Elastica degeneration and intimal hyperplasia lead to Contegra® conduit failure.

Armin Darius Peivandi1, Martina Seiler1, Klaus-Michael Mueller1, Sven Martens1, Edward Malec1, Boulos Asfour2, Sabrina Lueck1.   

Abstract

OBJECTIVES: Currently, Contegra® grafts (processed bovine jugular vein conduits) are widely used for reconstructive surgery of the right ventricular outflow tract in patients with congenital heart disease (CHD). We analysed explanted Contegra conduits from 2 institutions histologically to get a possible hint at the underlying pathomechanisms of degenerative alterations and to find histological correlations of graft failure. Additionally, we compared the explants with a non-implanted processed graft and a native jugular vein obtained from a young bull.
METHODS: The explanted Contegra grafts were gathered during reoperations of 13 patients (male: n = 9, 69.2%; female: n = 4, 30.8%). After standardized histological preparation, samples were stained with dyes haematoxylin and eosin and Elastica van Gieson. Additionally, X-ray pictures revealed the extent of calcification and chelaplex (III)-descaling agent was used to decalcify selected explants.
RESULTS: Processing of the native jugular vein leads to tissue loosening and a loss of elastic fibres. For graft failure after implantation, 2 pathomechanisms were identified: original graft alteration as well as intimal hyperplasia. Elastica degeneration and rearrangement with interfibrillary matrix structures were the main developments observed within the graft itself. Intimal hyperplasia was characterized by fibrous tissue apposition, calcification and heterotopic ossification.
CONCLUSIONS: Regression of the elastic fibre network leads to rigidification of the conduit. In Contegra grafts, atherosclerosis-like changes can be considered the leading cause of graft stenosis and insufficiency. We conclude that both observed mechanisms lead to early reoperation in CHD patients.
© The Author(s) 2019. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

Entities:  

Keywords:  Congenital heart disease; Elastica degeneration; Graft failure; Intimal hyperplasia; Right ventricular outflow tract

Year:  2019        PMID: 31280306     DOI: 10.1093/ejcts/ezz199

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  4 in total

1.  Industrial Processing Induces Pericardial Patch Degeneration.

Authors:  Armin Darius Peivandi; Sven Martens; Barbara Heitplatz; Alena Guseva; Klaus-Michael Mueller; Sabrina Martens
Journal:  Front Surg       Date:  2022-05-27

2.  Promotion of right ventricular outflow tract reconstruction using a novel cardiac patch incorporated with hypoxia-pretreated urine-derived stem cells.

Authors:  Long-Mei Zhao; Long Wang; Wen-Qian Zhang; Rui Wang; Xiu-Zhen Zhang; Xiong-Xin Lei; Yan Liang; Yu-Ting Song; Qing-Yi Zhang; Ke Lin; Hui-Qi Xie
Journal:  Bioact Mater       Date:  2021-11-30

Review 3.  Current development of bovine jugular vein conduit for right ventricular outflow tract reconstruction.

Authors:  Chenggang Li; Bo Xie; Ruizhe Tan; Lijin Liang; Zhaoxiang Peng; Qi Chen
Journal:  Front Bioeng Biotechnol       Date:  2022-08-04

4.  Risk factors for early adverse outcomes after bovine jugular vein conduit implantation: influence of oversized conduit on the outcomes.

Authors:  Dong-Hee Kim; Young Kern Kwon; Eun Seok Choi; Bo Sang Kwon; Chun Soo Park; Tae-Jin Yun
Journal:  Interact Cardiovasc Thorac Surg       Date:  2022-09-09
  4 in total