Yang-Kun Ou1, Ming-Jui Wu2,3, Wei-Siang Ciou3, Yi-Chun Du4.
Abstract
In hemodialysis, vascular access is usually achieved through an arteriovenous fistula, and a dislodged needle can cause varying degrees of injury to patients. In severe cases, the loss of blood can prove to be fatal. This study proposed a blood leakage detection device for patients during hemodialysis (HD). First, the device was tested on a phantom arm, and later in a clinical test on patients receiving HD. The thoughts of the patients and the nursing staff involved were surveyed before and after the introduction of the device. Analysis of the results indicated that the device achieved 100% and 98.9% accuracy rates on the phantom arm test and clinical test, respectively. The results suggested that patients believed the device could reduce their mental anxiety, and the nursing staff considered the device reliable and that it would enhance the quality of care. The proposed detection device can be extended to similar applications for preventing catheter dislodgement, and to improve patient safety and reduce the stress of clinical nursing staff.Entities:
Keywords: blood leakage detection device; hemodialysis (HD), quality of care; risk control
Mesh:
Year: 2019 PMID: 31284443 PMCID: PMC6651021 DOI: 10.3390/ijerph16132388
Source DB: PubMed Journal: Int J Environ Res Public Health ISSN: 1660-4601 Impact factor: 3.390
Figure 1Venous needle dislodgement (VND) blood leakage test on a phantom arm.
Figure 2Blood leakage test.
Figure 3Angles of blood ejection.
Questionnaires for the patients and nursing staff.
| Patients | Nursing Staff |
|---|---|
| I feel safe now. | I am able to provide the patients receiving HD with comprehensive care. |
| I feel nervous now. | I am able to pay attentive care to patients receiving HD. |
| I feel relaxed now. | I am satisfied with my interactions with patients receiving HD. |
| I am looking forward to the introduction of the blood leakage detection device. | The blood leakage detection device can affect my interactions with patients. |
| The detection device makes me feel safe. | The blood leakage detection device can increase my workload. |
| I am worried about blood leakage during HD right now. | The blood leakage detection device is complicated to install. |
| I am worried about other possible accidents during HD right now. | The blood leakage detection device is reliable. |
| HD is uncomfortable. | The blood leakage detection device makes me feel safe. |
| I feel fully rested during HD. | The blood leakage detection device can mitigate my stress when caring for patients. |
| I am satisfied with the HD. | I am looking forward to the introduction of the blood leakage detection device. |
| I am satisfied with the care I receive during HD. | |
| I am satisfied with the attentiveness of the nursing staff during HD. | |
| I am satisfied with the nurses’ interactions with me during HD. |
Results of the phantom arm test.
| Blood Flow Velocity (mL/min) | Dislodged Needle Length (mm) | Blood Leakage (Yes/No) | Detection (Yes/No) |
|---|---|---|---|
| 200 | 18 | Y | Y |
| 250 | 18 | Y | Y |
| 300 | 16 | Y | Y |
| 350 | 15 | Y | Y |
| 400 | 14 | Y | Y |
Results of the sensitivity test.
| Simulated Amount of Blood Leakage | Angle | Alarm S/F | Simulated Amount of Blood Leakage | Angle | Alarm S/F |
|---|---|---|---|---|---|
| 0.5 mL | 1 | 10/0 | 0.2 mL | 1 | 10/0 |
| 2 | 10/0 | 2 | 10/0 | ||
| 3 | 10/0 | 3 | 10/0 | ||
| 4 | 10/0 | 4 | 10/0 | ||
| 5 | 10/0 | 5 | 10/0 | ||
| 6 | 10/0 | 6 | 10/0 | ||
| 7 | 10/0 | 7 | 10/0 | ||
| 8 | 10/0 | 8 | 10/0 | ||
| 9 | 10/0 | 9 | 10/0 | ||
| 10 | 10/0 | 10 | 10/0 | ||
| 0.3 mL | 1 | 10/0 | 0.1 mL | 1 | 10/0 |
| 2 | 10/0 | 2 | 10/0 | ||
| 3 | 10/0 | 3 | 10/0 | ||
| 4 | 10/0 | 4 | 10/0 | ||
| 5 | 10/0 | 5 | 10/0 | ||
| 6 | 10/0 | 6 | 10/0 | ||
| 7 | 10/0 | 7 | 10/0 | ||
| 8 | 10/0 | 8 | 10/0 | ||
| 9 | 10/0 | 9 | 10/0 | ||
| 10 | 10/0 | 10 | 10/0 |
* S/F mean Success/Failure.
