Based on the optimum condition using differential pulse voltammetry, the limit of detection and sensitivity for phosphate was obtained as [0.011(± 0.004) μM] (S/N = 3) and [622.4(± 9.6) μA μM-1 cm-2], respectively. The sensor was successfully evaluated for phosphate determination in human serum samples. In practical terms, the construction of this sensor was exceptionally simple, fast, cost effective and reproducible.

Entities:  

Keywords:  Graphene oxide; pencil graphite electrode; phosphate; zirconium

Year:  2019        PMID: 31292353     DOI: 10.2116/analsci.18P548

Source DB:  PubMed          Journal:  Anal Sci        ISSN: 0910-6340            Impact factor:   2.081


  2 in total

1.  Square-wave Adsorptive Anodic Stripping Voltammetric Determination of Antidiabetic Drug Linagliptin in Pharmaceutical Formulations and Biological Fluids Using a Pencil Graphite Electrode.

Authors:  Ahmed H Naggar; Gamal A Saleh; Mahmoud A Omar; Ahmed M Haredy; Sayed M Derayea
Journal:  Anal Sci       Date:  2020-02-14       Impact factor: 1.967

Review 2.  Graphene-Based Composites for Phosphate Removal.

Authors:  Sanny Verma; Mallikarjuna N Nadagouda
Journal:  ACS Omega       Date:  2021-02-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.