Agreement between supine and standing bioimpedance spectroscopy devices and dual-energy X-ray absorptiometry for body composition determination.

Michael R Esco1, Michael V Fedewa1, Todd J Freeborn2, Jordan R Moon3, Jonathan E Wingo1, Zack Cicone1, Clifton J Holmes1, Bjoern Hornikel1, Bailey Welborn1.   

Abstract

BACKGROUND: Research comparing bioimpedance spectroscopy (BIS) to dual-energy X-ray absorptiometry (DXA) is limited, especially with newer BIS devices that take measures in a standing position instead of the traditional supine position.
PURPOSE: The purpose of this study was to compare a standing BIS device (BISSTA ) and a supine BIS device (BISSUP ) to DXA for measuring body fat percentage (BF%), fat mass (FM) and fat-free mass (FFM) in a cohort of male and female subjects displaying a wide range of ages and BMI levels.
METHODS: Ninety-five subjects (30 ± 15 years, 170 ± 8·0 cm, 72·6 ± 14·8 kg) participated in the study. Body composition measures were taken from BISSTA , BISSUP and DXA during a single visit to the laboratory following an 8- to 12-h fast in a euhydration state.
RESULTS: Supine BIS device and BISSTA produced r-values >0·91 and low SEE values for all measurements compared to DXA. Effect sizes were 'trivial' for FFM comparing both BISSUP and BISSTA to DXA (<0·1) and 'small' for FM and BF% (<0·39). Compared to DXA, BISSTA resulted in lower total (TE) and constant errors/mean differences (CE) (TE < 3·6 kg, CE < -1·82 kg) versus BISSUP (TE < 4·35 kg, CE < -3·10 kg) for FFM.
CONCLUSION: Fat-free mass values for BISSTA resulted in the most comparable measurements to DXA with no mean differences and the lowest total error and effect size. However, the findings indicated both BIS devices may be acceptable alternatives to DXA for BF%, FM and FFM in clinical practice.
© 2019 Scandinavian Society of Clinical Physiology and Nuclear Medicine. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  adiposity; bioimpedance spectroscopy; body fat; dual-energy X-ray absorptiometry; fat-free mass

Year:  2019        PMID: 31287228     DOI: 10.1111/cpf.12585

Source DB:  PubMed          Journal:  Clin Physiol Funct Imaging        ISSN: 1475-0961            Impact factor:   2.273


  2 in total

1.  Skeletal Muscle Myofibrillar Protein Abundance Is Higher in Resistance-Trained Men, and Aging in the Absence of Training May Have an Opposite Effect.

Authors:  Christopher G Vann; Paul A Roberson; Shelby C Osburn; Petey W Mumford; Matthew A Romero; Carlton D Fox; Johnathon H Moore; Cody T Haun; Darren T Beck; Jordan R Moon; Andreas N Kavazis; Kaelin C Young; Veera L D Badisa; Benjamin M Mwashote; Victor Ibeanusi; Rakesh K Singh; Michael D Roberts
Journal:  Sports (Basel)       Date:  2020-01-10

2.  The influence of body position on bioelectrical impedance spectroscopy measurements in young children.

Authors:  Jaz Lyons-Reid; Leigh C Ward; Mya-Thway Tint; Timothy Kenealy; Keith M Godfrey; Shiao-Yng Chan; Wayne S Cutfield
Journal:  Sci Rep       Date:  2021-05-14       Impact factor: 4.379

  2 in total