Demographic information of the participants.
|
|
|
|
|
|
|
|
|
|
| Sex | Female | Male | Male | Male | Female | Male | Male | Male |
| Age | 47 | 69 | 85 | 84 | 65 | 49 | 57 | 88 |
| HD history (years) | 22 | 1.5 | 1.0 | 7.0 | 21 | 1.0 | 8.5 | 2.0 |
|
|
|
|
| |||||
| Sex | Male | Male | Male | |||||
| Age | 64 | 55 | 52 | |||||
| HD history (years) | 9.0 | 1.5 | 3.5 |
Cluster of the classification output by the blood leakage detection device.
| True Condition | |||
|---|---|---|---|
| Condition Positive | Condition Negative | ||
| Predicted | Predicted condition positive | 471 | 6 |
| Predicted condition negative | 0 | 67 | |
Performance of distinguishing blood leakage.
| Precision | Recall |
| Accuracy Rate |
|---|---|---|---|
| 98.7% | 100% | 99.4% | 98.9% |
Patients’ subjective thoughts before and after using the device.
| Before | After | ||
|---|---|---|---|
| I feel safe now. | 2.08 | 2.42 | 0.046 * |
| I feel nervous now. | 2.42 | 2.08 | 0.046 * |
| I feel relaxed now. | 2.08 | 2.42 | 0.046 * |
| I am looking forward to the introduction of the blood leakage detection device. | 2.33 | 2.83 | 0.034 * |
| The detection device makes me feel safe. | 2.42 | 2.92 | 0.014 * |
| I am worried about blood leakage during HD right now. | 2.50 | 2.42 | 0.705 |
| I am worried about other possible accidents during HD right now. | 2.50 | 2.42 | 0.705 |
| HD is uncomfortable. | 2.42 | 2.58 | 0.414 |
| I feel fully rested during HD. | 2.33 | 2.67 | 0.102 |
| I am satisfied with the HD. | 2.67 | 2.83 | 0.317 |
| I am satisfied with the care I receive during HD. | 2.75 | 2.83 | 0.655 |
| I am satisfied with the attentiveness of the nursing staff during HD. | 2.75 | 2.83 | 0.655 |
| I am satisfied with the nurses’ interactions with me during HD. | 2.75 | 2.83 | 0.655 |
* p < 0.05
Nursing staffs’ subjective thoughts before and after using the device.
| Before | After | ||
|---|---|---|---|
| I am able to provide the patients receiving HD with comprehensive care. | 1.43 | 3.00 | 0.015 * |
| I am able to pay attentive care to patients receiving HD. | 1.29 | 3.00 | 0.014 * |
| I am satisfied with my interactions with patients receiving HD. | 1.29 | 3.00 | 0.014 * |
| The blood leakage detection device can affect my interactions with patients. | 2.00 | 2.29 | 0.746 |
| The blood leakage detection device can increase my workload. | 2.00 | 2.14 | 0.888 |
| The blood leakage detection device is complicated to install. | 2.14 | 2.00 | 0.739 |
| The blood leakage detection device is reliable. | 1.29 | 3.00 | 0.014 * |
| The blood leakage detection device makes me feel safe. | 1.57 | 3.00 | 0.015 * |
| The blood leakage detection device can mitigate my stress when caring for patients. | 1.71 | 2.86 | 0.038 * |
| I am looking forward to the introduction of the blood leakage detection device. | 1.29 | 3.00 | 0.014 * |
* p < 0.05